Model-Based Personalization Scheme of an Artificial Pancreas for Type 1 Diabetes Applications

被引:0
作者
Lee, Joon Bok [1 ]
Dassau, Eyal [1 ]
Seborg, Dale E. [1 ]
Doyle, Francis J., III [1 ]
机构
[1] Univ Calif Santa Barbara, Dept Chem Engn, Santa Barbara, CA 93106 USA
来源
2013 AMERICAN CONTROL CONFERENCE (ACC) | 2013年
关键词
AP; control-relevant modeling; closed-loop; internal model control (IMC); proportional-integral-derivative (PID) control; insulin feedback; type 1 diabetes mellitus (T1DM); INSULIN DELIVERY; ALGORITHM; MELLITUS;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Automated controllers designed to regulate blood glucose concentrations in people with Type 1 diabetes mellitus (T1DM) must avoid hypoglycemia (blood glucose < 70 mg/dl) while minimizing hyperglycemia (> 180 mg/dl), a challenging task. In this paper, a model-based control design approach with a personalized scheme based on readily available clinical factors is applied to a linearized control-relevant model of subject insulin-glucose response profiles. An insulin feedback strategy is included with specific personalization settings and variations in a tuning parameter, tau(c). The control strategy is challenged by an unannounced meal disturbance with 50g carbohydrate content. A set of metrics are introduced as a method of evaluating the performance of different controllers. In-silico simulations of ten subjects in the Food and Drug Administration accepted Universities of Virginia and Padova metabolic simulator indicate that the personalization strategy with a tau(c) setting of 270 minutes gives very good controller performance. Post-prandial glucose concentration peaks of 183 mg/dl were achieved with 97% of the total simulation time spent within a safe glycemic zone (70-180 mg/dl), without hypoglycemic incidents and without requiring a time-consuming model identification process.
引用
收藏
页码:2911 / 2916
页数:6
相关论文
共 24 条
[21]   The Effect of Insulin Feedback on Closed Loop Glucose Control [J].
Steil, Garry M. ;
Palerm, Cesar C. ;
Kurtz, Natalie ;
Voskanyan, Gayane ;
Roy, Anirban ;
Paz, Sachiko ;
Kandeel, Fouad R. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2011, 96 (05) :1402-1408
[22]   Control-Relevant Models for Glucose Control Using A Priori Patient Characteristics [J].
van Heusden, Klaske ;
Dassau, Eyal ;
Zisser, Howard C. ;
Seborg, Dale E. ;
Doyle, Francis J., III .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (07) :1839-1849
[23]  
Walsh J., 2006, Pumping insulin, Vfourth
[24]   Fully automated closed-loop insulin delivery versus sermautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas [J].
Weinzimer, Stuart A. ;
Steil, Garry M. ;
Swan, Karena L. ;
Dziura, Jim ;
Kurtz, Natalie ;
Tamborlane, Williani V. .
DIABETES CARE, 2008, 31 (05) :934-939