Biomechanics of a short stem: In vitro primary stability and stress shielding of a conservative cementless hip stem

被引:77
作者
Bieger, Ralf [1 ]
Ignatius, Anita [2 ]
Reichel, Heiko [1 ]
Duerselen, Lutz [2 ]
机构
[1] Univ Hosp Ulm, Ctr Musculoskeletal Res, Dept Orthopaed Surg, D-89081 Ulm, Germany
[2] Univ Ulm, Ctr Musculoskeletal Res, Inst Orthopaed Res & Biomech, D-89069 Ulm, Germany
关键词
arthroplasty; hip; cadaver; stress; mechanical; micromotion; BONE-MINERAL DENSITY; FEMORAL STEMS; STRAIN DISTRIBUTION; ARTHROPLASTY; IMPLANTS; PROSTHESIS; FAILURE; REPLACEMENT; MIGRATION; COMPONENT;
D O I
10.1002/jor.22349
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Short stem prostheses provide conservative surgery and favorable metaphyseal load transmission. However, clinical long-term results are lacking. Therefore, in vitro trials can be used to predict bone-implant performance. In this in vitro study, primary stability and stress shielding of a new cementless short stem implant was evaluated in comparison to a straight stem using nine pairs of human cadaver femurs. Primary stability, including reversible micromotion and irreversible migration, was assessed in a hip simulator. Furthermore, changes in the pattern of cortical strain were evaluated. The short stem was more resistant to reversible micromotion and irreversible migration into retroversion. Axial stability was similar, with mean reversible micromotions of 9 mu m for the short stem and 7 mu m for the straight stem. Proximal load transmission was more physiological with the short stem, though both implants could not avoid stress shielding in Gruen zones 1 and 7. Primary stability of the short stem prosthesis was not negatively influenced compared to the straight shaft. Furthermore, proximal femoral strain pattern was more physiological after insertion of the short stem prosthesis. (c) 2013 Orthopaedic Research Society Published by Wiley Periodicals, Inc. J Orthop Res 31:1180-1186, 2013
引用
收藏
页码:1180 / 1186
页数:7
相关论文
共 32 条
  • [1] Administration USFaD, 2007, GUID IND FDA STAFF N, V13
  • [2] Survival of the Cementless SpotornoA® Stem in the Second Decade
    Aldinger, Peter R.
    Jung, Alexander W.
    Breusch, Steffen J.
    Ewerbeck, Volker
    Parsch, Dominik
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (09) : 2297 - 2304
  • [3] HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (08) : 969 - 990
  • [4] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [5] Primary stability and strain distribution of cementless hip stems as a function of implant design
    Bieger, Ralf
    Ignatius, Anita
    Decking, Ralf
    Claes, Lutz
    Reichel, Heiko
    Duerselen, Lutz
    [J]. CLINICAL BIOMECHANICS, 2012, 27 (02) : 158 - 164
  • [6] Biomechanics of short hip endoprostheses - The risk of bone failure increases with decreasing implant size
    Bishop, Nicholas E.
    Burton, Andrew
    Maheson, Marci
    Morlock, Michael M.
    [J]. CLINICAL BIOMECHANICS, 2010, 25 (07) : 666 - 674
  • [7] Three-dimensional primary stability of cementless femoral stems
    Buhler, DW
    Berlemann, U
    Lippuner, K
    Jaeger, P
    Nolte, LP
    [J]. CLINICAL BIOMECHANICS, 1997, 12 (02) : 75 - 86
  • [8] Bone Remodeling Characteristics of a Short-Stemmed Total Hip Replacement
    Chen, Hsiang-Ho
    Morrey, Bernard F.
    An, Kai-Nan
    Luo, Zong-Ping
    [J]. JOURNAL OF ARTHROPLASTY, 2009, 24 (06) : 945 - 950
  • [9] MEDICAL DEVICES Europeans are left to their own devices
    Cohen, Deborah
    Billingsley, Matthew
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 2011, 342
  • [10] A critical analysis of stress shielding evaluation of hip prostheses
    Cristofolini, L
    [J]. CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) : 409 - 483