An assessment of the luminance and light field characteristics of used direct laryngoscopes

被引:19
作者
Crosby, E
Cleland, M
机构
[1] Ottawa Hosp, Ottawa, ON K1H 8L6, Canada
[2] Univ Ottawa, Dept Anesthesiol, Ottawa, ON, Canada
[3] Univ Ottawa, Dept Biomed Engn, Ottawa, ON, Canada
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1999年 / 46卷 / 08期
关键词
Light Field; Direct Laryngoscope; Colour Temperature; Test Fixture; Luminous Flux;
D O I
10.1007/BF03013918
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Purpose: To determine luminance and light field characteristics and the effect of residual battery potential and luminance on light colour temperature in our used Macintosh #3 and #4 fibre-light (FLB) and bulb-light (BLB) laryngoscopes. Methods: We used a power supply to provide laryngoscopes with potentials equivalent to those most commonly measured in the handles in use in our OR. Measurements were made under controlled, constant, conditions using a Pentax digital spotmeter (luminance) or a Minolta Color III colour temperature meter (light colour), Colour measurements were made while increasing the power source potential from 2-3 volts (v) in increments of 0.1 v. Light field measurements were made with a mm increment ruler mounted on the base of the test fixture. Results: At 2.5 and 2.8 v respectively, the #3 FLB produced luminance values of 23.9 +/- 11.4 and 41.7 +/- 17.2 cd.m(-2) (mean +/- SD), and the #4 FLB produced 58.6 +/- 21.4 and 90.9 +/- 32.2 cd.m(-2). Increasing potential increased luminance values (P < 0.001) and the #4FLB produced higher luminance values (P < 0.001). BLB produced higher luminance values than did FLB across all comparisons (P < 0.001). As potentials and luminance values decreased, light temperature was reduced (P < 0.001). There were no differences in light field dimensions noted in any comparison. Conclusion: Fifteen percent of the BLB did not meet the minimum luminance for laryngoscopy of 100 cd.m(-2) 92% of the FLB did not meet that same standard.
引用
收藏
页码:792 / 796
页数:5
相关论文
共 1 条
[1]   A study of the brightness of laryngoscope light [J].
Skilton, RWH ;
Parry, D ;
Arthurs, GJ ;
Hiles, P .
ANAESTHESIA, 1996, 51 (07) :667-672