摘要
This is an experimental study on human cadaver spines. The objective of this study is to compare the pullout forces between three screw augmentation methods and two different screw designs. Surgical interventions of patients with osteoporosis increase following the epidemiological development. Biomechanically the pedicle provides the strongest screw fixation in healthy bone, whereas in osteoporosis all areas of the vertebra are affected by the disease. This explains the high screw failure rates in those patients. Therefore PMMA augmentation of screws is often mandatory. This study involved investigation of the pullout forces of augmented transpedicular screws in five human lumbar spines (L1-L4). Each spine was treated with four different methods: non-augmented unperforated (solid) screw, perforated screw with vertebroplasty augmentation, solid screw with vertebroplasty augmentation and solid screw with balloon kyphoplasty augmentation. Screws were augmented with Polymethylmethacrylate (PMMA). The pullout forces were measured for each treatment with an Instron testing device. The bone mineral density was measured for each vertebra with Micro-CT. The statistical analysis was performed with a two-sided independent student t test. Forty screws (10 per group and level) were inserted. The vertebroplasty-augmented screws showed a significant higher pullout force (mean 918.5 N, P = 0.001) than control (mean 51 N), the balloon kyphoplasty group did not improve the pullout force significantly (mean 781 N, P > 0.05). However, leakage occurred in some cases treated with perforated screws. All spines showed osteoporosis on Micro-CT. Vertebroplasty-augmented screws, augmentation of perforated screws and balloon kyphoplasty augmented screws show higher pullout resistance than non-augmented screws. Significant higher pullout forces were only reached in the vertebroplasty augmented vertebra. The perforated screw design led to epidural leakage due to the position of the perforation in the screw. The position of the most proximal perforation is critical, depending on screw design and proper insertion depth. Nevertheless, using a properly designed perforated screw will facilitate augmentation and instrumentation in osteoporotic spines.
引用
收藏
页码:1462 / 1469
页数:8
相关论文
共 43 条
[1]   Temperature measurement during polymerization of polymethylmethacrylate cement used for vertebroplasty [J].
Belkoff, SM ;
Molloy, S .
SPINE, 2003, 28 (14) :1555-1559
[2]  
BOHNER M, 2008, BALLOON KYPHOPLASTY, P143
[3]   THE EFFECTS OF PEDICLE SCREW FIT - AN IN-VITRO STUDY [J].
BRANTLEY, AGU ;
MAYFIELD, JK ;
KOENEMAN, JB ;
CLARK, KR .
SPINE, 1994, 19 (15) :1752-1758
[4]   Primary pedicle screw augmentation in osteoporotic lumbar vertebrae - Biomechanical analysis of pedicle fixation strength [J].
Burval, Daniel J. ;
McLain, Robert F. ;
Milks, Ryan ;
Inceoglu, Serkan .
SPINE, 2007, 32 (10) :1077-1083
[5]   INFLUENCE OF BONE-MINERAL DENSITY ON THE FIXATION OF THORACOLUMBAR IMPLANTS - A COMPARATIVE-STUDY OF TRANSPEDICULAR SCREWS, LAMINAR HOOKS, AND SPINOUS PROCESS WIRES [J].
COE, JD ;
WARDEN, KE ;
HERZIG, MA ;
MCAFEE, PC .
SPINE, 1990, 15 (09) :902-907
[6]  
Cook Stephen D, 2004, Spine J, V4, P402, DOI 10.1016/j.spinee.2003.11.010
[7]   The Human Development Index and changes in standards of living: Some historical comparisons [J].
Crafts, N. F. R. .
EUROPEAN REVIEW OF ECONOMIC HISTORY, 1997, 1 :299-322
[8]   Instrumentation-related complications of multilevel fusions for adult spinal deformity patients over age 65 - Surgical considerations and treatment option in patients with poor bone quality [J].
DeWald, Christopher J. ;
Stanley, Thomas .
SPINE, 2006, 31 (19) :S144-S151
[9]   Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: A clinical evaluation [J].
Frankel, Bruce A. ;
Jones, Tanya ;
Wang, Chiang .
NEUROSURGERY, 2007, 61 (03) :531-537
[10]   A biomechanical cadaveric analysis of polymethylmethacrylate-augmented pedicle screw fixation [J].
Frankel, Bruce M. ;
D'Agostino, Sabino ;
Wang, Chiang .
JOURNAL OF NEUROSURGERY-SPINE, 2007, 7 (01) :47-53