Use of a biocompatible fiducial marker in evaluating the accuracy of computed tomography image registration

被引:24
作者
Ellis, R
ToksvigLarsen, S
Marcacci, M
Caramella, D
Fadda, M
机构
[1] UNIV LUND HOSP,DEPT ORTHOPAED SURG,S-22185 LUND,SWEDEN
[2] RIZZOLI ORTHOPAED INST,BIOMECH LAB,BOLOGNA,ITALY
关键词
registration; validation; computed tomography; computer-assisted surgery;
D O I
10.1097/00004424-199610000-00008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. Accurate registration of computed tomography (CI) images to the patient is crucial for computer-assisted surgery, Markers used in roentgen stereo-grammetric analysis (RSA) can be located in a CT scan using a novel approach and also can be located physically, Roentgen stereo-grammetric analysis data act as ''ground truth'' for the three-dimensional marker locations. METHODS. Two foam-bone phantoms were marked, The markers were scanned seven times with RSA, three times with axial CT, and contacted four times with a coordinate measuring arm, Root-mean-square (RMS) errors were derived for the registrations. RESULTS. Computed tomography and RSA data register to 0.15 mm RMS error, Computed tomography and arm data register to 0.25 mm, The markers are biocompatible, and the coordinate measuring arm is usable in an operating room. CONCLUSIONS. Typical in vitro registration errors are approximately 2 mm, The authors have developed a marker that provides superior registration, and a procedure that can be used for in vivo studies.
引用
收藏
页码:658 / 667
页数:10
相关论文
共 33 条
[1]   EFFECT OF LUMBAR ORTHOSIS ON INTERVERTEBRAL MOBILITY - A ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS [J].
AXELSSON, P ;
JOHNSSON, R ;
STROMQVIST, B .
SPINE, 1992, 17 (06) :678-681
[2]   USE OF A FRAMELESS, ARMLESS STEREOTAXIC WAND FOR BRAIN-TUMOR LOCALIZATION WITH 2-DIMENSIONAL AND 3-DIMENSIONAL NEUROIMAGING [J].
BARNETT, GH ;
KORMOS, DW ;
STEINER, CP ;
WEISENBERGER, J .
NEUROSURGERY, 1993, 33 (04) :674-678
[3]   A METHOD FOR REGISTRATION OF 3-D SHAPES [J].
BESL, PJ ;
MCKAY, ND .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1992, 14 (02) :239-256
[4]   EXTRACRANIAL APPLICATION OF THE FRAMELESS STEREOTAXIC OPERATING MICROSCOPE - EXPERIENCE WITH LUMBAR SPINE [J].
BRODWATER, BK ;
ROBERTS, DW ;
NAKAJIMA, T ;
FRIETS, EM ;
STROHBEHN, JW ;
KELLY, PJ ;
FRIEDMAN, WA .
NEUROSURGERY, 1993, 32 (02) :209-213
[5]   VARIABLES AFFECTING THE ACCURACY OF STEREOTAXIC LOCALIZATION USING COMPUTERIZED-TOMOGRAPHY [J].
BUCHOLZ, RD ;
HO, HW ;
RUBIN, JP .
JOURNAL OF NEUROSURGERY, 1993, 79 (05) :667-673
[6]   FRAMELESS STEREOTAXY IN CHILDREN [J].
DRAKE, JM ;
PRUDENCIO, J ;
HOLOWKA, S ;
RUTKA, JT ;
HOFFMAN, HJ ;
HUMPHREYS, RP .
PEDIATRIC NEUROSURGERY, 1994, 20 (02) :152-159
[7]  
FADDA M, 1994, P 1 INT S MED ROB CO, P26
[8]  
FIDEN T, 1992, ACTA ORTHOP SCAND, V63, P80
[9]   PREDICTION OF TOTAL KNEE MOTION USING A 3-DIMENSIONAL COMPUTER-GRAPHICS MODEL [J].
GARG, A ;
WALKER, PS .
JOURNAL OF BIOMECHANICS, 1990, 23 (01) :45-&
[10]  
GUTHRIE BL, 1992, CLIN NEUR, V38, P112