Evaluation with in vivo optical coherence tomography and histology of the vascular effects of the everolimus-eluting bioresorbable vascular scaffold at two years following implantation in a healthy porcine coronary artery model: implications of pilot results for future pre-clinical studies

被引:18
作者
Gogas, Bill D. [1 ]
Radu, Maria [1 ]
Onuma, Yoshinobu [1 ]
Perkins, Laura [2 ]
Powers, Jennifer C. [2 ]
Gomez-Lara, Josep [1 ]
Farooq, Vasim [1 ]
Garcia-Garcia, Hector M. [1 ]
Diletti, Roberto [1 ]
Rapoza, Richard [2 ]
Virmani, Renu [3 ]
Serruys, Patrick W. [1 ]
机构
[1] Erasmus Univ, Med Ctr, Thoraxctr, NL-3015 Rotterdam, Netherlands
[2] Abbott Vasc, Santa Clara, CA USA
[3] CVPath, Gaithersburg, MA USA
关键词
Bioresorbable vascular scaffold; Vascular healing; Optical coherence tomography; Histology; INTRAVASCULAR ULTRASOUND; STENT IMPLANTATION; ABSORB TRIAL; REPRODUCIBILITY; SYSTEM;
D O I
10.1007/s10554-011-9860-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To quantify with in vivo OCT and histology, the device/vessel interaction after implantation of the bioresorbable vascular scaffold (BVS). We evaluated the area and thickness of the strut voids previously occupied by the polymeric struts, and the neointimal hyperplasia (NIH) area covering the endoluminal surface of the strut voids (NIHEV), as well as the NIH area occupying the space between the strut voids (NIHBV), in healthy porcine coronary arteries at 2, 3 and 4 years after implantation of the device. Twenty-two polymeric BVS were implanted in the coronary arteries of 11 healthy Yucatan minipigs that underwent OCT at 2, 3 and 4 years after implantation, immediately followed by euthanasia. The areas and thicknesses of 60 corresponding strut voids previously occupied by the polymeric struts and the size of 60 corresponding NIHEV and 49 NIHBV were evaluated with both OCT and histology by 2 independent observers, using a single quantitative analysis software for both techniques. At 3 and 4 years after implantation, the strut voids were no longer detectable by OCT or histology due to complete polymer resorption. However, analysis performed at 2 years still provided clear delineation of these structures, by both techniques. The median [ranges] areas of these strut voids were 0.04 [0.03-0.16] and 0.02 [0.01-0.07] mm(2) by histology and OCT, respectively. The mean (+/- SD) thickness by histology and OCT was 220 +/- A 40 and 120 +/- A 20 mu m, respectively. The median [ranges] NIHEV by histology and OCT was 0.07 [0.04-0.20] and 0.03 [0.01-0.08] mm(2), while the mean (+/- SD) NIHBV by histology and OCT was 0.13 +/- A 0.07 and 0.10 +/- A 0.06 mm(2). Our study indicates that in vivo OCT of the BVS provides correlated measurements of the same order of magnitude as histomorphometry, and is reproducible for the evaluation of certain vascular and device-related characteristics. However, histology systematically gives larger values for all the measured structures compared to OCT, at 2 years post implantation.
引用
收藏
页码:499 / 511
页数:13
相关论文
共 20 条
[1]   VOLUME CHANGES OF TISSUES IN PHYSIOLOGICAL FLUIDS DURING FIXATION IN OSMIUM TETROXIDE OR FORMALDEHYDE AND DURING SUBSEQUENT TREATMENT [J].
BAHR, GF ;
BLOOM, G ;
FRIBERG, U .
EXPERIMENTAL CELL RESEARCH, 1957, 12 (02) :342-355
[2]  
Bruining Nico, 2009, Circ Cardiovasc Imaging, V2, pe8, DOI 10.1161/CIRCIMAGING.108.800862
[3]   Extracellular matrix changes in stented human coronary arteries [J].
Farb, A ;
Kolodgie, FD ;
Hwang, JY ;
Burke, AP ;
Tefera, K ;
Weber, DK ;
Wight, TN ;
Virmani, R .
CIRCULATION, 2004, 110 (08) :940-947
[4]   Reproducibility of quantitative optical coherence tomography for stent analysis [J].
Gonzalo, Nieves ;
Garcia-Garcia, Hector M. ;
Serruys, Patrick W. ;
Commissaris, Koen H. ;
Bezerra, Hiram ;
Gobbens, Pierre ;
Costa, Marco ;
Regar, Evelyn .
EUROINTERVENTION, 2009, 5 (02) :224-232
[5]   Quantitative Ex Vivo and In Vivo Comparison of Lumen Dimensions Measured by Optical Coherence Tomography and Intravascular Ultrasound in Human Coronary Arteries [J].
Gonzalo, Nieves ;
Serruys, Patrick W. ;
Garcia-Garcia, Hector M. ;
van Soest, Gijs ;
Okamura, Takayuki ;
Ligthart, Jurgen ;
Knaapen, Michiel ;
Verheye, Stefan ;
Bruining, Nico ;
Regar, Evelyn .
REVISTA ESPANOLA DE CARDIOLOGIA, 2009, 62 (06) :615-624
[6]   Comparison of nonuniform rotational distortion between mechanical IVUS and OCT using a phantom model [J].
Kawase, Yoshiaki ;
Suzuki, Yoriyasu ;
Ikeno, Fumiaki ;
Yoneyama, Ryuichi ;
Hoshino, Kozo ;
Ly, Hung Q. ;
Lau, George T. ;
Hayase, Motoya ;
Yeung, Alan C. ;
Hajjar, Roger J. ;
Jang, Ik-Kyung .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2007, 33 (01) :67-73
[7]   Accuracy of Optical Coherence Tomography in the Evaluation of Neointimal Coverage After Stent Implantation [J].
Murata, Akira ;
Wallace-Bradley, David ;
Tellez, Armando ;
Alviar, Carlos ;
Aboodi, Michael ;
Sheehy, Alexander ;
Coleman, Leslie ;
Perkins, Laura ;
Nakazawa, Gaku ;
Mintz, Gary ;
Kaluza, Greg L. ;
Virmani, Renu ;
Granada, Juan F. .
JACC-CARDIOVASCULAR IMAGING, 2010, 3 (01) :76-84
[8]  
Okamura T, HIGH SPEED INT UNPUB
[9]   Reproducibility of coronary Fourier domain optical coherence tomography: quantitative analysis of in vivo stented coronary arteries using three different software packages [J].
Okamura, Takayuki ;
Gonzalo, Nieves ;
Gutierrez-Chico, Juan Luis ;
Serruys, Patrick W. ;
Bruining, Nico ;
de Winter, Sebastiaan ;
Dijkstra, Jouke ;
Commossaris, Koen H. ;
Van Geuns, Robert-Jan ;
Van Soest, Gijs ;
Ligthart, Jurgen ;
Regar, Evelyn .
EUROINTERVENTION, 2010, 6 (03) :371-379
[10]  
Okamura T, 2010, EUROINTERVENTION, V5, P932, DOI 10.4244/