Evaluation of four postprocessing methods for determination of cerebral blood volume and mean transit time by dynamic susceptibility contrast imaging

被引:60
作者
Perkiö, J
Aronen, HJ
Kangasmäki, A
Liu, YW
Karonen, J
Savolainen, S
Ostergaard, L
机构
[1] Univ Helsinki, Cent Hosp, Dept Radiol, FIN-00029 Helsinki, Finland
[2] Univ Helsinki, Dept Phys Sci, Helsinki, Finland
[3] Kuopio Univ Hosp, Dept Clin Radiol, SF-70210 Kuopio, Finland
[4] Univ Helsinki, Cent Hosp, Funct Brain Imaging Unit, FIN-00029 Helsinki, Finland
[5] Mikkeli Cent Hosp, Dept Radiol, Mikkeli, Finland
[6] Univ Helsinki, Cent Hosp, Dept Lab Diagnost, FIN-00029 Helsinki, Finland
[7] Aarhus Univ Hosp, Dept Neuroradiol, DK-8000 Aarhus, Denmark
关键词
cerebral blood volume; mean transit time; dynamic susceptibility contrast; numerical simulation;
D O I
10.1002/mrm.10126
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Four different postprocessing methods to determine cerebral blood volume (CBV) and contrast agent mean transit time (MTT) by dynamic susceptibility contrast (DSC) MRI were compared. CBV was determined by two different methods that integrate tracer concentration-time curves numerically and by two other methods that take recirculation into account. For the two methods that use numerical integration, one method cuts the integration after the first pass while the other method integrates over the whole time series, For the two methods that account for recirculation, one method uses a gamma-variate fit, whereas the other method utilizes tissue impulse response. All four methods determine MTT as the ratio of CBV and cerebral blood flow (CBF). In each case, CBF was obtained as the height of the impulse response obtained by deconvolving the tissue concentration-time curves with a noninvasively determined arterial input function. Monte Carlo simulations were performed to determine the reliability of the methods and the validity of the simulations was supported by observation of similar trends in 13 acute stroke patients. The method of determining CBV and subsequently MTT was found to affect the measured value especially in areas where MTT is prolonged, but had no apparent effect on the visually determined hypoperfusion volumes. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:973 / 981
页数:9
相关论文
共 30 条
[1]   CEREBRAL BLOOD-VOLUME MAPS OF GLIOMAS - COMPARISON WITH TUMOR GRADE AND HISTOLOGIC-FINDINGS [J].
ARONEN, HJ ;
GAZIT, IE ;
LOUIS, DN ;
BUCHBINDER, BR ;
PARDO, FS ;
WEISSKOFF, RM ;
HARSH, GR ;
COSGROVE, GR ;
HALPERN, EF ;
HOCHBERG, FH ;
ROSEN, BR .
RADIOLOGY, 1994, 191 (01) :41-51
[2]  
Aronen HJ, 2000, CLIN CANCER RES, V6, P2189
[4]   FUNCTIONAL CEREBRAL IMAGING BY SUSCEPTIBILITY-CONTRAST NMR [J].
BELLIVEAU, JW ;
ROSEN, BR ;
KANTOR, HL ;
RZEDZIAN, RR ;
KENNEDY, DN ;
MCKINSTRY, RC ;
VEVEA, JM ;
COHEN, MS ;
PYKETT, IL ;
BRADY, TJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (03) :538-546
[5]   Accuracy of gamma-variate fits to concentration-time curves from dynamic susceptibility-contrast enhanced MRI: Influence of time resolution, maximal signal drop and signal-to-noise [J].
Benner, T ;
Heiland, S ;
Erb, G ;
Forsting, M ;
Sartor, K .
MAGNETIC RESONANCE IMAGING, 1997, 15 (03) :307-317
[6]   Signal-to-noise analysis of cerebral blood volume maps from dynamic NMR imaging studies [J].
Boxerman, JL ;
Rosen, BR ;
Weisskoff, RM .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (03) :528-537
[7]  
Calamante F, 2000, MAGN RESON MED, V44, P466, DOI 10.1002/1522-2594(200009)44:3<466::AID-MRM18>3.0.CO
[8]  
2-M
[9]   MR CONTRAST DUE TO MICROSCOPICALLY HETEROGENEOUS MAGNETIC-SUSCEPTIBILITY - NUMERICAL SIMULATIONS AND APPLICATIONS TO CEREBRAL PHYSIOLOGY [J].
FISEL, CR ;
ACKERMAN, JL ;
BUXTON, RB ;
GARRIDO, L ;
BELLIVEAU, JW ;
ROSEN, BR ;
BRADY, TJ .
MAGNETIC RESONANCE IN MEDICINE, 1991, 17 (02) :336-347
[10]   A DECONVOLUTION METHOD FOR EVALUATING INDICATOR-DILUTION CURVES [J].
GOBBEL, GT ;
FIKE, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (11) :1833-1854