Effects of enamel and dentine thickness on laser doppler blood-flow signals recorded from the underlying pulp cavity in human teeth in vitro

被引:8
作者
Banthitkhunanon, P. [1 ]
Chintakanan, S. [2 ]
Wanachantararak, S. [2 ]
Vongsavan, N. [3 ]
Matthews, B. [4 ]
机构
[1] Chiang Mai Univ, Fac Dent, Dept Prosthodont, Chiang Mai 50202, Thailand
[2] Chiang Mai Univ, Dept Oral Biol & Oral Diagnost Sci, Fac Dent, Chiang Mai 50202, Thailand
[3] Mahidol Univ, Dept Physiol & Biochem, Fac Dent, Bangkok 10400, Thailand
[4] Univ Bristol, Dept Physiol & Pharmacol, Sch Med Sci, Bristol BS8 1TH, Avon, England
关键词
Laser Doppler flow metre; Pulpal blood flow; Pulpal perfusion;
D O I
10.1016/j.archoralbio.2013.08.007
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To determine the effect of enamel and dentine thickness on laser Doppler blood-flow (LDF) signals recorded from dental pulp. Design: Observations were made on 18 human premolars that had been extracted from young patients during orthodontic treatment. The apical 2/3 of the root was cut off and the remaining pulp removed. Blood flow signals were recorded from the buccal surface of the crown with a laser Doppler flow metre while dilute blood was pumped at 10 ml/min. through a cannula inserted into the pulp cavity. Recordings were made from the enamel surface and at 0.5 mm steps through the enamel and dentine. Results: The blood flow signal increased significantly as the cavity depth increased and at 2.0 mm, the median flux signal was more than ten times greater than that obtained on the enamel surface. The backscattered light intensity did not change with cavity depth. Conclusion: When recording pulpal blood flow from a human tooth with a laser Doppler flow metre, a substantially better signal-to-noise ratio should be obtained by placing the probe on dentine in the floor of a cavity than on the enamel surface. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1692 / 1695
页数:4
相关论文
共 6 条
[1]   Effect of Gingiva on laser Doppler pulpal blood flow measurements [J].
Akpinar, KE ;
Er, K ;
Polat, S ;
Polat, NT .
JOURNAL OF ENDODONTICS, 2004, 30 (03) :138-140
[2]   Laser Doppler flowmetry in endodontics: a review [J].
Jafarzadeh, H. .
INTERNATIONAL ENDODONTIC JOURNAL, 2009, 42 (06) :476-490
[3]   Magnification - a new look at a long-known optical property of dentin [J].
Kienle, A. ;
Michels, R. ;
Hibst, R. .
JOURNAL OF DENTAL RESEARCH, 2006, 85 (10) :955-959
[4]   A comparison between red and infrared light for recording pulpal blood flow from human anterior teeth with a laser Doppler flow meter [J].
Kijsamanmith, Kanittha ;
Timpawat, Siriporn ;
Vongsavan, Noppakun ;
Matthews, Bruce .
ARCHIVES OF ORAL BIOLOGY, 2011, 56 (06) :614-618
[5]   The sources of laser Doppler blood-flow signals recorded from human teeth [J].
Soo-ampon, S ;
Vongsavan, N ;
Soo-ampon, M ;
Chuckpaiwong, S ;
Matthews, B .
ARCHIVES OF ORAL BIOLOGY, 2003, 48 (05) :353-360
[6]  
VONGSAVAN N, 1993, EXP PHYSIOL, V78, P1