Management of patients with implantable cardioverter-defibrillators and pacemakers who require radiation therapy

被引:43
作者
Brambatti, Michela [1 ,2 ]
Mathew, Rebecca [3 ]
Strang, Barbara [4 ]
Dean, Joan
Goyal, Anuja [1 ]
Hayward, Joseph E. [4 ]
Long, Laurene
DeMeis, Patty [4 ]
Smoke, Marcia [4 ]
Connolly, Stuart J. [1 ]
Morillo, Carlos A. [1 ]
Amit, Guy [1 ]
Capucci, Alessandro
Healey, Jeff S. [1 ]
机构
[1] McMaster Univ, Populat Hlth Res Inst, Hamilton, ON, Canada
[2] Univ Politecn Marche, Clin Cardiol & Aritmol, I-60100 Ancona, Italy
[3] McMaster Univ, Dept Internal Med, Hamilton, ON, Canada
[4] Hamilton Hlth Sci, Ctr Canc, Hamilton, ON, Canada
关键词
Pacemaker; Implantable cardioverter-defibrillator; Radiation therapy; Malfunction; Care path; CARDIAC-PACEMAKERS; LATEST GENERATION; ONCOLOGY PATIENTS; RADIOTHERAPY; IRRADIATION; DEVICE; SOCIETY; PROJECT;
D O I
10.1016/j.hrthm.2015.06.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Radiation therapy (RT) may pose acute and long-term risks for patients with cardiac implantable electronic devices (CIEDs), including pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs). However, the frequency of these problems has not been accurately defined. OBJECTIVE The purpose of this study was to determine the prevalence of CIEDs among patients requiring RT and report the common CIED-related problems when patients are managed according to a standard clinical care path. METHODS In a single tertiary-care center, we prospectively screened all patients requiring RT and identified patients with ICDs or PMs. We collected clinical data about their cancer, RT treatment plan, and CIED. Radiation dose to the device was estimated in all patients, and any device malfunction during RT was documented. RESULTS Of the 34,706 consecutive patients receiving RT, 261 patients (0.8%, mean age 77.9 +/- 9.4 years) had an implantable cardiac device: 54 (20.7%) ICDs and 207 (79.3%) PMs. The site of RT was head and neck (27.4%), chest (30.0%), and abdomen/pelvis (32.6%). Using our care path, 63.2% of patients required continuous cardiac monitoring, 14.6% required device reprogramming, 18.8% required magnet application during RT, and 3.4%, required device repositioning to the contralateral side before RT. Four patients (1.5%) had inappropriate device function during RT: 3 experienced hemodynamically tolerated ventricular pacing at the maximum sensor rate, and 1 experienced a device power-on-reset. No patient died or suffered permanent device failure. CONCLUSION Nearly 1% of patients receiving RT in this series has a PM or ICD. However, with a systematic policy of risk assessment and patient management, significant device-related complications are rare.
引用
收藏
页码:2148 / 2154
页数:7
相关论文
共 25 条
  • [1] Brady LW, 2007, PEREZ BRADYS PRINCIP, P2043
  • [2] The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the Perioperative Management of Patients with Implantable Defibrillators, Pacemakers and Arrhythmia Monitors: Facilities and Patient Management This document was developed as a joint project with the American Society of Anesthesiologists (ASA), and in collaboration with the American Heart Association (AHA), and the Society of Thoracic Surgeons (STS)
    Crossley, George H.
    Poole, Jeanne E.
    Rozner, Marc A.
    Asirvatham, Samuel J.
    Cheng, Alan
    Chung, Mina K.
    Ferguson, T. Bruce, Jr.
    Gallagher, John D.
    Gold, Michael R.
    Hoyt, Robert H.
    Irefin, Samuel
    Kusumoto, Fred M.
    Moorman, Liza Prudente
    Thompson, Annemarie
    [J]. HEART RHYTHM, 2011, 8 (07) : 1114 - 1154
  • [3] Irradiation of pacemakers and cardio-defibrillators in patients submitted to radiotherapy: a clinical experience
    Ferrara, Teresa
    Baiotto, Barbara
    Malinverni, Giuseppe
    Caria, Nicola
    Garibaldi, Elisabetta
    Barboni, Gilberto
    Stasi, Michele
    Gabriele, Pietro
    [J]. TUMORI, 2010, 96 (01) : 76 - 83
  • [4] Frizzel B, 2009, COMMUNITY ONCOL, V6, P469
  • [5] IMPLANTED CARDIAC DEFIBRILLATOR CARE IN RADIATION ONCOLOGY PATIENT POPULATION
    Gelblum, Daphna Y.
    Amols, Howard
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (05): : 1525 - 1531
  • [6] Malfunctions of Implantable Cardiac Devices in Patients Receiving Proton Beam Therapy: Incidence and Predictors
    Gomez, Daniel R.
    Poenisch, Falk
    Pinnix, Chelsea C.
    Sheu, Tommy
    Chang, Joe Y.
    Memon, Nada
    Mohan, Radhe
    Rozner, Marc A.
    Dougherty, Anne H.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (03): : 570 - 575
  • [7] Management of radiation oncology patients with a pacemaker or ICD: A new comprehensive practical guideline in The Netherlands
    Hurkmans, Coen W.
    Knegjens, Joost L.
    Oei, Bing S.
    Maas, Ad J. J.
    Uiterwaal, G. J.
    van der Borden, Arnoud J.
    Ploegmakers, Marleen M. J.
    van Erven, Lieselot
    [J]. RADIATION ONCOLOGY, 2012, 7
  • [8] Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators
    Hurkmans, CW
    Scheepers, E
    Springorum, BGF
    Uiterwaal, H
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (01): : 282 - 289
  • [9] Influence of radiotherapy on the latest generation of pacemakers
    Hurkmans, CW
    Scheepers, E
    Springorum, BGF
    Uiterwaal, H
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) : 93 - 98
  • [10] Radiotherapy in patients with cardiac pacemakers
    Last, A