Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT

被引:11
|
作者
Taherparvar, Payvand [1 ]
Sadremomtaz, Alireza [1 ]
机构
[1] Univ Guilan, Dept Phys, Fac Sci, Namjoo Ave,POB 41635-1914, Rasht 4193833697, Gilan, Iran
关键词
GATE; Pixelated gamma camera; Spatial resolution; PET; VALIDATION; MODEL;
D O I
10.1007/s13246-017-0607-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
GATE is currently considered in scintigraphic imaging as a powerful tool to develop, design and optimize nuclear medicine modalities. This paper describes the GATE simulation of a pixelated gamma camera which is dedicated to high resolution of small animals imaging. It consists of a CsI(Na) crystal array coupled to position sensitive photomultiplier tube. The simulation model includes photon tracking through low energy high resolution hexagonal parallel holes collimator, CsI(Na) pixelated crystal, back-compartment, and camera shielding. Simulations were compared with experimental results by some criteria such as energy spectrum, energy resolution, spatial resolution, sensitivity and count profiles obtained from line and point sources imaging. The acquired energy resolution show good agreement with measured spectra. Difference between calculated and experimental values is about 0.3% for absolute sensitivity measurement. The result of the image uniformity is more consistent after implementation of non-uniformity correction. These values were about 1.3 and 1.2% for experimental and simulation study in the central field of view, respectively. Measurements showed that the spatial resolutions differences at the head surface along the long dimensions of gamma camera for simulation and experimental differed by no more than 4%.Differences along the short axis were about 6%. The FWHMs of images of point and line sources show good consistency between experimental images and corresponding simulated ones. The difference between experimental and simulated system parameters was within 11%. Our results demonstrate the ability and flexibility of the Monte Carlo simulation for modeling pixelated gamma camera with position sensitive detector by selecting the appropriate parameters for digitizer chain and collimator position on the detector surface.
引用
收藏
页码:31 / 39
页数:9
相关论文
共 50 条
  • [1] Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT
    Payvand Taherparvar
    Alireza Sadremomtaz
    Australasian Physical & Engineering Sciences in Medicine, 2018, 41 : 31 - 39
  • [2] Monte Carlo simulation of a CsI(Tl) gamma camera dedicated to small animal imaging using GATE.
    Lazaro, D
    Buvat, I
    Donnarieix, D
    Loudos, G
    Giokaris, N
    Strul, D
    Santin, G
    Maigne, L
    Spanoudaki, V
    Styliaris, S
    Staelens, S
    Breton, V
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 288P - 288P
  • [3] Comparison of a GATE Monte Carlo Simulation Predictions to the Performance of a High-Resolution LYSO Based Dedicated Animal PET Camera
    Baghaei, Hossain
    Zhang, Yuxuan
    Li, Hongdi
    Ramirez, Rocio A.
    Liu, Shitao
    Wang, Chao
    An, Shaohui
    Wong, Wai-Hoi
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2388 - 2392
  • [4] A Monte Carlo Simulation Study of Optimization for Collimator in a Pixelated SPECT Camera
    Telikani, Zeinab
    Sadremomtaz, Alireza
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2019, 50 (01) : 163 - 170
  • [5] Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small-animal imaging
    Lazaro, D
    Buvat, I
    Loudos, G
    Strul, D
    Santin, G
    Giokaris, N
    Donnarieix, D
    Maigne, L
    Spanoudaki, V
    Styliaris, S
    Staelens, S
    Breton, V
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (02): : 271 - 285
  • [6] Development of GATE Monte Carlo simulation for a dual-head gamma camera
    Momennezhad, Mehdi
    Sadeghi, Ramin
    Nasseri, Shahrokh
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2012, 5 (02) : 222 - 228
  • [7] Development of GATE Monte Carlo simulation for a dual-head gamma camera
    Mehdi Momennezhad
    Ramin Sadeghi
    Shahrokh Nasseri
    Radiological Physics and Technology, 2012, 5 (2) : 222 - 228
  • [8] Quantification of Inter-crystal Scattering and Parallax Effect in Pixelated High Resolution Small Animal Gamma Camera: A Monte Carlo Study
    Adibpour, F.
    Ay, M. R.
    Sarkar, S.
    Loudos, G.
    5TH KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2011 (BIOMED 2011), 2011, 35 : 708 - +
  • [9] GATE Monte Carlo simulation of a high-sensitivity and high-resolution LSO-based small animal PET camera
    Baghaei, Flossain
    Zhang, Yuxuan
    Li, Hongdi
    Wang, Yu
    Kim, Soonseok
    Ramirez, Rocio A.
    Liu, Jiguo
    Liu, Shitao
    Wong, Wai-Hoi
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (05) : 1568 - 1573
  • [10] A GATE Monte Carlo Simulation of the Performance of a High-Sensitivity and High-Resolution LSO Based Small Animal PET Camera
    Baghaei, Hossain
    Zhang, Yuxuan
    Li, Hongdi
    Wang, Yu
    Kim, Soonseok
    Ramirez, Rocio
    Liu, Jiguo
    Liu, Shitao
    Wong, Wai-Hoi
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2476 - 2479