Corrosion at the Cone/Taper Interface Leads to Failure of Large-diameter Metal-on-metal Total Hip Arthroplasties

被引:124
作者
Meyer, Heiko [1 ]
Mueller, Tina [1 ]
Goldau, Gesine [2 ]
Chamaon, Kathrin [1 ]
Ruetschi, Marcel [3 ]
Lohmann, Christoph H. [1 ]
机构
[1] Univ Magdeburg, Dept Orthopaed, D-39108 Magdeburg, Germany
[2] Univ Appl Sci, Magdeburg, Germany
[3] Loretto Hosp, Dept Orthopaed, Freiburg, Germany
关键词
WEAR PRODUCTS; ION LEVELS; REPLACEMENT; BEARINGS; DEBRIS; BLOOD; HYPERSENSITIVITY; COMPONENTS; CHROMIUM; COBALT;
D O I
10.1007/s11999-012-2502-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Metal-on-metal (MoM) THAs have reduced wear rates compared with metal-on-polyethylene. However, elevated serum metal ion levels and pseudotumors have been reported in large MoM articulations. Questions/purposes We therefore determined (1) if corrosion occurred at the cone/taper interface leading to instability in patients with large-diameter THAs; (2) how patients presented clinically and radiographically; (3) if adverse periprosthetic tissue reactions occurred; (4) whether metal was released from the implants into the periprosthetic tissues; and (5) if head size correlated with metal release. Methods We reviewed 114 patients who had revisions of large-diameter head MoM articulations. Mean time of implantation was 46 months. To identify adverse reactions and particle load, tissues were stained by hematoxylin and eosin and CD3/CD20/CD68 antibodies. Periprosthetic tissues were analyzed for metal content and distribution in different regions. Electrochemical reactions between the stem and adapter were investigated by a minicell electrode. Results Electrochemical studies on the stem and the head adapter showed a risk for galvanic corrosion. Ninety-four percent of patients had instability at the cone/taper interface. All patients presented with early clinical symptoms; 59 patients had radiographic signs of loosening. One hundred four patients had foreign body reactions and necrosis. The largest amounts of metal released were titanium or iron. We found no correlation between head size and metal ion release. Conclusions These findings suggest that in modular cone/taper connections, friction of the MoM articulations may cause failure of the cone/taper interface leading to galvanic corrosion and loosening. It is unclear whether the design of this MoM system provides sufficient stability at the taper.
引用
收藏
页码:3101 / 3108
页数:8
相关论文
共 45 条
[1]  
AMSTUTZ HC, 1992, CLIN ORTHOP RELAT R, P7
[2]   Metal ion levels in the blood of patients after hip resurfacing: A comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses [J].
Antoniou, John ;
Zukor, David J. ;
Mwale, Fackson ;
Minarik, William ;
Petit, Alain ;
Huk, Olga L. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A :142-148
[3]   Reproducible assessment of radiolucent lines in total knee arthroplasty [J].
Bach, CM ;
Biedermann, R ;
Goebel, G ;
Mayer, E ;
Rachbauer, F .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (434) :183-188
[4]  
Boehler N, 1997, Acta Orthop Belg, V63 Suppl 1, P96
[5]   High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[6]   Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements [J].
Clarke, IC ;
Good, V ;
Williams, P ;
Schroeder, D ;
Anissian, L ;
Stark, A ;
Oonishi, H ;
Schuldies, J ;
Gustafson, G .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H4) :331-347
[7]   CONCENTRATION OF WEAR PRODUCTS IN HAIR, BLOOD, AND URINE AFTER TOTAL HIP REPLACEMENT [J].
COLEMAN, RF ;
HERRINGTON, J ;
SCALES, JT .
BMJ-BRITISH MEDICAL JOURNAL, 1973, 1 (5852) :527-529
[8]   CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[9]   The argument for the use of Metasul as an articulation surface in total hip replacement [J].
Dorr, LD ;
Long, WT ;
Sirianni, L ;
Campana, M ;
Wan, ZN .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :80-85
[10]   The John Charnley Award Metal-on-Metal Hip Resurfacing versus Large-diameter Head Metal-on-Metal Total Hip Arthroplasty A Randomized Clinical Trial [J].
Garbuz, Donald S. ;
Tanzer, Michael ;
Greidanus, Nelson V. ;
Masri, Bassam A. ;
Duncan, Clive P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (02) :318-325