The wear pattern in metal-on-metal hip prostheses

被引:48
作者
Anissian, HL [1 ]
Stark, A
Good, V
Dahlstrand, H
Clarke, IC
机构
[1] Karolinska Hosp, Dept Orthopaed Surg, S-17176 Stockholm, Sweden
[2] Loma Linda Univ, Med Ctr, Dept Orthopaed Surg, Loma Linda, CA 92350 USA
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 2001年 / 58卷 / 06期
关键词
implants; hip-joint simulator; wear; run-in phase; steady-state phase; metal-on-metal prosthesis;
D O I
10.1002/jbm.1068
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accumulated data suggest that there is a strong correlation between polyethylene wear and osteolysis, which ultimately leads to prosthetic loosening. Second-generation metal-on-metal prostheses have been introduced, with an eye toward resolving this wear-induced osteolysis problem. The metal particles and ions are biologically active and can affect the cell homeostasis. Thus, defining the wear pattern and ratio of a given metal-on-metal prosthesis system is desirable. An early high-wear or run-in phase followed by a low-wear phase or steady state has been suggested for metal-on-metal hip prostheses. The aim of this study was to define the wear pattern of metal-on-metal bearings. The prosthesis systems were tested in a joint simulator. An early accelerated wear phase transformed to a slower wear phase after 700,000 cycles. The run-in and steady-state wear rates for combined head and liner averaged 2.22 mm(3)/Mc and 1.0 mm(3)/ million cycles, respectively. The metal-on-metal prosthesis featured a biphasic wear trend with the wear ratio in run in being more than twofold higher than the steady-state phase. (C) 2001 John Wiley & Sons, Inc.
引用
收藏
页码:673 / 678
页数:6
相关论文
共 50 条
  • [41] Metal-on-metal articulation in total hip arthroplasty
    Marchetti, PG
    Binazzi, R
    Vaccari, V
    Manca, A
    Rossi, R
    Campagna, A
    Magnani, M
    Baldarelli, M
    Pasini, E
    BIOCERAMICS, 2000, 192-1 : 975 - 977
  • [42] The effect of microstructure on the wear of cobalt-based alloys used in metal-on-metal hip implants
    Varano, R
    Bobyn, JD
    Medley, JB
    Yue, S
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 145 - 159
  • [43] Lubrication of metal-on-metal hip joints: The effect of protein content and load on film formation and wear
    Myant, C.
    Underwood, R.
    Fan, J.
    Cann, P. M.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 6 : 30 - 40
  • [44] Total hip replacements with metal-on-metal bearings
    Milosev, Ingrid
    Kovac, Simon
    Topolovec, Matevz
    Pisot, Venceslav
    Trebse, Rihard
    ZDRAVNISKI VESTNIK-SLOVENIAN MEDICAL JOURNAL, 2009, 78 : 34 - 40
  • [45] The Effect of Bottom Profile Dimples on the Femoral Head on Wear in Metal-on-Metal Total Hip Arthroplasty
    Jamari, J.
    Ammarullah, Muhammad Imam
    Saad, Amir Putra Md
    Syahrom, Ardiyansyah
    Uddin, Mohammad
    van der Heide, Emile
    Basri, Hasan
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2021, 12 (02)
  • [46] A comparative joint simulator study of the wear of metal-on-metal and alternative material combinations in hip replacements
    Goldsmith, AAJ
    Dowson, D
    Isaac, GH
    Lancaster, JG
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H1) : 39 - 47
  • [47] Assessment of a recent tribocorrosion model for wear of metal-on-metal hip joints: Comparison between model predictions and simulator results
    Cao, Shoufan
    Mischler, Stefano
    WEAR, 2016, 362 : 170 - 178
  • [48] Investigation of Taper Failure in a Contemporary Metal-on-Metal Hip Arthroplasty System Through Examination of Unused and Explanted Prostheses
    Langton, David
    Ahmed, Imran
    Avery, Peter
    Bone, Martin
    Cooke, Nicholas
    Deehan, David
    Duffy, Patrick
    Foguet, Pedro
    Green, Stephen
    Holland, James
    Jafri, Anwar
    Longstaff, Lee
    Lord, James
    Loughead, Jonathan
    Meek, R. M. Dominic
    Murray, Helen
    Nanu, Ananda
    Nargol, A. V. F.
    Scholes, Susan Claire
    Sidaginamale, Raghavendra Prasad
    Waller, Susan
    Joyce, Thomas
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2017, 99 (05) : 427 - 436
  • [49] On the matter of synovial fluid lubrication: Implications for Metal-on-Metal hip tribology
    Myant, Connor
    Cann, Philippa
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 34 : 338 - 348
  • [50] Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream
    Dalila De Pasquale
    Susanna Stea
    Stefano Squarzoni
    Barbara Bordini
    Marilina Amabile
    Simona Catalani
    Pietro Apostoli
    Aldo Toni
    International Orthopaedics, 2014, 38 : 469 - 475