In Vivo Blood Glucose Quantification Using Raman Spectroscopy

被引:125
作者
Shao, Jingwei [2 ]
Lin, Manman [1 ,3 ]
Li, Yongqing [4 ]
Li, Xue [1 ,3 ]
Liu, Junxian [3 ]
Liang, Jianpin [1 ]
Yao, Huilu [1 ]
机构
[1] Guangxi Acad Sci, Biophys Lab, Nanning, Peoples R China
[2] Fuzhou Univ, Coll Chem & Chem Engn, Fuzhou 350002, Peoples R China
[3] Guangxi Normal Univ, Coll Phys & Technol, Guilin, Peoples R China
[4] E Carolina Univ, Dept Phys, Greenville, NC 27858 USA
基金
中国博士后科学基金;
关键词
NEAR-INFRARED SPECTROSCOPY; SENSORS; SERUM;
D O I
10.1371/journal.pone.0048127
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
We here propose a novel Raman spectroscopy method that permits the noninvasive measurement of blood glucose concentration. To reduce the effects of the strong background signals produced by surrounding tissue and to obtain the fingerprint Raman lines formed by blood analytes, a laser was focused on the blood in vessels in the skin. The Raman spectra were collected transcutaneously. Characteristic peaks of glucose (1125 cm(-1)) and hemoglobin (1549 cm(-1)) were observed. Hemoglobin concentration served as an internal standard, and the ratio of the peaks that appeared at 1125 cm(-1) and 1549 cm(-1) peaks was used to calculate the concentration of blood glucose. We studied three mouse subjects whose blood glucose levels became elevated over a period of 2 hours using a glucose test assay. During the test, 25 Raman spectra were collected transcutaneously and glucose reference values were provided by a blood glucose meter. Results clearly showed the relationship between Raman intensity and concentration. The release curves were approximately linear with a correlation coefficient of 0.91. This noninvasive methodology may be useful for the study of blood glucose in vivo.
引用
收藏
页数:6
相关论文
共 34 条
[1]  
Alfano RR, 2008, DETECTION GLUCOSE LE
[2]   Accurate Spectroscopic Calibration for Noninvasive Glucose Monitoring by Modeling the Physiological Glucose Dynamics [J].
Barman, Ishan ;
Kong, Chae-Ryon ;
Singh, Gajendra P. ;
Dasari, Ramachandra R. ;
Feld, Michael S. .
ANALYTICAL CHEMISTRY, 2010, 82 (14) :6104-6114
[3]  
Bruulsema JT, 1996, C BIOM OPT SPECTR DI
[4]   Effect of hemoglobin concentration variation on the accuracy and precision of glucose analysis using tissue modulated, noninvasive, in vivo Raman spectroscopy of human blood:: a small clinical study -: art. no. 031111 [J].
Chaiken, J ;
Finney, W ;
Knudson, PE ;
Weinstock, RS ;
Khan, M ;
Bussjager, RJ ;
Hagrman, D ;
Hagrman, P ;
Zhao, YW ;
Peterson, CM ;
Peterson, K .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (03)
[5]   LASER PHOTOACOUSTIC DETERMINATION OF PHYSIOLOGICAL GLUCOSE-CONCENTRATIONS IN HUMAN WHOLE-BLOOD [J].
CHRISTISON, GB ;
MACKENZIE, HA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1993, 31 (03) :284-290
[6]   A two-level partial least squares system for non-invasive blood glucose concentration prediction [J].
Chuah, Zheng-Ming ;
Paramesran, Raveendran ;
Thambiratnam, Kavintheran ;
Poh, Sin-Chew .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2010, 104 (02) :347-351
[7]  
Davidson M.B., 1991, Diabetes Mellitus: Diagnosis and Treatment
[8]   Raman spectroscopy for noninvasive glucose measurements [J].
Enejder, AMK ;
Scecina, TG ;
Oh, J ;
Hunter, M ;
Shih, WC ;
Sasic, S ;
Horowitz, GL ;
Feld, MS .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (03) :1-9
[9]   Blood analysis by Raman spectroscopy [J].
Enejder, AMK ;
Koo, TW ;
Oh, J ;
Hunter, M ;
Sasic, S ;
Feld, MS ;
Horowitz, GL .
OPTICS LETTERS, 2002, 27 (22) :2004-2006
[10]   Noninvasive monitoring of glucose concentration with optical coherence tomography [J].
Esenaliev, RO ;
Larin, KV ;
Larina, IV ;
Motamedi, M .
OPTICS LETTERS, 2001, 26 (13) :992-994