OPTIMIZED 2D ARRAY DESIGN FOR ULTRASOUND IMAGING

被引:0
作者
Diarra, Bakary [1 ,2 ,3 ,4 ,5 ,6 ]
Liebgott, Herve [1 ,2 ,3 ,4 ,5 ]
Robini, Marc [1 ,2 ,3 ,4 ,5 ]
Tortoli, Piero [6 ]
Cachard, Christian [1 ,2 ,3 ,4 ,5 ]
机构
[1] Univ Lyon, CREATIS, Villeurbanne, France
[2] INSERM, U1044, F-69365 Lyon, France
[3] CNRS, UMR 5220, F-69365 Lyon, France
[4] Univ Lyon 1, F-69365 Lyon, France
[5] INSA Lyon, F-69365 Lyon, France
[6] Univ Florence, Dept Elect & Telecommun, Florence, Italy
来源
2012 PROCEEDINGS OF THE 20TH EUROPEAN SIGNAL PROCESSING CONFERENCE (EUSIPCO) | 2012年
关键词
ultrasound; 2D array; sparse array; simulated annealing;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Ultrasound imaging is one of the least expensive and safest diagnostic modalities routinely used. An attractive recent development in this field is three-dimensional (3D) imaging with two-dimensional (2D) matrix probes. The difficulty in implementing these probes comes from their large number of elements; for instance, the probe considered in this paper is composed of 1024 elements, whereas the number of channels of most current beamformers ranges from 64 to 256. To reduce the number of active elements, we propose a new sparse array design technique based on simulated annealing. Our method is capable of significantly reducing the number of probe elements as well as the side lobe level in a reasonable amount of computing time. Experiments in the context of hepatic biopsy show that good imaging performance can be obtained with only 177 active elements out of the total of 1024.
引用
收藏
页码:2718 / 2722
页数:5
相关论文
共 15 条
[1]   Sparse 2-D Arrays for 3-D phased array imaging - Design methods [J].
Austeng, A ;
Holm, S .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (08) :1073-1086
[3]   Optimized simulated annealing algorithm for thinning and weighting large planar arrays [J].
Chen, Peng ;
Shen, Bin-jian ;
Zhou, Li-sheng ;
Chen, Yao-wu .
JOURNAL OF ZHEJIANG UNIVERSITY-SCIENCE C-COMPUTERS & ELECTRONICS, 2010, 11 (04) :261-269
[4]  
Diarra B., 2011, IEEE INT UL IN PRESS
[5]  
Ezhilarasi M., 2008, CALICUT MED J, V6
[6]  
HAJEK B., MATH OPERATIONS RES, V13, P311
[7]  
Holm S., 2009, SPARSE SAMPLING ARRA
[8]  
Jensen J. A., 1996, Medical & Biological Engineering & Computing, V34, P351
[9]   CALCULATION OF PRESSURE FIELDS FROM ARBITRARILY SHAPED, APODIZED, AND EXCITED ULTRASOUND TRANSDUCERS [J].
JENSEN, JA ;
SVENDSEN, NB .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (02) :262-267
[10]   Transthoracic hepatic radiofrequency ablation [J].
Lee, NN ;
O'Rourke, RW ;
Cheng, J ;
Hansen, PD .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2004, 18 (11) :1672-1674