Using diagnostic reference levels to evaluate the improvement of patient dose optimisation and the influence of recent technologies in radiography and computed tomography

被引:51
作者
Roch, Patrice [1 ]
Celier, David [1 ]
Dessaud, Cecile [2 ]
Etard, Cecile [1 ]
机构
[1] Inst Radioprotect & Surete Nucl, 31 Ave Div Leclerc, F-92260 Fontenay Aux Roses, France
[2] CERAP, Domaine Technol Saclay, 4 Rue Rene Razel, F-91400 Saclay, France
关键词
Patient dose; Radiography; CT; Optimisation; Diagnostic reference levels;
D O I
10.1016/j.ejrad.2017.11.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Twelve years since the implementation of Diagnostic Reference Levels (DRL) process in France, the Nuclear Safety and Radiation Protection French Institute (IRSN) presents its latest analyses performed on the most recent national data. Methods: Statutorily, each year, medical imaging departments must perform patient exposure evaluation from their clinical practice for at least 2 types of radiographic and computed tomography (CT) examinations freely chosen in the regulatory list. The samples of dosimetric data used for the evaluations must be sent to IRSN for national assessment using a dedicated and secured web portal. The analyses of collected data for radiography and CT allow IRSN to estimate the representativeness of current DRLs in terms of target practices and examinations, dosimetric quantities and numerical values. Technical data are transmitted, such as detector type in radiography or commissioning date of CT, and are included in some complementary analyses in order to evaluate their influence on patient exposure. Results: Since 2004 the involvement of professionals in the DRL process has highly increased in CT (about 80% in 2015) but remains quite weak in radiography (almost 30%). Analyses show some discordance between regulation references and clinical practice leading to clinical doses data which are 40% lower than DRLs in 2015. As a consequence, the list of examinations types and some numerical values should be updated in the regulation. Focused analyses show a significant patient exposure reduction when digital radiography is used and when CT equipment is under five years old. Conclusions: Based on these findings, IRSN recommends to update DRL regulation with current and relevant examination lists, dosimetric quantities and numerical values. In addition, this study shows that technology and generation of equipment, such as detector type in radiography or image reconstruction algorithm in CT, take an important place in the dose optimisation process, enabling significant patient exposure reduction when it is associated with protocols optimisation.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 27 条
[1]   Optimization of dose and image quality for computed radiography and digital radiography [J].
Aldrich, JE ;
Duran, E ;
Dunlop, P ;
Mayo, JR .
JOURNAL OF DIGITAL IMAGING, 2006, 19 (02) :126-131
[2]  
[Anonymous], 1996, Ann ICRP, V26, P1
[3]  
[Anonymous], 2016, DIAGNOSTISCHE REFERE
[4]  
[Anonymous], 2011, R0604 OFSP
[5]  
[Anonymous], J OFFICIEL REPUBLIQU
[6]  
[Anonymous], 1997, Official Journal of the European Union
[7]  
[Anonymous], DIAGN REF LEV MED IM, p201X
[8]  
[Anonymous], 2010, R0606 OFSP
[9]  
[Anonymous], 2012, J OFFICIEL REPUBLIQU
[10]  
[Anonymous], 2011, Journal Officiel de la Republique Algerienne