Do Obesity and/or Stripe Wear Increase Ceramic Liner Fracture Risk? An XFEM Analysis

被引:23
作者
Elkins, Jacob M. [1 ,2 ]
Pedersen, Douglas R. [1 ,2 ]
Callaghan, John J. [1 ,2 ,3 ]
Brown, Thomas D. [1 ,2 ]
机构
[1] Univ Iowa, Dept Orthopaed & Rehabil, Orthopaed Biomech Lab, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
[3] Iowa City Vet Adm Med Ctr, Iowa City, IA USA
基金
美国国家卫生研究院;
关键词
TOTAL HIP-ARTHROPLASTY; IMPINGEMENT; CONTACT; SQUEAKING; BEARINGS; FAILURE; HEADS;
D O I
10.1007/s11999-012-2562-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Hypothesized risk factors for fracture of ceramic liners include impingement, edge-loading, and cup malpositioning. These risk factors are similar to those for generation of stripe wear. However, it is unclear whether the biomechanical conditions contributing to stripe wear generation also increase the risk for ceramic liner fracture We asked whether (1) head stripe wear propensity; and (2) cup orientation would correlate with alumina liner fracture risk for instances of normal and elevated body weight. An eXtended Finite Element Method (XFEM) model was developed to investigate these mechanisms. Liner fracture risk for 36-mm alumina bearings was studied by simulating two fracture-prone motions: stooping and squatting. Twenty-five distinct cup orientations were considered with variants of both acetabular inclination and anteversion. Four separate body mass indices were considered: normal (25 kg/m(2)) and three levels of obesity (33, 42, and 50 kg/m(2)). Material properties were modified to simulate alumina with and without the presence of dispersed microflaws. The model was validated by corroboration with two previously published ceramic liner fracture studies. Of 200 XFEM simulations with flaw-free alumina, fracture occurred in eight instances, all of them involving obesity. Each of these occurred with cups in a parts per thousand currency sign 37A degrees inclination and in 0A degrees anteversion. For 200 corresponding simulations with microflawed alumina, fracture propensity was greatest for cups with higher (edge loading-associated) scraping wear. Fracture risk was greatest for cups with lower inclination (average 42A degrees for fractured cases versus 48A degrees for nonfractured cases) and lower anteversion (9A degrees versus 20A degrees). Fracture propensity for 36-mm liners was elevated for cups with decreased anteversion and/or inclination and under conditions of patient obesity. Factors causing stripe wear, including obesity and cup malpositioning, also involve increased risk of ceramic liner fracture and merit heightened concern.
引用
收藏
页码:527 / 536
页数:10
相关论文
共 42 条
[1]   Comparison of Surgical Outcomes and Implant Wear Between Ceramic-Ceramic and Ceramic-Polyethylene Articulations in Total Hip Arthroplasty [J].
Amanatullah, Derek F. ;
Landa, Joshua ;
Strauss, Eric J. ;
Garino, Jonathan P. ;
Kim, Sunny H. ;
Di Cesare, Paul E. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (06) :72-77
[2]   CONTACT AND RUBBING OF FLAT SURFACES [J].
ARCHARD, JF .
JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) :981-988
[3]   The Epidemiology of Bearing Surface Usage in Total Hip Arthroplasty in the United States [J].
Bozic, Kevin J. ;
Kurtz, Steven ;
Lau, Edmund ;
Ong, Kevin ;
Chiu, Vanessa ;
Vail, Thomas P. ;
Rubash, Harry E. ;
Berry, Daniel J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (07) :1614-1620
[4]   Friction of total hip replacements with different bearings and loading conditions [J].
Brockett, Claire ;
Williams, Sophie ;
Jin, Zhongmin ;
Isaac, Graham ;
Fisher, John .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 81B (02) :508-515
[5]   Multiple scale modeling for cortical bone fracture in tension using X-FEM [J].
Budyn, Elisa ;
Hoc, Thierry .
EUROPEAN JOURNAL OF COMPUTATIONAL MECHANICS, 2007, 16 (02) :213-236
[6]   The John Charnley Award Risk Factors for Cup Malpositioning Quality Improvement Through a Joint Registry at a Tertiary Hospital [J].
Callanan, Mark C. ;
Jarrett, Bryan ;
Bragdon, Charles R. ;
Zurakowski, David ;
Rubash, Harry E. ;
Freiberg, Andrew A. ;
Malchau, Henrik .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) :319-329
[7]   Ceramic-on-Ceramic Total Hip Arthroplasty: Update [J].
Capello, William N. ;
D'Antonio, James A. ;
Feinberg, Judy R. ;
Manley, Michael T. ;
Naughton, Marybetb .
JOURNAL OF ARTHROPLASTY, 2008, 23 (07) :39-43
[8]  
CeramTec AG, 2008, SCI INF PERF DAT BIO, P11
[9]   Fracture Propagation Propensity of Ceramic Liners During Impingement-Subluxation A Finite Element Exploration [J].
Elkins, Jacob M. ;
Pedersen, Douglas R. ;
Callaghan, John J. ;
Brown, Thomas D. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (04) :520-526
[10]   Edge-loading severity as a function of cup lip radius in metal-on-metal total hipsuA finite element analysis [J].
Elkins, Jacob M. ;
Kruger, Karen M. ;
Pedersen, Douglas R. ;
Callaghan, John J. ;
Brown, Thomas D. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (02) :169-177