The Endothelial Sample Size Analysis in Corneal Specular Microscopy Clinical Examinations

被引:29
作者
Abib, Fernando C. [1 ,2 ]
Holzchuh, Ricardo [3 ]
Schaefer, Artur [3 ]
Schaefer, Tania
Godois, Ronialci [4 ]
机构
[1] Erast Gaertner Hosp, Dept Ophthalmol, Curitiba, Parana, Brazil
[2] Univ Fed Parana, BR-80060000 Curitiba, Parana, Brazil
[3] Univ Sao Paulo, Dept Ophthalmol Santa Casa Misericorida Sao Paulo, Sao Paulo, Brazil
[4] Araucaria Ambiental, Cianorte, Brazil
关键词
cornea; endothelium; specular microscopy; specular microscope; analysis; sample; sample error;
D O I
10.1097/ICO.0b013e3181cc7961
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate endothelial cell sample size and statistical error in corneal specular microscopy (CSM) examinations. Methods: One hundred twenty examinations were conducted with 4 types of corneal specular microscopes: 30 with each BioOptics, CSO, Konan, and Topcon corneal specular microscopes. All endothelial image data were analyzed by respective instrument software and also by the Cells Analyzer software with a method developed in our lab(US Patent). A reliability degree (RD) of 95% and a relative error (RE) of 0.05 were used as cut-off values to analyze images of the counted endothelial cells called samples. The sample size mean was the number of cells evaluated on the images obtained with each device. Only examinations with RE<0.05 were considered statistically correct and suitable for comparisons with future examinations. The Cells Analyzer software was used to calculate the RE and customized sample size for all examinations. Results: Bio-Optics: sample size, 97 +/- 22 cells; RE, 6.52 +/- 0.86; only 10% of the examinations had sufficient endothelial cell quantity (RE<0.05); customized sample size, 162 +/- 34 cells. CSO: sample size, 110 +/- 20 cells; RE, 5.98 +/- 0.98; only 16.6% of the examinations had sufficient endothelial cell quantity (RE<0.05); customized sample size, 157 +/- 45 cells. Konan: sample size, 80 +/- 27 cells; RE, 10.6 +/- 3.67; none of the examinations had sufficient endothelial cell quantity (RE>0.05); customized sample size, 336 +/- 131 cells. Topcon: sample size, 87 +/- 17 cells; RE, 10.1 +/- 2.52; none of the examinations had sufficient endothelial cell quantity (RE>0.05); customized sample size, 382 +/- 159 cells. Conclusions: A very high number of CSM examinations had sample errors based on Cells Analyzer software. The endothelial sample size (examinations) needs to include more cells to be reliable and reproducible. The Cells Analyzer tutorial routine will be useful for CSM examination reliability and reproducibility.
引用
收藏
页码:546 / 550
页数:5
相关论文
共 23 条
[1]   Behavior of corneal endothelial density over a lifetime [J].
Abib, FC ;
Barreto, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (10) :1574-1578
[2]  
Abib FC, 2000, POS OPERATORIO CATAR, P101
[3]  
ABIB FC, 2006, CORNEAL SPECULAR MIC
[4]  
Abib FC, 2000, MICROSCOPIA ESPECULA, P35
[5]  
Altman DG, 2000, Statistics with confidence, V2nd
[6]  
[Anonymous], 2006, CELLS ANAL CORNEAL E
[7]  
Bisquerra R., 2004, INTRO ESTATISTICA
[8]  
BOLFARINE H., 2005, ELEMENTOS AMOSTRAGEM
[9]  
*FDA, 1999, SUMM SAF EFF DAT
[10]  
HIRST LW, 1980, INVEST OPHTH VIS SCI, V19, P2