Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom

被引:51
作者
Bamba, J. [1 ]
Araki, K. [1 ]
Endo, A. [1 ]
Okano, T. [1 ]
机构
[1] Showa Univ, Sch Dent, Dept Oral Diagnost Sci, Div Radiol, Tokyo 1458515, Japan
关键词
cone beam CT; image quality; quality assurance; Radiographic phantom; COMPUTED-TOMOGRAPHY SCANNERS;
D O I
10.1259/dmfr.20120445
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The SEDENTEXCT Project proposed quality assurance (QA) methods and introduced a QA image quality phantom. A new prototype was recently introduced that may be improved according to previous reports. The purpose of this study is to evaluate image quality in various protocols of three cone beam CT (CBCT) machines using the proposed QA phantom. Methods: Using three CBCT machines, nine image quality parameters, including image homogeneity (noise), uniformity, geometrical distortion, pixel intensity value, contrast resolution, spatial resolution [line pair (LP) chart, point spread function (PSF) and modulation transfer function (MTF)] and metal artefacts, were evaluated using a QA phantom proposed by SEDENTEXCT. Exposure parameters, slice thickness and field of view position changed variously, and the number of total protocols was 22. Results: Many protocols showed, a uniform gray value distribution except in the minimum slice thickness image acquired using 3D Accuitomo 80 (Morita, Kyoto, Japan) and Veraviewepocs 3Df (Morita). Noise levels differed among the protocols. There was no geometric distortion, and the pixel intensity values were correlated with the CT value. Low contrast resolution differed among the protocols, but high contrast resolution performed well in all. Many protocols showed that the maximum line pair was larger than 1 LP mm(-1) but smaller than 3 LP mm(-1) PSF and MTF did not correlate well with the pixel size. The measured metal artefact areas varied for each device. Conclusions: We studied the image quality of three CBCT machines using the SEDENTEXCT phantom. Image quality varied with exposure protocols and machines.
引用
收藏
页数:13
相关论文
共 20 条
[1]   The effect of surrounding conditions on pixel value of cone beam computed tomography [J].
Araki, Kazuyuki ;
Okano, Tomohiro .
CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (08) :862-865
[2]   Quantitative analysis of metallic artifacts caused by dental metals: comparison of cone-beam and multi-detector row CT scanners [J].
Chindasombatjaroen, Jira ;
Kakimoto, Naoya ;
Murakami, Shumei ;
Maeda, Yoshinobu ;
Furukawa, Souhei .
ORAL RADIOLOGY, 2011, 27 (02) :114-120
[3]  
European Commission, 2004, Radiation protection No.136. European guidelines in dental practice
[4]  
International Electrotechnical Commission, 2012, 606012442009A1 IEC
[5]   Cone-beam computed tomography with a flat-panel imager: Initial performance characterization [J].
Jaffray, DA ;
Siewerdsen, JH .
MEDICAL PHYSICS, 2000, 27 (06) :1311-1323
[6]   A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT) Part I. On subjective image quality [J].
Liang, Xin ;
Jacobs, Reinhilde ;
Hassan, Bassam ;
Li, Limin ;
Pauwels, Ruben ;
Corpas, Livia ;
Souza, Paulo Couto ;
Martens, Wendy ;
Shahbazian, Maryam ;
Alonso, Arie ;
Lambrichts, Ivo .
EUROPEAN JOURNAL OF RADIOLOGY, 2010, 75 (02) :265-269
[7]   Image quality vs radiation dose of four cone beam computed tomography scanners [J].
Loubele, M. ;
Jacobs, R. ;
Maes, F. ;
Denis, K. ;
White, S. ;
Coudyzer, W. ;
Lambrichts, I. ;
van Steenberghe, D. ;
Suetens, P. .
DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (06) :309-318
[8]   Comparative localized linear accuracy of small-field cone-beam CT and multislice CT for alveolar bone measurements [J].
Loubele, M. ;
Van Assche, N. ;
Carpentier, K. ;
Maes, F. ;
Jacobs, R. ;
van Steenberghe, D. ;
Suetens, P. .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2008, 105 (04) :512-518
[9]   Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications [J].
Loubele, M. ;
Bogaerts, R. ;
Van Dijck, E. ;
Pauwels, R. ;
Vanheusden, S. ;
Suetens, P. ;
Marchal, G. ;
Sanderink, G. ;
Jacobs, R. .
EUROPEAN JOURNAL OF RADIOLOGY, 2009, 71 (03) :461-468
[10]   Assessment of bone segmentation quality of cone-beam CT versus multislice spiral CT: a pilot study [J].
Loubele, Miet ;
Maes, Frederik ;
Schutyser, Filip ;
Marchal, Guy ;
Jacobs, Reinhilde ;
Suetens, Paul .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2006, 102 (02) :225-234