Contrast Media-Doped Hydrodissection During Thermal Ablation: Optimizing Contrast Media Concentration for Improved Visibility on CT Images

被引:34
作者
Campbell, Calista [2 ]
Lubner, Meghan G. [2 ]
Hinshaw, J. Louis [2 ]
del Rio, Alejandro Munoz [2 ]
Brace, Christopher L. [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Biomed Engn, Madison, WI 53705 USA
[2] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
基金
美国国家卫生研究院;
关键词
hydrodissection; interventional oncology; microwave ablation; radiofrequency ablation; thermal ablation; PERCUTANEOUS RADIOFREQUENCY ABLATION; MINIMIZING DIAPHRAGMATIC INJURY; 5-PERCENT DEXTROSE; LIVER-TUMORS; HEPATOCELLULAR-CARCINOMA; RENAL MASSES; WATER; EFFICACY; SALINE; MODEL;
D O I
10.2214/AJR.11.7999
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to determine a concentration of iodinated contrast media in saline and 5% dextrose in water (D5W) for organ hydrodissection, a technique used to physically separate and protect tissues adjacent to thermal ablations. MATERIALS AND METHODS. A total of 28 samples were prepared from 1:1000-1:1 iohexol or iothalamate meglumine contrast media in either normal saline or D5W. Samples alone or juxtaposed with a homogeneous liver-mimicking phantom were imaged by CT using 80-120 kVp and 10-300 mAs. Mean CT numbers and noise were measured from the fluid, background air, phantom adjacent to the fluid, and phantom distant from the fluid. Visibility was determined from the contrast-to-noise ratio between the fluid and phantom, whereas streaking artifact was quantified by relative noise in the phantom. Measures were individually fit using multiple linear regression to determine an optimal contrast-to-fluid ratio for increased visualization without streaking. Contrast media-and blood-doped saline and D5W were also tested to determine whether such doping altered their electrical conductivity. RESULTS. Iohexol concentration most influenced CT number; volumetric ratios of 1: 1000-1:1 produced 20 HU to over 3000 HU. CT numbers were weakly dependent on x-ray tube voltage, whereas contrast-to-noise ratio and streaking artifacts were somewhat dependent on tube output. An optimal ratio of iohexol in fluid was determined to be 1:50. There was no significant difference between the electrical impedances of doped and pure saline or D5W (p > 0.5, all cases). CONCLUSION. A 1:50 ratio of iohexol in saline or D5W provides an optimal combination of increased visibility on CT without streaking artifacts.
引用
收藏
页码:677 / 682
页数:6
相关论文
共 29 条
  • [1] Percutaneous CT-Guided Radiofrequency Ablation of Renal Cell Carcinoma: Efficacy of Organ Displacement by Injection of 5% Dextrose in Water Into the Retroperitoneum
    Arellano, Ronald S.
    Garcia, Rodrigo G.
    Gervais, Debra A.
    Mueller, Peter R.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (06) : 1686 - 1690
  • [2] Brace Christopher L, 2006, Conf Proc IEEE Eng Med Biol Soc, V2006, P5021
  • [3] Hypertensive crisis in a patient undergoing percutaneous radiofrequency ablation of an adrenal mass under general anesthesia
    Chini, EN
    Brown, MJ
    Farrell, MA
    Charboneau, JW
    [J]. ANESTHESIA AND ANALGESIA, 2004, 99 (06) : 1867 - 1869
  • [4] CT-guided radiofrequency ablation of osteoid osteoma: long-term results
    Cioni, R
    Armillotta, N
    Bargellini, I
    Zampa, V
    Cappelli, C
    Vagli, P
    Boni, G
    Marchetti, S
    Consoli, V
    Bartolozzi, C
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (07) : 1203 - 1208
  • [5] Radiofrequency ablation in the musculoskeletal system
    Davis, KW
    Choi, JJ
    Blankenbaker, DG
    [J]. SEMINARS IN ROENTGENOLOGY, 2004, 39 (01) : 129 - 144
  • [6] Utility of Iodinated Contrast Medium in Hydrodissection Fluid when Performing Renal Tumor Ablation
    DeBenedectis, Carolynn M.
    Beland, Michael D.
    Dupuy, Damian E.
    Mayo-Smith, William W.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 21 (05) : 745 - 747
  • [7] Radiofrequency ablation of liver tumours: systematic review
    Decadt, B
    Siriwardena, AK
    [J]. LANCET ONCOLOGY, 2004, 5 (09) : 550 - 560
  • [8] Paranephric water instillation: A technique to prevent bowel injury during percutaneous renal radiofrequency ablation
    Farrell, MA
    Charboneau, JW
    Callstrom, MR
    Reading, CC
    Engen, DE
    Blute, ML
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (05) : 1315 - 1317
  • [9] CT mapping of the distribution of saline during radiofrequency ablation with perfusion electrodes
    Gillams, AR
    Lees, WR
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 28 (04) : 476 - 480
  • [10] Bowel Displacement for CT-Guided Tumor Radiofrequency Ablation: Techniques and Anatomic Considerations
    Ginat, Daniel T.
    Saad, Wael
    Davies, Mark
    Walman, David
    Erturk, Erdal
    [J]. JOURNAL OF ENDOUROLOGY, 2009, 23 (08) : 1259 - 1264