Systematic Comparisons of Electrocardiographic Morphology Increase the Precision of QT Interval Measurement

被引:36
作者
Hnatkova, Katerina [1 ]
Smetana, Peter [2 ]
Toman, Ondrej [3 ]
Bauer, Axel [4 ]
Schmidt, Georg [4 ]
Malik, Marek [1 ]
机构
[1] St Pauls Cardiac Electrophysiol, Surrey CR8 3NQ, England
[2] Wilhelminenspital Stadt Wien, Vienna, Austria
[3] Masaryk Univ, Univ Hosp Brno, Brno, Czech Republic
[4] Tech Univ Munich, Med Klin 1, Munich, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2009年 / 32卷 / 01期
关键词
thorough QT; QTc study; QTc variability; ECG morphology; pattern matching; HEART-RATE CORRECTION; HEALTHY-SUBJECTS; NOISE;
D O I
10.1111/j.1540-8159.2009.02185.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Decreased intrasubject variability of QTc values is needed to increase the power and reduce the size of the so-called thorough QT studies. One source of QTc variability is the lack of systematic measurements when electrocardiograms (ECG) with closely matching morphologies are not measured in an exactly corresponding way. The inaccuracy can be eliminated by postprocessing of QT measurements by ECG pattern matching. This study tested the effects of pattern matching in ECG measurements in two populations of healthy subjects (n = 48 + 56) and in a population of patients with advanced Parkinson's disease (n = 130) in whom both day-time and night-time data were available. Intrasubject QTc variability was measured by intrasubject standard deviations (SD) of QTc values obtained with manual measurements before and after pattern-matching measurement alignments. In each subject, QT values (n = 230-320) in one drug-free long-term ECG recording were evaluated. The pattern-matching adjustment of the QT measurement decreased the intrasubject QTc variability from 5.2 +/- 1.0 to 4.5 +/- 1.0 ms (P < 10(-14)) from 6.4 +/- 1.7 to 5.5 +/- 1.6 ms (P < 10(-10)) from 5.6 +/- 1.5 to 4.6 +/- 1.4 ms (P < 10(-34)) and from 6.1 +/- 1.9 to 5.0 +/- 1.7 ms (P < 10(-33)), in the two populations of healthy subjects and in the day-time and night-time recordings of Parkinson's disease patients, respectively. Hence, morphological pattern adjustment of QT interval measurements improves the quality of the QT data with substantial practical implications. Reductions in intrasubject QTc variability were reproducibly found in different populations and thus the technology might be recommended for every thorough QT/QTc study. Noticeable reductions of necessary study size are likely achievable in this way. (PACE 2009; 32:119-130).
引用
收藏
页码:119 / 130
页数:12
相关论文
共 19 条
[1]   Assessment of noise in digital electrocardiograms [J].
Batchvarov, V ;
Hnatkova, K ;
Malik, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (04) :499-503
[2]   QT-RR relationship in healthy subjects exhibits substantial intersubject variability and high intrasubject stability [J].
Batchvarov, VN ;
Ghuran, A ;
Smetana, P ;
Hnatkova, K ;
Harries, M ;
Dilaveris, P ;
Camm, AJ ;
Malik, M .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2002, 282 (06) :H2356-H2363
[3]   Developments in ECG acquisition, preprocessing, parameter measurement, and recording [J].
Daskalov, IK ;
Dotsinsky, IA ;
Christov, II .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1998, 17 (02) :50-58
[4]   Circadian modulation of QT rate dependence in healthy volunteers - Gender and age differences [J].
Extramiana, F ;
Maison-Blanche, P ;
Badilini, F ;
Pinoteau, J ;
Deseo, T ;
Coumel, P .
JOURNAL OF ELECTROCARDIOLOGY, 1999, 32 (01) :33-43
[5]  
Food and Drug Administration International Conference on Harmonisation, 2005, FED REG, V70, P61134
[6]  
Fridericia LS., 1920, Acta Med Scand, V53, P469, DOI [10.1111/j.0954-6820.1920.tb18267.x, DOI 10.1111/J.0954-6820.1920.TB18266.X, 10.1111/j.0954-6820.1920.tb18266.x]
[7]   Precision of QT interval measurement by advanced electrocardiographic equipment [J].
Hnatkova, Katerina ;
Gang, Yi ;
Batchvarov, Velislav N. ;
Malik, Marek .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (11) :1277-1284
[8]   Levocetirizine does not prolong the QT/QTc interval in healthy subjects: results from a thorough QT study [J].
Hulhoven, Reginald ;
Rosillon, Dominique ;
Letiexhe, Michel ;
Meeus, Marie-Anne ;
Daoust, Agnes ;
Stockis, Armel .
EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 2007, 63 (11) :1011-1017
[9]   Randomized, double-blind, crossover study to investigate the effect of rivaroxaban on QT-interval prolongation [J].
Kubitza, Dagmar ;
Mueck, Wolfgang ;
Becka, Michael .
DRUG SAFETY, 2008, 31 (01) :67-77
[10]  
Malik M, 2007, EUR HEART J, V28, P624