Reliability of an infrared auditory thermometer in the measurement of oral temperature

被引:0
作者
Moya Villaescusa, M. Jose [2 ]
Perez, A. Sanchez [2 ]
Bermejo Fenoll, Ambrosio [1 ]
机构
[1] Univ Murcia, Dept Oral Med, Univ Dent Clinic, Murcia, Spain
[2] Univ Murcia, Dept Periodontol, Univ Dent Clinic, Murcia, Spain
来源
MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL | 2008年 / 13卷 / 06期
关键词
body temperature; temperature; fever; thermometers; thermography;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim: To evaluate the reliability of the measurements obtained from different parts of the oral cavity using an infrared auditory thermometer, and to contrast these results with those obtained from the axilla and auditory canal. Study design: A comparative descriptive design was used to establish differences between the temperature recordings made in 66 healthy volunteers with an infrared auditory thermometer applied to different parts of the body (oral cavity and left ear) versus the recordings obtained with a glass mercury thermometer in the axillary zone. The study sample was balanced with respect to gender, and the mean age was 21.15 (+/- 1.61) years. Results: The largest standard deviation of all the locations corresponded to the left ear. This variable did not present a normal distribution. However, there were no statistically significant differences among the consecutive measurements made in the different anatomical (oral or axillary) regions. Conclusions: The thermal device used in this study seems to fulfill most of the requirements of an ideal clinical thermometer for yielding objective information on body temperature. It is easy to use, noninvasive, inexpensive and rapid. The temperature in the oral cavity is very representative of body temperature.
引用
收藏
页码:E385 / E389
页数:5
相关论文
共 23 条
[1]   THERMAL ENVIRONMENT OF TEETH DURING OPEN-MOUTH RESPIRATION [J].
BOEHM, RF .
JOURNAL OF DENTAL RESEARCH, 1972, 51 (01) :75-&
[2]   Simultaneous assessment of autonomic function and anxiety-related behavior in BALB/c and C57BL/6 mice [J].
Depino, Amaicha Mara ;
Gross, Cornelius .
BEHAVIOURAL BRAIN RESEARCH, 2007, 177 (02) :254-260
[3]   CERUMEN OCCLUSION LOWERS INFRARED TYMPANIC MEMBRANE TEMPERATURE-MEASUREMENT [J].
DOEZEMA, D ;
LUNT, M ;
TANDBERG, D .
ACADEMIC EMERGENCY MEDICINE, 1995, 2 (01) :17-19
[4]   THE GENIUS INFRARED TYMPANIC THERMOMETER - AN EVALUATION FOR CLINICAL USE [J].
EDGE, G ;
MORGAN, M .
ANAESTHESIA, 1993, 48 (07) :604-607
[5]   Temperature measurement: comparison of non-invasive methods used in adult critical care [J].
Farnell, S ;
Maxwell, L ;
Tan, S ;
Rhodes, A ;
Philips, B .
JOURNAL OF CLINICAL NURSING, 2005, 14 (05) :632-639
[6]   Infrared thermometry in the auditory canal: Evaluation on the DIATEK 9000 Instatemp and DIATEK 9000 Thermoguide [J].
Fritz, U ;
Rohrberg, M ;
Lange, C ;
Weyland, W ;
Brauer, A ;
Braun, U .
ANAESTHESIST, 1996, 45 (11) :1059-1066
[7]   The effect of factors such as handedness, sex and age on body temperature measured by an infrared 'tympanic' thermometer [J].
School of Applied Science, University of Glamorgan, Mid Glamorgan, CF37 1DL, United Kingdom ;
不详 .
J. Med. Eng. Technol., 2006, 4 (235-241) :235-241
[8]   THE RELATIONSHIP BETWEEN GINGIVAL TISSUE TEMPERATURES AND VARIOUS INDICATORS OF GINGIVAL INFLAMMATION [J].
HOLTHUIS, AF ;
GELSKEY, SC ;
CHEBIB, FS .
JOURNAL OF PERIODONTOLOGY, 1981, 52 (04) :187-189
[9]   Screening for fever in an adult emergency department: Oral vs tympanic thermometry [J].
Hooker, EA ;
Houston, H .
SOUTHERN MEDICAL JOURNAL, 1996, 89 (02) :230-234
[10]  
HORVATH STEVEN M., 1950, JOUR AMER MED ASSOC, V144, P1562