Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT

被引:11
|
作者
Taherparvar, Payvand [1 ]
Sadremomtaz, Alireza [1 ]
机构
[1] Univ Guilan, Dept Phys, Fac Sci, Namjoo Ave,POB 41635-1914, Rasht 4193833697, Gilan, Iran
关键词
GATE; Pixelated gamma camera; Spatial resolution; PET; VALIDATION; MODEL;
D O I
10.1007/s13246-017-0607-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
GATE is currently considered in scintigraphic imaging as a powerful tool to develop, design and optimize nuclear medicine modalities. This paper describes the GATE simulation of a pixelated gamma camera which is dedicated to high resolution of small animals imaging. It consists of a CsI(Na) crystal array coupled to position sensitive photomultiplier tube. The simulation model includes photon tracking through low energy high resolution hexagonal parallel holes collimator, CsI(Na) pixelated crystal, back-compartment, and camera shielding. Simulations were compared with experimental results by some criteria such as energy spectrum, energy resolution, spatial resolution, sensitivity and count profiles obtained from line and point sources imaging. The acquired energy resolution show good agreement with measured spectra. Difference between calculated and experimental values is about 0.3% for absolute sensitivity measurement. The result of the image uniformity is more consistent after implementation of non-uniformity correction. These values were about 1.3 and 1.2% for experimental and simulation study in the central field of view, respectively. Measurements showed that the spatial resolutions differences at the head surface along the long dimensions of gamma camera for simulation and experimental differed by no more than 4%.Differences along the short axis were about 6%. The FWHMs of images of point and line sources show good consistency between experimental images and corresponding simulated ones. The difference between experimental and simulated system parameters was within 11%. Our results demonstrate the ability and flexibility of the Monte Carlo simulation for modeling pixelated gamma camera with position sensitive detector by selecting the appropriate parameters for digitizer chain and collimator position on the detector surface.
引用
收藏
页码:31 / 39
页数:9
相关论文
共 50 条
  • [41] Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study
    Park, Chan Rok
    Kang, Seong-Hyeon
    Lee, Youngjin
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2021, 53 (06) : 1947 - 1954
  • [42] Optimization of pinhole collimator for small animal SPECT using Monte Carlo simulation
    Song, TY
    Choi, Y
    Chung, YH
    Jung, JH
    Choe, YS
    Lee, KH
    Kim, SE
    Kim, BT
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (03) : 327 - 332
  • [43] Feasibility of a CdTe-based SPECT for high-resolution low-dose small animal imaging: a Monte Carlo simulation study
    Park, S. -J.
    Yu, A. R.
    Lee, Y. -J.
    Kim, Y. -S.
    Kim, H. -J.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [44] Ultra-high-resolution SPECT with CdTe for small-animal imaging applications: A Monte Carlo simulation study using a voxelized phantom
    Su-Jin Park
    Chang-Lae Lee
    Hyo-Min Cho
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2012, 60 : 1145 - 1149
  • [45] Ultra-high-resolution SPECT with CdTe for Small-animal Imaging Applications: A Monte Carlo Simulation Study Using A Voxelized Phantom
    Park, Su-Jin
    Lee, Chang-Lae
    Cho, Hyo-Min
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 60 (07) : 1145 - 1149
  • [46] Evaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study
    Sadremomtaz, Alireza
    Telikani, Zeinab
    IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2016, 24 (02): : 136 - 143
  • [47] Monte Carlo Simulation of Intrinsic Count Rate Performance for Scintillation Gamma Camera
    Mowlavi, A. A.
    de Denaro, M.
    Fornasier, M. R.
    Binesh, A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 209 - 211
  • [48] Monte carlo simulation study of a novel parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Lee, Young-Jin
    Kim, Dae-Hong
    Rhee, Yong-Chun
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (11) : 1737 - 1744
  • [49] Monte carlo simulation study of a novel parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Young-Jin Lee
    Dae-Hong Kim
    Yong-Chun Rhee
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2014, 64 : 1737 - 1744
  • [50] A clinical gamma camera-based pinhole collimated system for high resolution small animal SPECT imaging
    Mejia, J.
    Galvis-Alonso, O. Y.
    de Castro, A. A.
    Braga, J.
    Leite, J. P.
    Simoes, M. V.
    BRAZILIAN JOURNAL OF MEDICAL AND BIOLOGICAL RESEARCH, 2010, 43 (12) : 1160 - 1166