Quality assurance of asymmetric jaw alignment using 2D diode array

被引:1
|
作者
Kim, Sun Mo [1 ,2 ]
Chmielewski, Renata [1 ]
Abbas, Ahmar [1 ]
Yeung, Ivan W. T. [1 ,2 ,3 ]
Moseley, Douglas J. [1 ,2 ,3 ]
机构
[1] Southlake Reg Hlth Ctr, Stronach Reg Canc Ctr, Dept Med Phys, Newmarket, ON L3Y 2P9, Canada
[2] Univ Hlth Network, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[3] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
关键词
asymmetric jaw alignment; junction gap; 2D diode array; MapCHECK2; RADIOCHROMIC FILM; ABUTTED FIELDS; DOSIMETRY; HEAD; COLLIMATORS; TIME; NECK;
D O I
10.1118/1.4828784
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A method using a 2D diode array is proposed to measure the junction gap (or overlap) and dose with high precision for routine quality assurance of the asymmetric jaw alignment. Methods: The central axis (CAX) of the radiation field was determined with a 15 x 15 cm(2) photon field at four cardinal collimator angles so that the junction gap (or overlap) can be measured with respect to the CAX. Two abutting fields having a field size of 15 cm (length along the axis parallel to the junction) x 7.5 cm (width along the axis perpendicular to the junction) were used to irradiate the 2D diode array (MapCHECK2) with 100 MU delivered at the photon energy of 6 MV. The collimator was slightly rotated at 15 degrees with respect to the beam central axis to increase the number of diodes effective on the measurement of junction gap. The junction gap and dose measured in high spatial resolution were compared to the conventional methods using an electronic portal imaging device (EPID) and radiochromic film, respectively. In addition, the reproducibility and sensitivity of the proposed method to the measurements of junction gap and dose were investigated. Results: The junction gap (or overlap) and dose measured by MapCHECK2 agreed well to those measured by the conventional methods of EPID and film (the differences ranged from -0.01 to 0 cm and from -1.34% to 0.6% for the gap and dose, respectively). No variation in the repeat measurements of the junction gap was found whereas the measurements of junction dose were found to vary in quite a small range over the days of measurement (0.21%-0.35%). While the sensitivity of the measured junction gap to the actual junction gap applied was the ideal value of 1 cm/cm as expected, the sensitivity of the junction dose to the actual junction gap increased as the junction gap (or overlap) decreased (maximum sensitivity: 201.7%/cm). Conclusions: The initial results suggest that the method is applicable for a comprehensive quality assurance of the asymmetric jaw alignment. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Developing new extension of GafChromic RTQA2 film to patient quality assurance field using a plan-based calibration method
    Peng, Jiayuan
    Zhang, Zhen
    Wang, Jiazhou
    Xie, Jiang
    Chen, Junchao
    Hu, Weigang
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (19): : 7533 - 7542
  • [42] Automatic MRI 2D brain segmentation using graph searching technique
    Pedoia, Valentina
    Binaghi, Elisabetta
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2013, 29 (09) : 887 - 904
  • [43] Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array
    Wagner, Daniela
    Vorwerk, Hilke
    RADIATION ONCOLOGY, 2011, 6
  • [44] 3D printing for dosimetric optimization and quality assurance in small animal irradiations using megavoltage X-rays
    Koutsouvelis, Nikolaos
    Rouzaud, Michel
    Dubouloz, Angele
    Nouet, Philippe
    Jaccard, Maud
    Garibotto, Valentina
    Tournier, Benjamin B.
    Zilli, Thomas
    Dipasquale, Giovanna
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2020, 30 (03): : 227 - 235
  • [45] A 2D strip ionization chamber array with high spatiotemporal resolution for proton pencil beam scanning FLASH radiotherapy
    Yang, Yunjie
    Shi, Chengyu
    Chen, Chin-Cheng
    Tsai, Pingfang
    Kang, Minglei
    Huang, Sheng
    Lin, Chih-Hsun
    Chang, Fu-Xiong
    Chhabra, Arpit M.
    Choi, J. Isabelle
    Tome, Wolfgang A.
    Simone, Charles B.
    Lin, Haibo
    MEDICAL PHYSICS, 2022, 49 (08) : 5464 - 5475
  • [47] Predictive gamma passing rate of 3D detector array-based volumetric modulated arc therapy quality assurance for prostate cancer via deep learning
    Matsuura, Takaaki
    Kawahara, Daisuke
    Saito, Akito
    Miura, Hideharu
    Yamada, Kiyoshi
    Ozawa, Shuichi
    Nagata, Yasushi
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (04) : 1073 - 1081
  • [48] Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques
    Newton, Joseph
    Oldham, Mark
    Thomas, Andrew
    Li, Yifan
    Adamovics, John
    Kirsch, David G.
    Das, Shiva
    MEDICAL PHYSICS, 2011, 38 (12) : 6754 - 6762
  • [49] Shape reconstruction of 2D perfectly conducting cylinder targets using the particle swarm optimisation
    Chiu, C-C
    Sun, C-H
    Fan, Y-S
    IMAGING SCIENCE JOURNAL, 2012, 60 (02): : 83 - 89
  • [50] The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
    Sim, G. S.
    Wong, J. H. D.
    Ng, K. H.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (04): : 85 - 95