Delayed coil migration from a small wide-necked aneurysm after stent-assisted embolization: case report and literature review

被引:26
作者
Gao, Bu-Lang [1 ]
Li, Ming-Hua [1 ]
Wang, Yong-Li [1 ]
Fang, Chun [1 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 6, Dept Diagnost & Intervent Radiol, Shanghai 200233, Peoples R China
关键词
coil escape; intracranial aneurysm; wide-necked aneurysm; stent-assisted technique; embolization;
D O I
10.1007/s00234-005-0044-1
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Introduction: We present a previously unreported complication following the treatment of a patient with two small, wide-necked, posterior communicating artery aneurysms. Methods: Endovascular embolization of one aneurysm was performed using a stent-assisted technique. Follow-up angiography 5 months later revealed that a coil had escaped the confinement of the stent and migrated distally without occluding any arterial branches or causing symptoms. This case report demonstrates that although a rare occurrence, a coil can break loose from the stent. Discussion: We discuss the potential mechanisms of this phenomenon and review the literature on stent-assisted aneurysm coiling in order to raise awareness of this event when embolizing small, wide-necked aneurysms with a stent-assisted technique.
引用
收藏
页码:333 / 337
页数:5
相关论文
共 28 条
[1]  
Akpek S, 2005, AM J NEURORADIOL, V26, P1223
[2]   Treatment of basilar artery bifurcation aneurysms by using Guglielmi detachable coils: a 6-year experience [J].
Bavinzski, G ;
Killer, M ;
Gruber, A ;
Reinprecht, A ;
Gross, CE ;
Richling, B .
JOURNAL OF NEUROSURGERY, 1999, 90 (05) :843-852
[3]   Endovascular occlusion of wide-necked aneurysms with a new intracranial microstent (neuroform) and detachable coils [J].
Benitez, RP ;
Silva, MT ;
Klem, J ;
Veznedaroglu, E ;
Rosenwasser, RH .
NEUROSURGERY, 2004, 54 (06) :1359-1367
[4]   Intracranial aneurysms: Endovascular treatment with mechanical detachable spirals in 60 aneurysms [J].
Cognard, C ;
Pierot, L ;
Boulin, A ;
Weill, A ;
Toevi, M ;
Castaings, L ;
Rey, A ;
Moret, J .
RADIOLOGY, 1997, 202 (03) :783-792
[5]  
Doerfler A, 2005, AM J NEURORADIOL, V26, P862
[6]   Double-stent method: therapeutic alternative for small wide-necked aneurysms [J].
Doerfler, A ;
Wanke, I ;
Egelhof, T ;
Stolke, D ;
Forsting, M .
JOURNAL OF NEUROSURGERY, 2004, 100 (01) :150-154
[7]   Usefulness of the neuroform stent for the treatment of cerebral aneurysms: Results at initial (3-6-mo) follow-up [J].
Fiorella, D ;
Albuquerque, FC ;
Deshmukh, VR ;
McDougall, CG .
NEUROSURGERY, 2005, 56 (06) :1191-1201
[8]   Preliminary experience using the neuroform stent for the treatment of cerebral aneurysms [J].
Fiorella, D ;
Albuquerque, FC ;
Han, P ;
McDougall, CG .
NEUROSURGERY, 2004, 54 (01) :6-16
[9]   Percutaneous intracranial stent placement for aneurysms [J].
Han, PP ;
Albuquerque, FC ;
Ponce, FA ;
MacKay, CI ;
Zabramski, JM ;
Spetzler, RF ;
McDougall, CG .
JOURNAL OF NEUROSURGERY, 2003, 99 (01) :23-30
[10]   Endovascular coil occlusion of intracranial aneurysms assisted by a novel self-expandable nitinol microstent (Neuroform) [J].
Henkes, H ;
Bose, A ;
Felber, S ;
Miloslavski, E ;
Berg-Dammer, E ;
Kühne, D .
INTERVENTIONAL NEURORADIOLOGY, 2002, 8 (02) :107-119