Utilization of a diol-stationary phase column in ion chromatographic separation of inorganic anions

被引:9
|
作者
Arai, Kaori [1 ]
Mori, Masanobu [1 ]
Kozaki, Daisuke [1 ,2 ]
Nakatani, Nobutake [3 ]
Itabashi, Hideyuki [1 ]
Tanaka, Kazuhiko [4 ]
机构
[1] Gunma Univ, Grad Sch Engn, Kiryu, Gunma 3768515, Japan
[2] Japan Soc Promot Sci, Chiyoda Ku, Tokyo 1028472, Japan
[3] Rakuno Gakuen Univ, Coll Agr Food & Environm Sci, Dept Environm & Symbiot Sci, Ebetsu, Hokkaido 0698501, Japan
[4] Hiroshima Univ, Grad Sch Int Dev & Cooperat, Higashihiroshima, Hiroshima 7398529, Japan
关键词
Non-suppressed ion chromatography; Diol-stationary phase; Anion-exchange; Inorganic anions; Acidic eluent; CATION-EXCHANGE RESIN; ALIPHATIC CARBOXYLIC-ACIDS; LIQUID-CHROMATOGRAPHY; COMMON ANIONS; POLYOXYETHYLENE; CONDUCTIVITY; BEHAVIORS; MECHANISM; SEAWATER; ELUENT;
D O I
10.1016/j.chroma.2012.10.058
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We describe the ion chromatographic separation of inorganic anions using a diol-stationary phase column (-CH(OH)CH2OH; diol-column) without charged functional groups. Anions were separated using acidic eluent as in typical anion-exchange chromatography. The retention volumes of anions on the diol-column increased with increasing H+ concentration in the eluent. The anion-exchange capacities of diol-columns in the acidic eluent(pH 2.8)were larger than that of zwitterionic stationary phase column but smaller than that of an anion-exchange column. The separation of anions using the diol-column was strongly affected by the interaction of H+ ions with the diol-functional groups and by the types of the eluents. In particular, the selection of the eluent was very important for controlling the retention time and resolution. Good separation was obtained using a diol-column (HILIC-10) with 5 mM phthalic acid as eluent. The limits of detection at a signal-to-noise ratio of 3 ranged from 1.2 to 2.7 mu M with relative standard deviations (RSD, n=5) of 0.04-0.07% for the retention time and 0.4-2.0% for the peak areas. This method was successfully applied to the determination of H2PO4-, Cl-, and NO3- in a liquid fertilizer sample. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:147 / 152
页数:6
相关论文
共 50 条
  • [41] Application of the bifunctional switch-type cryptand ion chromatographic stationary phase for ion separation
    Tsai, SS
    Shih, JS
    SEPARATION SCIENCE AND TECHNOLOGY, 1998, 33 (10) : 1407 - 1421
  • [42] CHROMATOGRAPHIC BEHAVIOR OF INORGANIC ANIONS ON A SEPHADEX G-15 COLUMN
    DEGUCHI, T
    HISANAGA, A
    NAGAI, H
    JOURNAL OF CHROMATOGRAPHY, 1977, 133 (01): : 173 - 179
  • [43] Separation of 8 Anions by Porous Monodisperse Poly ( styrene-co-divinyl benzene) Ion-pair Chromatographic Stationary Phase
    Jia Lei-Na
    Wang Fu-Qiang
    Gong Bo-Lin
    Hu Wen-Zhi
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2010, 38 (05) : 638 - 642
  • [44] Interpretative optimization of the isocratic ion chromatographic separation of anions
    Todorovic, Zaklina N.
    Rajakovic, Ljubinka V.
    Onjia, Antonije E.
    JOURNAL OF THE SERBIAN CHEMICAL SOCIETY, 2016, 81 (06) : 661 - 672
  • [45] REVERSED-PHASE CHROMATOGRAPHIC-SEPARATION OF HIGHLY CHARGED INORGANIC CATIONS AND ANIONS USING ION INTERACTION REAGENTS AND COMPETING IONS
    KIRK, AD
    HEWAVITHARANA, AK
    ANALYTICAL CHEMISTRY, 1988, 60 (08) : 797 - 801
  • [46] Improved separation of micro gas chromatographic column using mesoporous silica as a stationary phase support
    Luo, Fan
    Zhao, Bin
    Feng, Fei
    Hou, Lei
    You, Wenbin
    Xu, Pengcheng
    Zhou, Haimei
    Li, Xinxin
    TALANTA, 2018, 188 : 546 - 551
  • [47] DUAL FIBER-IN-CAPILLARY ANNULAR COLUMN WITH TERNARY STATIONARY PHASE FOR GAS CHROMATOGRAPHIC SEPARATION
    Qiu, Wenxi
    Diao, Peiyu
    Yuan, Hongyan
    Xiao, Dan
    Choi, Martin M. F.
    ANALYTICAL LETTERS, 2011, 44 (16) : 2721 - 2731
  • [48] SIMULTANEOUS DETERMINATION OF SOME INORGANIC AND ORGANIC-ANIONS BY MEANS OF AN ION-CHROMATOGRAPHIC TRIPLE COLUMN SYSTEM
    PIMMINGER, M
    PUXBAUM, H
    KOSSINA, I
    WEBER, M
    FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 320 (05): : 445 - 450
  • [49] ION-CHROMATOGRAPHIC DETERMINATION OF INORGANIC ANIONS IN VINEGAR SAMPLES
    DOLENC, J
    GORENC, D
    NAHRUNG-FOOD, 1994, 38 (04): : 434 - 438
  • [50] Ion Chromatographic Determination of Inorganic Anions in Environmental Samples of Korea
    Hyun-Mee Park
    Young-Man Kim
    Dai Woon Lee
    Su-won Lee
    Kang-Bong Lee
    Analytical Sciences, 2002, 18 : 343 - 346