Quality assurance of asymmetric jaw alignment using 2D diode array

被引:1
|
作者
Kim, Sun Mo [1 ,2 ]
Chmielewski, Renata [1 ]
Abbas, Ahmar [1 ]
Yeung, Ivan W. T. [1 ,2 ,3 ]
Moseley, Douglas J. [1 ,2 ,3 ]
机构
[1] Southlake Reg Hlth Ctr, Stronach Reg Canc Ctr, Dept Med Phys, Newmarket, ON L3Y 2P9, Canada
[2] Univ Hlth Network, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[3] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
关键词
asymmetric jaw alignment; junction gap; 2D diode array; MapCHECK2; RADIOCHROMIC FILM; ABUTTED FIELDS; DOSIMETRY; HEAD; COLLIMATORS; TIME; NECK;
D O I
10.1118/1.4828784
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A method using a 2D diode array is proposed to measure the junction gap (or overlap) and dose with high precision for routine quality assurance of the asymmetric jaw alignment. Methods: The central axis (CAX) of the radiation field was determined with a 15 x 15 cm(2) photon field at four cardinal collimator angles so that the junction gap (or overlap) can be measured with respect to the CAX. Two abutting fields having a field size of 15 cm (length along the axis parallel to the junction) x 7.5 cm (width along the axis perpendicular to the junction) were used to irradiate the 2D diode array (MapCHECK2) with 100 MU delivered at the photon energy of 6 MV. The collimator was slightly rotated at 15 degrees with respect to the beam central axis to increase the number of diodes effective on the measurement of junction gap. The junction gap and dose measured in high spatial resolution were compared to the conventional methods using an electronic portal imaging device (EPID) and radiochromic film, respectively. In addition, the reproducibility and sensitivity of the proposed method to the measurements of junction gap and dose were investigated. Results: The junction gap (or overlap) and dose measured by MapCHECK2 agreed well to those measured by the conventional methods of EPID and film (the differences ranged from -0.01 to 0 cm and from -1.34% to 0.6% for the gap and dose, respectively). No variation in the repeat measurements of the junction gap was found whereas the measurements of junction dose were found to vary in quite a small range over the days of measurement (0.21%-0.35%). While the sensitivity of the measured junction gap to the actual junction gap applied was the ideal value of 1 cm/cm as expected, the sensitivity of the junction dose to the actual junction gap increased as the junction gap (or overlap) decreased (maximum sensitivity: 201.7%/cm). Conclusions: The initial results suggest that the method is applicable for a comprehensive quality assurance of the asymmetric jaw alignment. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] An EPID based method for efficient and precise asymmetric jaw alignment quality assurance
    Clews, Luke
    Greer, Peter B.
    MEDICAL PHYSICS, 2009, 36 (12) : 5488 - 5496
  • [2] The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance
    Espinoza, A.
    Petasecca, M.
    Fuduli, I.
    Howie, A.
    Bucci, J.
    Corde, S.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2015, 42 (02) : 663 - 673
  • [3] Performance assessment of a 2D array of plastic scintillation detectors for IMRT quality assurance
    Guillot, Mathieu
    Gingras, Luc
    Archambault, Louis
    Beddar, Sam
    Beaulieu, Luc
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (13): : 4439 - 4454
  • [4] A 2D silicon detector array for quality assurance in small field dosimetry: DUO
    Al Shukaili, Khalsa
    Petasecca, Marco
    Newall, Matthew
    Espinoza, Anthony
    Perevertaylo, Vladimir L.
    Corde, Stephanie
    Lerch, Michael
    Rosenfeld, Anatoly B.
    MEDICAL PHYSICS, 2017, 44 (02) : 628 - 636
  • [5] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Diaz Moreno, Rogelio Manuel
    Venencia, Daniel
    Garrigo, Edgardo
    Pipman, Yakov
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 145 - 151
  • [6] Evaluation of a high resolution diode array for CyberKnife quality assurance
    Garcia-Fuentes, Juan D. D.
    Sevillano, David
    Colmenares, Rafael
    Capuz, Ana B. B.
    Moris, Rafael
    Camara, Miguel
    Galiano, Pablo
    Williamson, Sandra
    Bejar, Maria J.
    Prieto, Daniel
    Garcia-Vicente, Feliciano
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09):
  • [7] Quality assurance of carbon ion and proton beams: A feasibility study for using the 2D MatriXX detector
    Anvar, M. Varasteh
    Attili, A.
    Ciocca, M.
    Donetti, M.
    Guarachi, L. K. Fanola
    Fausti, F.
    Giordanengo, S.
    Marchetto, F.
    Molinelli, S.
    Monaco, V.
    Sacchi, R.
    Vignati, A.
    Cirio, R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (06): : 831 - 837
  • [8] Determination of correction factors for a 2D diode array device in MV photon beams
    Zhou, Xiaodong
    Both, Stefan
    Zhu, Timothy C.
    MEDICAL PHYSICS, 2009, 36 (02) : 523 - 529
  • [9] Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate
    Alrowaili, Ziyad A.
    Lerch, Michael L. F.
    Petasecca, Marco
    Carolan, Martin G.
    Metcalfe, Peter E.
    Rosenfeld, Anatoly B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 85 - 98
  • [10] Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT
    Korevaar, Erik W.
    Wauben, David J. L.
    van der Hulst, Peter C.
    Langendijk, Johannes A.
    Van't Veld, Aart A.
    RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) : 446 - 452