Patient-specific hip geometry has greater effect on THA wear than femoral head size

被引:4
作者
Daniel, Matej [1 ]
Rijavec, Boris [2 ]
Dolinar, Drago [3 ]
Pokorny, David [4 ]
Iglic, Ales [5 ]
Kralj-Iglic, Veronika [2 ]
机构
[1] Czech Tech Univ, Fac Mech Engn, Prague, Czech Republic
[2] Univ Ljubljana, Fac Hlth Sci, Lab Clin Biophys, Ljubljana 61000, Slovenia
[3] Univ Med Ctr Ljubljana, Dept Orthopaed Surg, Ljubljana, Slovenia
[4] Charles Univ Prague, Fac Med 1, Prague, Czech Republic
[5] Univ Ljubljana, Fac Elect Engn, Biophys Lab, Ljubljana 61000, Slovenia
关键词
UHMWPE; Linear penetration; Contact pressure; Design; CROSS-LINKED POLYETHYLENE; CEMENTLESS ACETABULAR COMPONENTS; BEARING SURFACES; FOLLOW-UP; CONVENTIONAL POLYETHYLENE; PENETRATION RATES; CONTACT MECHANICS; ARTHROPLASTY; REPLACEMENT; PREDICTION;
D O I
10.1016/j.jbiomech.2016.10.030
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In vivo linear penetration in total hip arthroplasty (THA) exhibits similar values for 28 mm and 32 mm femoral head diameter with considerable variations between and within the studies. It indicates factors other than femoral head diameter influence polyethylene wear. This study is intended to test the effect of patient's individual geometry of musculoskeletal system, acetabular cup orientation, and radius of femoral head on wear. Variation in patient's musculoskeletal geometry and acetabular cup placement is evaluated in two groups of patients implanted with 28 mm and 32 mm THA heads. Linear wear rate estimated by mathematical model is 0.165-0.185 mm/year and 0.157-0.205 mm/year for 28 and 32 mm THA heads, respectively. Simulations show little influence femoral head size has on the estimated annual wear rate. Predicted annual linear wear depends mostly on the abduction angle of the acetabular cup and individual geometry of the musculoskeletal system of the hip, with the latter having the greatest affect on variation in linear wear rate. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:3996 / 4001
页数:6
相关论文
共 60 条
[1]   Quantification of the effect of cross-shear and applied nominal contact pressure on the wear of moderately cross-linked polyethylene [J].
Abdelgaied, Abdellatif ;
Brockett, Claire L. ;
Liu, Feng ;
Jennings, Louise M. ;
Fisher, John ;
Jin, Zhongmin .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (H1) :18-26
[2]   Effect of Femoral Head Size on Metal-on-HXLPE Hip Arthroplasty Outcome in a Combined Analysis of Six National and Regional Registries [J].
Allepuz, Alex ;
Havelin, Leif ;
Barber, Thomas ;
Sedrakyan, Art ;
Graves, Stephen ;
Bordini, Barbara ;
Hoeffel, Daniel ;
Cafri, Guy ;
Paxton, Elizabeth .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A :12-18
[3]   Fifteen-Year to Twenty-Year Results of Cementless Harris-Galante Porous Femoral and Harris-Galante Porous I and II Acetabular Components [J].
Anseth, Scott D. ;
Pulido, Pamela A. ;
Adelson, Wendy S. ;
Patil, Shantanu ;
Sandwell, Julie C. ;
Colwell, Clifford W., Jr. .
JOURNAL OF ARTHROPLASTY, 2010, 25 (05) :687-691
[4]   Role of polyethylene particles in peri-prosthetic osteolysis: A review [J].
Atkins, Gerald J. ;
Haynes, David R. ;
Howie, Donald W. ;
Findlay, David M. .
WORLD JOURNAL OF ORTHOPEDICS, 2011, 2 (10) :93-101
[5]  
Bankston A B, 1993, J Arthroplasty, V8, P315
[6]  
BANKSTON AB, 1995, CLIN ORTHOP RELAT R, P7
[7]   Number of pedometer-assessed steps taken per day by adults: A descriptive meta-analysis [J].
Bohannon, Richard W. .
PHYSICAL THERAPY, 2007, 87 (12) :1642-1650
[8]  
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[9]   What design factors influence wear behavior at the bearing surfaces in total joint replacements? [J].
Brown, Thomas D. ;
Bartel, Donald L. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2008, 16 :S101-S106
[10]   A Double-Blind, Prospective, Randomized Controlled Trial Comparing Highly Cross-Linked and Conventional Polyethylene in Primary Total Hip Arthroplasty [J].
Calvert, George T. ;
Devane, Peter A. ;
Fielden, Jan ;
Adams, Kathryn ;
Horne, J. Geoffrey .
JOURNAL OF ARTHROPLASTY, 2009, 24 (04) :505-510