An error in the CoaguChek S® device results in prosthetic heart valve thrombosis

被引:3
作者
van den Brule, Judith M. D. [2 ]
Romkes, Johannes H. [2 ]
Brouwer, Marc A. [2 ]
Noyez, Luc [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Ctr Heart, Dept Cardiothorac Surg, NL-6500 HB Nijmegen, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Ctr Heart, Dept Cardiol, NL-6500 HB Nijmegen, Netherlands
关键词
Anticoagulation; CoaguChek S (R); Prosthetic heart valve; Thrombosis;
D O I
10.1510/icvts.2010.233288
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The CoaguChek S (R) device offers patients on oral anticoagulant treatment, a simple and fast method of direct monitoring of their international normalized ratio (INR). It is comparable with the glucose monitoring system, a fingerprick for a drop of capillary blood and some minutes to get the INR result for further anticoagulation monitoring. We present a case of tricuspid prosthetic valve thrombosis due to inadequate anticoagulation because of an error in the point-of-care device. (C) 2010 Published by European Association for Cardio-Thoracic Surgery. All rights reserved.
引用
收藏
页码:123 / 124
页数:2
相关论文
共 6 条
  • [1] Prosthetic valve thrombosis:: Twenty-year experience at the Montreal Heart Institute
    Dürrleman, N
    Pellerin, M
    Bouchard, D
    Hébert, Y
    Cartier, R
    Perrault, LP
    Basmadjian, A
    Carrier, M
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2004, 127 (05) : 1388 - 1392
  • [2] Role of cine-fluoroscopy, transthoracic, and transesophageal echocardiography in patients with suspected prosthetic heart valve thrombosis
    Montorsi, P
    De Bernardi, F
    Muratori, M
    Cavoretto, D
    Pepi, M
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2000, 85 (01) : 58 - 64
  • [3] Biological or mechanical prostheses in tricuspid position? A meta-analysis of intra-institutional results
    Rizzoli, G
    Vendramin, I
    Nesseris, G
    Bottio, T
    Guglielmi, C
    Schiavon, L
    [J]. ANNALS OF THORACIC SURGERY, 2004, 77 (05) : 1607 - 1614
  • [4] Fibrinolysis of mechanical prosthetic valve thrombosis -: A single-center study of 127 cases
    Roudaut, R
    Lafitte, S
    Roudaut, MF
    Courtault, C
    Perron, JM
    Jaïs, C
    Pillois, X
    Coste, P
    DeMaria, A
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (04) : 653 - 658
  • [5] Accuracy and clinical usefulness of the CoaguChek S and XS Point of Care devices when starting warfarin in a hospital outreach setting
    Sobieraj-Teague, Magdalena
    Daniel, Dolly
    Farrelly, Barbara
    Coghlan, Douglas
    Gallus, Alexander
    [J]. THROMBOSIS RESEARCH, 2009, 123 (06) : 909 - 913
  • [6] Guidelines on the management of valvular heart disease - The Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology
    Vahanian, Alec
    Baumgartner, Helmut
    Bax, Jeroen
    Butchart, Eric
    Dion, Robert
    Filippatos, Gerasimos
    Flachskampf, Frank
    Hall, Roger
    Iung, Bernard
    Kasprzak, Jaroslaw
    Nataf, Patrick
    Tornos, Pilar
    Torracca, Lucia
    Wenink, Arnold
    Priori, Silvia G.
    Blanc, Jean-Jacques
    Budaj, Andrzej
    Camm, John
    Dean, Veronica
    Deckers, Jaap
    Dickstein, Kenneth
    Lekakis, John
    McGregor, Keith
    Metra, Marco
    Morais, Joao
    Osterspey, Ady
    Tamargo, Juan
    Luis Zamorano, Jose
    Angelini, Annalisa
    Antunes, Manuel
    Garcia Fernandez, Miguel Angel
    Gohlke-Baerwolf, Christa
    Habib, Gilbert
    McMurray, John
    Otto, Catherine
    Pierard, Luc
    Pomar, Jose L.
    Prendergast, Bernard
    Rosenhek, Raphael
    Uva, Miguel Sousa
    Tamargo, Juan
    [J]. EUROPEAN HEART JOURNAL, 2007, 28 (02) : 230 - 268