Test method for empirically determining inertial properties of manual wheelchairs

被引:16
作者
Eicholtz, Matthew R. [1 ,2 ]
Caspall, Jayme J. [1 ,2 ]
Dao, Phuc V. [1 ,2 ]
Sprigle, Stephen [2 ,3 ]
Al Ferri [1 ]
机构
[1] Georgia Inst Technol, Dept Mech Engn, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, Coll Architecture, Rehabil Engn & Appl Res Lab, Atlanta, GA 30332 USA
[3] Georgia Inst Technol, Dept Appl Physiol, Atlanta, GA 30332 USA
关键词
center of mass; free vibration; iMachine; mass; mathematical models; mechanical design; moment of inertia; rigid-body dynamics; simple harmonic oscillator; wheelchair testing;
D O I
10.1682/JRRD.2011.03.0045
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
The iMachine is a spring-loaded turntable used to measure inertial properties of irregularly shaped rigid bodies, specifically manual wheelchairs. We used a Newton-Euler approach to calculate wheelchair mass and center of mass (CM) location from static force measurements using load cells. We determined the moment of inertia about the vertical axis from the natural frequency of the system in simple harmonic motion. The device was calibrated to eliminate the effects of platform components on measurement error. For objects with known inertial properties, the average relative error of the mass and the CM coordinates (x and y) were 0.76%, 0.89%, and 1.99%, respectively. The resolution of the moment of inertia calculation depends on the ratio of test piece inertia to system inertia, such that the higher the ratio, the more accurate the measurements. We conducted a Gage Repeatability and Reproducibility (Gage R&R) test using three manual wheelchairs measured three times by three operators; the results showed that over 90% of the variance in inertia was caused by differences in the wheelchairs being measured. Gage R&R analysis indicated that measurement system operation was acceptable using criteria from the Automobile Industry Action Group for both inertia and mass measurements.
引用
收藏
页码:51 / 62
页数:12
相关论文
共 15 条
[1]   Further developments on the estimation of rigid body properties from experimental data [J].
Almeida, Raquel A. B. ;
Urgueira, Antonio P. V. ;
Maia, Nuno M. M. .
MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2010, 24 (05) :1391-1408
[2]  
Boynton Richard S, 1972, United States patent US 3,693,413, Patent No. 3693413
[3]  
Carpenter David A, 1979, United States patent US 4,161,876., Patent No. 4161876
[4]  
DaimlerChrysler Ford Motor Company General Motors Corporation., 2002, MEAS SYST AN REF MAN
[5]   Analysis of driving backward in an electric-powered wheelchair [J].
Ding, D ;
Cooper, RA ;
Guo, SF ;
Corfman, TA .
IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 2004, 12 (06) :934-943
[6]   Error Analysis in Trifilar Inertia Measurements [J].
du Bois, J. L. ;
Lieven, N. A. J. ;
Adhikari, S. .
EXPERIMENTAL MECHANICS, 2009, 49 (04) :533-540
[7]  
Frigo M, 1998, INT CONF ACOUST SPEE, P1381, DOI 10.1109/ICASSP.1998.681704
[8]   Measuring the moment of inertia of the human body by a rotating platform method [J].
Griffiths, IW ;
Watkins, J ;
Sharpe, D .
AMERICAN JOURNAL OF PHYSICS, 2005, 73 (01) :85-93
[9]  
Kauzlarich J J, 1987, J Rehabil Res Dev, V24, P67
[10]  
Mastinu Gianpiero, 2007, United States patent US 7,278,295, Patent No. 7278295