Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator

被引:4
|
作者
Martins, L. Lages [1 ]
Ribeiro, A. Silva [1 ]
Alves e Sousa, J. [2 ]
Forbes, Alistair B. [3 ]
机构
[1] LNEC, Lisbon, Portugal
[2] LREC, Funchal, Portugal
[3] NPL, Teddington, Middx, England
关键词
Dew-point temperature; Iterative model; Measurement uncertainty;
D O I
10.1007/s10765-011-1005-z
中图分类号
O414.1 [热力学];
学科分类号
摘要
This article describes the measurement uncertainty evaluation of the dew-point temperature when using a two-pressure humidity generator as a reference standard. The estimation of the dew-point temperature involves the solution of a non-linear equation for which iterative solution techniques, such as the Newton-Raphson method, are required. Previous studies have already been carried out using the GUM method and the Monte Carlo method but have not discussed the impact of the approximate numerical method used to provide the temperature estimation. One of the aims of this article is to take this approximation into account. Following the guidelines presented in the GUM Supplement 1, two alternative approaches can be developed: the forward measurement uncertainty propagation by the Monte Carlo method when using the Newton-Raphson numerical procedure; and the inverse measurement uncertainty propagation by Bayesian inference, based on prior available information regarding the usual dispersion of values obtained by the calibration process. The measurement uncertainties obtained using these two methods can be compared with previous results. Other relevant issues concerning this research are the broad application to measurements that require hygrometric conditions obtained from two-pressure humidity generators and, also, the ability to provide a solution that can be applied to similar iterative models. The research also studied the factors influencing both the use of the Monte Carlo method (such as the seed value and the convergence parameter) and the inverse uncertainty propagation using Bayesian inference (such as the pre-assigned tolerance, prior estimate, and standard deviation) in terms of their accuracy and adequacy.
引用
收藏
页码:1568 / 1582
页数:15
相关论文
共 50 条
  • [21] Consistency of the National Realization of Dew-Point Temperature Using Standard Humidity Generators
    R. Benyon
    T. Vicente
    International Journal of Thermophysics, 2012, 33 : 1550 - 1558
  • [22] Consistency of the National Realization of Dew-Point Temperature Using Standard Humidity Generators
    Benyon, R.
    Vicente, T.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2012, 33 (8-9) : 1550 - 1558
  • [23] The effect of water contamination on the dew-point temperature scale realization with humidity generators
    Vilbaste, M.
    Heinonen, M.
    Saks, O.
    Leito, I.
    METROLOGIA, 2013, 50 (04) : 329 - 336
  • [24] Dew-point measurement in industrial applications
    Soleyn, K
    MEASUREMENTS & CONTROL, 1996, (177): : 121 - 125
  • [25] PROPAGATION OF VARIANCE IN DEW-POINT AND PSYCHROMETER HUMIDITY GAUGES
    CHAPMAN, GD
    METROLOGIA, 1993, 29 (06) : 379 - 387
  • [26] AUTOMATIC DEW-POINT TEMPERATURE SENSOR
    GRAICHEN, H
    RASCATI, R
    GONZALEZ, RR
    JOURNAL OF APPLIED PHYSIOLOGY, 1982, 52 (06) : 1658 - 1660
  • [27] A New Primary Dew-Point Generator at TUBITAK UME
    Aytekin, S. Oguz
    Karaboce, N.
    Heinonen, M.
    Sairanen, H.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2018, 39 (05)
  • [28] A New Primary Dew-Point Generator at TUBITAK UME
    S. Oğuz Aytekin
    N. Karaböce
    M. Heinonen
    H. Sairanen
    International Journal of Thermophysics, 2018, 39
  • [29] SEMICONDUCTOR SENSOR MIRROR FOR DEW-POINT MEASUREMENT
    DZHAMAGIDZE, SZ
    KOBAKHIDZE, KA
    TODUA, AM
    CHATOV, VA
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1989, 32 (02) : 503 - 504
  • [30] Development of the High-Temperature Dew-Point Generator Over the Past 15 Years
    R. Bosma
    J. Nielsen
    A. Peruzzi
    International Journal of Thermophysics, 2017, 38