Detection of Endotoxin Concentration Using Piezoelectric Based Biosensor System

被引:5
|
作者
Xiong, Xingliang [1 ]
Wang, Shiming [1 ]
Zhang, Yan [1 ]
Chen, Longcong [1 ]
机构
[1] Chongqing Med Univ, Biomed Engn Lab, Chongqing, Peoples R China
来源
MECHANICAL ENGINEERING AND INTELLIGENT SYSTEMS, PTS 1 AND 2 | 2012年 / 195-196卷
关键词
Endotoxin; Quartz crystal; Biosensors; LIMULUS AMEBOCYTE LYSATE;
D O I
10.4028/www.scientific.net/AMM.195-196.874
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Endotoxins are ubiquitous in the environment and represent important pathogenic molecules. In this paper, we present a sensitive and reliable method for quantitation and detection of endotoxin based on piezoelectric biosensors, which monitor the gel formation (causing viscosity change) when Limulus Amoebocyte Lysate (LAL) mixed with endotoxin. The resonance frequency shifts of the sensors in contact with different concentration endotoxin were recorded as a function of time. Results showed that there has good relationship between the logarithmic concentration of endotoxin and the maximum clotting rate as well as the corresponding clotting time. The detection limits could be gone down to 0.1pg/ml and the time-consuming is about 1 hour using this system. By comparison of the traditional methods (rabbit pyrogen test and LAL test) for detection of endotoxin, the proposed sensor is much simpler, more precise and lower detection limits.
引用
收藏
页码:874 / 878
页数:5
相关论文
共 50 条
  • [1] Novel integrated and portable endotoxin detection system based on an electrochemical biosensor
    Zuzuarregui, Ana
    Souto, David
    Perez-Lorenzo, Eva
    Arizti, Fernando
    Sanchez-Gomez, Susana
    Martinez de Tejada, Guillermo
    Brandenburg, Klaus
    Arana, Sergio
    Mujika, Maite
    ANALYST, 2015, 140 (02) : 654 - 660
  • [2] Recent advances in biosensor based endotoxin detection
    Das, A. P.
    Kumar, P. S.
    Swain, S.
    BIOSENSORS & BIOELECTRONICS, 2014, 51 : 62 - 75
  • [3] Capacitive biosensor for detection of endotoxin
    Limbut, Warakorn
    Hedstrom, Martin
    Thavarungkul, Panote
    Kanatharana, Proespichaya
    Mattiasson, Bo
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 389 (02) : 517 - 525
  • [4] Capacitive biosensor for detection of endotoxin
    Warakorn Limbut
    Martin Hedström
    Panote Thavarungkul
    Proespichaya Kanatharana
    Bo Mattiasson
    Analytical and Bioanalytical Chemistry, 2007, 389 : 517 - 525
  • [5] Piezoelectric Biosensor based on ultrasensitive MEMS system
    He, Ji-Huan
    He, Chun-Hui
    Qian, Man-Yu
    Alsolami, Abdulrahman Ali
    SENSORS AND ACTUATORS A-PHYSICAL, 2024, 376
  • [6] A Rapid Detection of Pesticide Residue Based on Piezoelectric Biosensor
    Shang, Zhan-jiang
    Xu, Yan-li
    Gu, Yun-xu-zi
    Wang, Yi
    Wei, Dong-xu
    Zhan, Li-li
    CEIS 2011, 2011, 15
  • [8] A dielectric loss angle based portable biosensor system for bacterial concentration detection
    Chen, Jingyao
    Cai, Jie
    Hu, Hao
    Huang, Xingjian
    Yi, Tian
    Wang, Kexing
    Pan, Siyi
    RSC ADVANCES, 2015, 5 (104) : 85919 - 85927
  • [9] PIEZOELECTRIC CRYSTAL BIOSENSOR SYSTEM FOR DETECTION OF ESCHERICHIA-COLI
    MURAMATSU, H
    WATANABE, Y
    HIKUMA, M
    ATAKA, T
    KUBO, I
    TAMIYA, E
    KARUBE, I
    ANALYTICAL LETTERS, 1989, 22 (09) : 2155 - 2166
  • [10] An embedded portable biosensor system for bacterial concentration detection
    Grossi, M.
    Lanzoni, M.
    Pompei, A.
    Lazzarini, R.
    Matteuzzi, D.
    Ricco, B.
    BIOSENSORS & BIOELECTRONICS, 2010, 26 (03): : 983 - 990