A Method for Improving Signal-to-Noise Ratio of Tissue Harmonic Imaging Based on Bayesian Inference Using Information of Fundamental Echoes

被引:12
作者
Yamamura, Takuya [1 ]
Tanabe, Masayuki [2 ]
Okubo, Kan [1 ]
Tagawa, Norio [1 ]
机构
[1] Tokyo Metropolitan Univ, Grad Sch Syst Design, Hino, Tokyo 1910065, Japan
[2] Kumamoto Univ, Grad Sch Sci & Technol, Kumamoto 8608555, Japan
关键词
D O I
10.1143/JJAP.51.07GF01
中图分类号
O59 [应用物理学];
学科分类号
摘要
A medical ultrasound imaging technique with a high spatial resolution and a high signal-to-noise ratio (SNR) is strongly required for precise diagnosis. To obtain high-resolution images, tissue harmonic imaging (THI) can be effectively adopted because of its wide bandwidth characteristics compared with fundamental imaging. However, the amplitude of harmonic components is much smaller than that of a fundamental component. Additionally, frequency dependent attenuation is severe, especially for harmonic components. These phenomena mean that THI is susceptible to noise. From these considerations, in this study, we propose a method for improving the SNR of THI by reducing noise using a signal processing technique. The proposed method is constructed based on the Bayesian inference using fundamental components of echoes, and it can hold the high-resolution nature of THI. We show the effectiveness of the proposed method through simulations. (C) 2012 The Japan Society of Applied Physics
引用
收藏
页数:11
相关论文
共 22 条
[1]   SPATIAL PROCESSING FOR COHERENT NOISE-REDUCTION IN ULTRASONIC-IMAGING [J].
BILGUTAY, NM ;
MURTHY, R ;
BENCHARIT, U ;
SANIIE, J .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 87 (02) :728-736
[2]  
Christopher LA, 2002, 2002 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING, PROCEEDINGS, P86, DOI 10.1109/ISBI.2002.1029199
[3]   Adaptive spectral sensitive filter for tissue harmonic imaging [J].
Demirbilek, FN ;
Yamakoshi, Y .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2003, 42 (01) :342-349
[4]   MAXIMUM LIKELIHOOD FROM INCOMPLETE DATA VIA EM ALGORITHM [J].
DEMPSTER, AP ;
LAIRD, NM ;
RUBIN, DB .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-METHODOLOGICAL, 1977, 39 (01) :1-38
[5]   Image Quality Improvement of Ultrasonic Computed Tomography on the Basis of Maximum Likelihood Expectation Maximization Algorithm Considering Anisotropic Acoustic Property and Time-of-Flight Interpolation [J].
Fan, Honghui ;
Yanagida, Hirotaka ;
Tamura, Yasutaka ;
Guo, Shuqiang ;
Saitoh, Tadashi ;
Takahashi, Tatsuhisa .
JAPANESE JOURNAL OF APPLIED PHYSICS, 2010, 49 (07)
[6]   Intravascular ultrasound tissue harmonic imaging in vivo [J].
Frijlink, Martijn E. ;
Goertz, David E. ;
van Damme, Luc C. A. ;
Krams, Rob ;
van der Steen, Antonius F. W. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (10) :1844-1852
[7]   Preliminary Study of Broadband Transducer for Measurement of Bone Characteristics [J].
Irie, Takasuke ;
Koumoto, Kouhei ;
Tanabe, Masayuki ;
Tagawa, Norio ;
Okubo, Kan ;
Akiyama, Iwaki ;
Itoh, Kouichi .
JAPANESE JOURNAL OF APPLIED PHYSICS, 2010, 49 (07)
[8]   Understanding bone responses in B-mode ultrasound images and automatic bone surface extraction using a Bayesian probabilistic framework [J].
Jain, AK ;
Taylor, RH .
MEDICAL IMAGING 2004: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2004, 5373 :131-142
[9]   Numerical calculation and measurement of nonlinear acoustic fields in ultrasound diagnosis [J].
Kawagishi, T ;
Saito, S ;
Mine, Y .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2002, 41 (5B) :3549-3554
[10]   Inverse scattering image reconstruction based on the multi-frequency reflection and transmission observation data from limited views [J].
Kim, JS ;
Yamada, A .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2002, 41 (5B) :3592-3597