A differential dielectric affinity glucose sensor

被引:20
作者
Huang, Xian [1 ]
Leduc, Charles [2 ]
Ravussin, Yann [2 ]
Li, Siqi [3 ]
Davis, Erin [3 ]
Song, Bing [1 ]
Li, Dachao [4 ]
Xu, Kexin [4 ]
Accili, Domenico [5 ]
Wang, Qian [3 ]
Leibel, Rudolph [2 ]
Lin, Qiao [1 ]
机构
[1] Columbia Univ, Dept Mech Engn, New York, NY 10027 USA
[2] Columbia Univ, Dept Pediat, Med Ctr, New York, NY 10032 USA
[3] Univ S Carolina, Dept Chem & Biochem, Columbia, SC 29208 USA
[4] Tianjin Univ, Coll Precis Instrument & Optoelect Engn, Tianjin 300072, Peoples R China
[5] Columbia Univ, Med Ctr, Dept Med, New York, NY 10032 USA
基金
美国国家卫生研究院;
关键词
IMPEDANCE SPECTROSCOPY; BIOCOMPATIBILITY; POLYMER; VISCOSITY;
D O I
10.1039/c3lc51026c
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A continuous glucose monitor with a differential dielectric sensor implanted within the subcutaneous tissue that determines the glucose concentration in the interstitial fluid is presented. The device, created using microelectromechanical systems (MEMS) technology, consists of sensing and reference modules that are identical in design and placed in close proximity. Each module contains a microchamber housing a pair of capacitive electrodes residing on the device substrate and embedded in a suspended, perforated polymer diaphragm. The microchambers, enclosed in semi-permeable membranes, are filled with either a polymer solution that has specific affinity to glucose or a glucose-insensitive reference solution. To accurately determine the glucose concentration, changes in the permittivity of the sensing and the reference solutions induced by changes in glucose concentration are measured differentially. In vitro characterization demonstrated the sensor was capable of measuring glucose concentrations from 0 to 500 mg dL(-1) with resolution and accuracy of similar to 1.7 mu g dL(-1) and similar to 1.74 mg dL(-1), respectively. In addition, device drift was reduced to 1.4% (uncontrolled environment) and 11% (5 degrees C of temperature variation) of that from non-differential measurements, indicating significant stability improvements. Preliminary animal testing demonstrated that the differential sensor accurately tracks glucose concentration in blood. This sensor can potentially be used clinically as a subcutaneously implanted continuous monitoring device in diabetic patients.
引用
收藏
页码:294 / 301
页数:8
相关论文
共 47 条
[1]   A membrane-moderated, conductimetric sensor for the detection and measurement of specific organic solutes in aqueous solutions [J].
Arnold, FH ;
Zheng, WG ;
Michaels, AS .
JOURNAL OF MEMBRANE SCIENCE, 2000, 167 (02) :227-239
[2]   KINETICS AND ENERGETICS OF CALCIUM EFFLUX FROM INTACT SQUID GIANT-AXONS [J].
BAKER, PF ;
MCNAUGHTON, PA .
JOURNAL OF PHYSIOLOGY-LONDON, 1976, 259 (01) :103-144
[3]  
Ballerstadt Ralph, 2007, J Diabetes Sci Technol, V1, P384
[4]   Detection of protein-DNA interaction with a DNA probe: distinction between single-strand and double-strand DNA-protein interaction [J].
Ban, CG ;
Chung, SM ;
Park, DS ;
Shim, YB .
NUCLEIC ACIDS RESEARCH, 2004, 32 (13) :e110
[5]  
Barriet D., 2009, Journal of Medical Devices, V3, DOI 10.1115/1.3136753
[6]   DIELECTRIC RELAXATION IN POLYMER SOLUTIONS [J].
BLOCK, H ;
NORTH, AM .
ADVANCES IN MOLECULAR RELAXATION PROCESSES, 1970, 1 (04) :309-&
[7]   A viscosity-dependent affinity sensor for continuous monitoring of glucose in biological fluids [J].
Boss, Christophe ;
Meurville, Eric ;
Sallese, Jean-Michel ;
Ryser, Peter .
BIOSENSORS & BIOELECTRONICS, 2011, 30 (01) :223-228
[8]   A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring [J].
Cengiz, Eda ;
Tamborlane, William V. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 :S11-S16
[9]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[10]   Label-free impedance biosensors: Opportunities and challenges [J].
Daniels, Jonathan S. ;
Pourmand, Nader .
ELECTROANALYSIS, 2007, 19 (12) :1239-1257