A microwave resonance dew-point hygrometer

被引:15
|
作者
Underwood, R. J. [1 ]
Cuccaro, R. [2 ]
Bell, S. [1 ]
Gavioso, R. M. [2 ]
Ripa, D. Madonna [2 ]
Stevens, M. [1 ]
de Podesta, M. [1 ]
机构
[1] Natl Phys Lab, Teddington TW11 0LW, Middx, England
[2] Ist Nazl Ric Metrol, I-10135 Turin, Italy
关键词
hygrometry; dew point; microwave resonator; condensation detection; surface layers; quasi-spherical resonator; WATER; PERMITTIVITY; RESONATORS; SENSOR; VAPOR;
D O I
10.1088/0957-0233/23/8/085905
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We report the first measurements of a quasi-spherical microwave resonator used as a dew-point hygrometer. In conventional dew-point hygrometers, the condensation of water from humid gas flowing over a mirror is detected optically, and the mirror surface is then temperature-controlled to yield a stable condensed layer. In our experiments we flowed moist air from a humidity generator through a quasi-spherical resonator and detected the onset of condensation by measuring the frequency ratio of selected microwave modes. We verified the basic operation of the device over the dew-point range 9.5-13.5 degrees C by comparison with calibrated chilled-mirror hygrometers. These tests indicate that the microwave method may allow a quantitative estimation of the volume and thickness of the water layer which is condensed on the inner surface of the resonator. The experiments reported here are preliminary due to the limited time available for the work, but show the potential of the method for detecting not only water but a variety of other liquid or solid condensates. The robust all-metal construction should make the device appropriate for use in industrial applications over a wide range of temperatures and pressures.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] LASER DEW-POINT HYGROMETER
    MATSUMOTO, S
    TOYOOKA, S
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1995, 34 (01): : 316 - 320
  • [2] A DEW-POINT HYGROMETER FOR FIELD USE
    HOLBO, HR
    AGRICULTURAL METEOROLOGY, 1981, 24 (02): : 117 - 130
  • [3] HUMIDITY MONITOR AND DEW-POINT HYGROMETER
    PARKER, HW
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1982, 338 : 4 - 4
  • [4] AN IMPROVED ELECTRONIC DEW-POINT HYGROMETER
    BARRETT, EW
    HERNDON, LR
    JOURNAL OF METEOROLOGY, 1951, 8 (01): : 40 - 51
  • [5] A new form of a dew-point hygrometer
    Williams, SR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1933, 4 (10): : 549 - 550
  • [6] MEASUREMENT OF DEW-POINT TEMPERATURE WITH LICL DEW-POINT HYGROMETER BY FEUTRON COMPANY, GREIZ
    HENNING, H
    ZEITSCHRIFT FUR METEOROLOGIE, 1971, 22 (8-10): : 283 - &
  • [7] Fast dew-point hygrometer for laryngolgy applications
    Jachowicz, R
    Weremczuk, J
    Paczesny, D
    Rapiejko, P
    PROCEEDINGS OF THE IEEE SENSORS 2004, VOLS 1-3, 2004, : 1488 - 1491
  • [8] FURTHER IMPROVEMENTS IN THE ELECTRONIC DEW-POINT HYGROMETER
    BARRETT, EW
    SLATER, RL
    NEWTON, KE
    JOURNAL OF METEOROLOGY, 1955, 12 (04): : 308 - 313
  • [9] HYGROMETER WITH FIBER OPTIC DEW-POINT DETECTOR
    JACHOWICZ, RS
    ZALEWSKI, D
    SENSORS AND ACTUATORS A-PHYSICAL, 1994, 42 (1-3) : 503 - 507
  • [10] FAST-RESPONSE DEW-POINT HYGROMETER
    BOVSHEVEROV, VM
    IZVESTIYA AKADEMII NAUK SSSR FIZIKA ATMOSFERY I OKEANA, 1976, 12 (12): : 1314 - 1317