Influence of contaminants in the stem-ball interface on the static fracture load of ceramic hip joint ball heads

被引:28
作者
Weisse, B. [1 ]
Affolter, C. [1 ]
Stutz, A. [1 ]
Terrasi, G. P. [1 ]
Koebel, S. [2 ]
Weber, W. [2 ]
机构
[1] Empa Swiss Fed Labs Mat Testing & Res, Lab Mech Syst Engn, CH-8600 Dubendorf, Switzerland
[2] Metoxit AG, Thayngen, Switzerland
关键词
hip; ceramic; fracture load; contaminant; taper; entrapped debris;
D O I
10.1243/09544119JEIM374
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The probability of in-vivo failure of ceramic hip joint implants is very low (0.05-0.004 per cent). Besides material flaws and overloading, improper handling during implantation may induce fractures of the ceramic ball head in the long term. This study focuses on the influence of contaminants located in the stern-ball interface and on the use of damaged metal tapers on the strength of ceramic ball heads. Mechanical tests on alumina ball heads according to the standard ISO 7206-1.0 were performed to identify their effect on the static fracture load. A decrease of up to 90 per cent with respect to the reference static fracture load was found when contaminants Such as bone chips, soft tissue, or blood were present. Reductions of 57 per cent and 27 per cent were observed for deformed stem cross-sections (from circular to elliptical) and for flattened stems respectively, making deformed stems another influential parameter. Since any alteration of the interface between the metal taper and the ceramic ball head yields a non-uniform load introduction and hence results in stress concentrations, its presence has to he avoided.
引用
收藏
页码:829 / 835
页数:7
相关论文
共 23 条
[1]  
[Anonymous], BONE IMPLANT INTERFA
[2]  
[Anonymous], 7206102003 ISO
[3]   Hip joint contact forces during stumbling [J].
Bergmann, G ;
Graichen, F ;
Rohlmann, A .
LANGENBECKS ARCHIVES OF SURGERY, 2004, 389 (01) :53-59
[4]  
BLACK J, 2005, P 10 INT BIOLOX S, P3
[5]   20 Years Experience of Zirconia Total Hip Replacements [J].
Clarke, I. C. ;
Green, D. D. ;
Pezzotti, G. ;
Donaldson, D. .
BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2005, :67-78
[6]  
DIETRICH M, 2004, CERAMIC IMPLANTS JOI, P1
[7]  
Elke R., 2002, BIOCERAMICS JOINT AR, P1
[8]   Ceramic Component Fracture: Trends and Recommendations with Modern Components based on improved Reporting Methods [J].
Garino, J. P. .
BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2005, :157-168
[9]   Fracture of an alumina ceramic head in total hip arthroplasty [J].
Higuchi, F ;
Shiba, N ;
Inoue, A ;
Wakebe, I .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :851-854
[10]   Ceramic head failure [J].
Krikler, S ;
Schatzker, J .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :860-862