UPLC-ESI-QTOF/MS and multivariate data analysis for blood plasma and serum metabolomics: Effect of experimental artefacts and anticoagulant

被引:169
作者
Barri, Thaer [1 ]
Dragsted, Lars Ove [1 ]
机构
[1] Univ Copenhagen, Fac Sci, Dept Human Nutr, DK-1958 Copenhagen, Denmark
关键词
Metabolomics; Mass spectrometry; Blood plasma; Serum; Anticoagulant; Matrix effect; TANDEM MASS-SPECTROMETRY; LIQUID-CHROMATOGRAPHY; GAS-CHROMATOGRAPHY; COLLECTION TUBES; INTERFERENCE; METABONOMICS; VALIDATION; H-1-NMR; SAMPLES; CANCER;
D O I
10.1016/j.aca.2013.01.015
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Clotting and anticoagulation of blood samples may give rise to different metabolic profiles of serum and plasma samples, respectively. The anticoagulant used for blood plasma preparation may affect the resulting metabolic profile due to different mechanisms involved in anticoagulation by various agents, e.g. heparin, EDTA and citrate. In the present study, we looked into metabolite and other differences in matched serum and plasma samples and different plasma preparations by using untargeted UPLC-ESI-QTOF/MS profiling and multivariate data analysis (PCA and OPLS-DA). Metabolite differences between serum and plasma samples were mainly related to small peptides reflecting presence or absence of coagulation. Only subtle metabolite differences between the different plasma preparations were noticed, which were primarily related to ion suppression or enhancement caused by citrate and EDTA anticoagulants. For the first time, we also report that anticoagulant counter cation (Na+ or K+) in Na-citrate and K-EDTA plasma can make some metabolites more dominant in ESI-MS. Polymeric material residues originating from blood collection tubes for serum preparation were observed only in serum samples. Hypoxanthine and xanthine were found at higher levels in serum than in plasma samples, possibly due to release from the clot. Mass spectral features of sodium formate and potassium formate ion clusters were detected in citrate and EDTA plasma samples, respectively, originating from formate in mobile phase and Na+ (in Na-citrate tubes) and K+ (in K-EDTA tubes). Among the anticoagulants, heparin is recommended for plasma samples used for LC-ESI/MS-based metabolomics of hydrophilic compounds because no plasma interferences or matrix effects were noticed for this polarity range. Citrate and EDTA should be avoided since interferences and serious matrix effects were encountered on some co-eluting polar metabolites. Serum is recommended as a second choice and an alternative to plasma. In conclusion, heparin plasma or serum should be the order of best choice for LC-ESI/MS-based metabolomics research. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:118 / 128
页数:11
相关论文
共 31 条
  • [1] The ion suppression phenomenon in liquid chromatography-mass spectrometry and its consequences in the field of residue
    Antignac, JP
    de Wasch, K
    Monteau, F
    De Brabander, H
    Andre, F
    Le Bizec, B
    [J]. ANALYTICA CHIMICA ACTA, 2005, 529 (1-2) : 129 - 136
  • [2] Influences of biofluid sample collection and handling procedures on GC-MS based metabolomic studies
    Bando, Kiyoko
    Kawahara, Rui
    Kunimatsu, Takeshi
    Sakai, Jun
    Kimura, Juki
    Funabashi, Hitoshi
    Seki, Takaki
    Bamba, Takeshi
    Fukusaki, Eiichiro
    [J]. JOURNAL OF BIOSCIENCE AND BIOENGINEERING, 2010, 110 (04) : 491 - 499
  • [3] Metabolic fingerprinting of high-fat plasma samples processed by centrifugation- and filtration-based protein precipitation delineates significant differences in metabolite information coverage
    Barri, Thaer
    Holmer-Jensen, Jens
    Hermansen, Kjeld
    Dragsted, Lars O.
    [J]. ANALYTICA CHIMICA ACTA, 2012, 718 : 47 - 57
  • [4] Large-scale human metabolomics studies: A strategy for data (pre-) processing and validation
    Bijlsma, S
    Bobeldijk, L
    Verheij, ER
    Ramaker, R
    Kochhar, S
    Macdonald, IA
    van Ommen, B
    Smilde, AK
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (02) : 567 - 574
  • [5] Immunoassay interference by a commonly used blood collection tube additive, the organosilicone surfactant Silwet L-720
    Bowen, RAR
    Chan, Y
    Ruddel, ME
    Hortin, GL
    Csako, G
    Demosky, SJ
    Remaley, AT
    [J]. CLINICAL CHEMISTRY, 2005, 51 (10) : 1874 - 1882
  • [6] Effect of blood collection tubes on total triiodothyronine and other laboratory assays
    Bowen, RAR
    Chan, Y
    Cohen, J
    Rehak, NN
    Hortin, GL
    Csako, G
    Remaley, AT
    [J]. CLINICAL CHEMISTRY, 2005, 51 (02) : 424 - 433
  • [7] Metabonomics study of liver cancer based on ultra performance liquid chromatography coupled to mass spectrometry with HILIC and RPLC separations
    Chen, Jing
    Wang, Wenzhao
    Lv, Shen
    Yin, Peiyuan
    Zhao, Xinjie
    Lu, Xin
    Zhang, Fengxia
    Xu, Guowang
    [J]. ANALYTICA CHIMICA ACTA, 2009, 650 (01) : 3 - 9
  • [8] Characterization of Differences between Blood Sample Matrices in Untargeted Metabolomics
    Denery, Judith R.
    Nunes, Ashlee A. K.
    Dickerson, Tobin J.
    [J]. ANALYTICAL CHEMISTRY, 2011, 83 (03) : 1040 - 1047
  • [9] Optimisation of collection, storage and preparation of rat plasma for 1H NMR spectroscopic analysis in toxicology studies to determine inherent variation in biochemical profiles
    Deprez, S
    Sweatman, BC
    Connor, SC
    Haselden, JN
    Waterfield, CJ
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 30 (04) : 1297 - 1310
  • [10] Comparison of serum versus plasma collection in gas chromatography - Mass spectrometry-based metabolomics
    Dettmer, Katja
    Almstetter, Martin F.
    Appel, Inka J.
    Nuernberger, Nadine
    Schlamberger, Gregor
    Gronwald, Wolfram
    Meyer, Heinrich H. D.
    Oefner, Peter J.