Accuracy of Image-Guided Surgical Systems at the Lateral Skull Base as Clinically Assessed Using Bone-Anchored Hearing Aid Posts as Surgical Targets

被引:21
作者
Balachandran, Ramya [1 ]
Fitzpatrick, J. Michael [2 ]
Labadie, Robert F. [1 ]
机构
[1] Vanderbilt Univ, Med Ctr, Dept Otolaryngol Head & Neck Surg, Nashville, TN 37203 USA
[2] Vanderbilt Univ, Dept Elect Engn & Comp Sci, Nashville, TN USA
关键词
Accuracy; BAHA; BrainLab; EarMark; Fiducial system; Image-guided surgical system; Surgical target; Target registration error;
D O I
10.1097/MAO.0b013e3181859a08
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: Image-guided surgical (IGS) technology has been clinically available for more than a decade. To date, no acceptable standard exists for reporting the accuracy of IGS systems, especially for lateral skull base applications. We present a validation method that uses the post from bone-anchored hearing aid (BAHA) patients as a target. We then compare the accuracy of 2 IGS systems-one using laser skin-surface scanning (LSSS) and another using a noninvasive fiducial frame (FF) attached to patient via dental bite-block (DBB) for registration. Study Design: Prospective. Setting: Tertiary referral center. Patients: Six BAHA patients who had adequate dentition for creation of a DBB. Intervention(s): For each patient, a dental impression was obtained, and a customized DBB was made to hold an FF. Temporal bone computed tomographic (CT) scans were obtained with the patient wearing the DBB, FF, and a customized marker on the BAHA post. In a mock operating room, CT scans were registered to operative anatomy using 2 methods: 1) LSSS and 2) FF. Main Outcome Measure(s): For each patient and each system, the position of the BAHA marker in the CT scan and in the mock operating room (using optical measurement technology) was compared, and the distances between them are reported to demonstrate accuracy. Results: Accuracy (mean +/- standard deviation) was 1.54 +/- 0.63 mm for DBB registration and 3.21 +/- 1.02 mm for LSSS registration. Conclusion: AD IGS system using FF registration is more accurate than one using LSSS (p = 0.03, 2-sided Student's t test). BAHA patients provide a unique patient population upon which IGS systems may be validated.
引用
收藏
页码:1050 / 1055
页数:6
相关论文
共 21 条
[1]  
BALACHANDRAN R, 2007, SPIE MED IMAGING, V6509
[2]  
Balachandran R, 2006, I S BIOMED IMAGING, P518
[3]   Efficacy of computed tomographic image-guided endoscopic sinus surgery in residency training programs [J].
Casiano, RR ;
Numa, WA .
LARYNGOSCOPE, 2000, 110 (08) :1277-1282
[4]  
Cleary Kevin, 2005, J Laparoendosc Adv Surg Tech A, V15, P497
[5]   Locking acrylic resin dental stent for image-guided surgery [J].
Fenlon, MR ;
Jusczyzck, AS ;
Edwards, PJ ;
King, AP .
JOURNAL OF PROSTHETIC DENTISTRY, 2000, 83 (04) :482-485
[6]  
Fitzpatrick J.M., 2000, Handbook of Medical Imaging, VII, P447, DOI DOI 10.1117/3.831079.CH8
[7]   The distribution of target registration error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (09) :917-927
[8]   Image-guided endoscopic surgery: Results of accuracy and performance in a multicenter clinical study using an electromagnetic tracking system [J].
Fried, MP ;
Kleefield, J ;
Gopal, H ;
Reardon, E ;
Ho, BT ;
Kuhn, FA .
LARYNGOSCOPE, 1997, 107 (05) :594-601
[9]  
Hajnal J. V., 2001, Medical Image Registration
[10]   In vitro assessment of image-guided otologic surgery: Submillimeter accuracy within the region of the temporal bone [J].
Labadie, RF ;
Shah, RJ ;
Harris, SS ;
Cetinkaya, E ;
Haynes, DS ;
Fenlon, MR ;
Juszczyk, AS ;
Galloway, RL ;
Fitzpatrick, JM .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2005, 132 (03) :435-442