Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film

被引:2
作者
Ghorbani, Mahdi [1 ]
Toossi, Mohammad Taghi Bahreyni [2 ]
Mowlavi, Ali Asghar [3 ]
Roodi, Shahram Bayani [2 ]
Meigooni, Ali Soleimani [4 ]
机构
[1] N Khorasan Univ Med Sci, Bojnurd, North Khorasan, Iran
[2] Mashhad Univ Med Sci, Fac Med, Dept Med Phys, Med Phys Res Ctr, Mashhad, Iran
[3] Hakim Sabzevari Univ, Sch Sci, Dept Phys, Sabzevar, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
关键词
color scanner; GZP6 brachytherapy source; Co-60 high dose rate source; HDR; EBT radiochromic film; FLAT-BED SCANNER; ENERGY-DEPENDENCE; SENSITIVITY;
D O I
10.2478/v10019-012-0015-1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. The aim of this study is to evaluate the performance of a color scanner as a radiochromic film reader in two dimensional dosimetry around a high dose rate brachytherapy source. Materials and methods. A Microtek Scan Maker 1000XL film scanner was utilized for the measurement of dose distribution around a high dose rate GZP6 Co-60 brachytherapy source with GafChromic (R) EBT radiochromic films. In these investigations, the non-uniformity of the film and scanner response, combined, as well as the films sensitivity to scanner's light source was evaluated using multiple samples of films, prior to the source dosimetry. The results of these measurements were compared with the Monte Carlo simulated data using MCNPX code. In addition, isodose curves acquired by radiochromic films and Monte Carlo simulation were compared with those provided by the GZP6 treatment planning system. Results. Scanning of samples of uniformly irradiated films demonstrated approximately 2.85% and 4.97% non-uniformity of the response, respectively in the longitudinal and transverse directions of the film. Our findings have also indicated that the film response is not affected by the exposure to the scanner's light source, particularly in multiple scanning of film. The results of radiochromic film measurements are in good agreement with the Monte Carlo calculations (4%) and the corresponding dose values presented by the GZP6 treatment planning system (5%). Conclusions. The results of these investigations indicate that the Microtek Scan Maker 1000XL color scanner in conjunction with GafChromic EBT film is a reliable system for dosimetric evaluation of a high dose rate brachytherapy source.
引用
收藏
页码:363 / 369
页数:7
相关论文
共 22 条
  • [1] The use of a reflective scanner to study radiochromic film response
    Alva, H
    Mercado-Uribe, H
    Rodríguez-Villafuerte, M
    Brandan, ME
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) : 2925 - 2933
  • [2] Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range
    Butson, MJ
    Cheung, T
    Yu, PKN
    [J]. APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) : 60 - 62
  • [3] Radiochromic film for medical radiation dosimetry
    Butson, MJ
    Yu, PKN
    Cheung, T
    Metcalfe, P
    [J]. MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) : 61 - 120
  • [4] Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies
    Chiu-Tsao, ST
    Ho, Y
    Shankar, R
    Wang, L
    Harrison, LB
    [J]. MEDICAL PHYSICS, 2005, 32 (11) : 3350 - 3354
  • [5] High-precision GAFCHROMIC EBT film-based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non-uniformity correction
    Chung, Heeteak
    Lynch, Bart
    Samant, Sanjiv
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (02): : 101 - 115
  • [6] Dosimetric evaluation of a 320 detector row CT scanner unit
    de Denaro, Mario
    Bregant, Paola
    [J]. RADIOLOGY AND ONCOLOGY, 2011, 45 (01) : 64 - 67
  • [7] Precise radiochromic film dosimetry using a flat-bed document scanner
    Devic, S
    Seuntjens, J
    Sham, E
    Podgorsak, EB
    Schmidtlein, CR
    Kirov, AS
    Soares, CG
    [J]. MEDICAL PHYSICS, 2005, 32 (07) : 2245 - 2253
  • [8] Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry
    Ferreira, B. C.
    Lopes, M. C.
    Capela, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (04) : 1073 - 1085
  • [9] Monte Carlo calculations of the dose distribution around a commercial gynecologic tandem applicator
    Gifford, KA
    Mourtada, F
    Cho, SH
    Lawyer, A
    Horton, JL
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 77 (02) : 210 - 215
  • [10] A novel method of radiochromic film dosimetry using a color scanner
    Hupe, Oliver
    Brunzendorf, Jens
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4085 - 4094