Comparison between intermittent and continuous measurement of cardiac output after acute normovolemic hemodilution in pigs

被引:4
作者
Otsuki, Denise A. [1 ]
Cardieri, Fernando A. [1 ]
Schmidt, Andre P. [1 ]
Brudniewski, Marcio [1 ]
Fantoni, Denise T. [1 ]
Auler, Jose O. C., Jr. [1 ]
机构
[1] Univ Sao Paulo, Sch Med, Dept Anesthesia, LIMOS, Sao Paulo, Brazil
关键词
normovolemic hemodilution; continuous cardiac output; mixed venous oxygen saturation; volumetric pulmonary artery catheter;
D O I
10.1111/j.1525-1594.2006.00242.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Continuous cardiac output (CO) and mixed venous oxygen saturation (SvO(2)) determined through the pulmonary artery catheter may be helpful in monitoring hemodynamic conditions in critically ill patients. This study aimed to evaluate CO and SvO(2) in a model of acute normovolemic hemodilution (ANH), analyzing the accuracy of the continuous versus intermittent method for CO and SvO(2) measurement in pigs. Twenty-three pigs were enrolled to three groups: control, ANH with 6% hydroxyethyl starch (HES), or ANH with lactated Ringer's (LR) solution. After hemodilution, we showed that SvO(2) was reduced in both groups, mainly in LR animals (P < 0.05). Regarding the evaluation of CO, we showed an increase in both groups submitted to ANH (P < 0.05). Through Bland-Altman analysis, we showed that the continuous CO catheter presented lower values than the intermittent method after hemodilution, mainly with HES (P < 0.001), and there was no difference in the measurement of SvO(2). The ANH promoted a decrease in SvO(2) and an increase in CO values, mainly in animals submitted to hemodilution with HES. The use of continuous and intermittent (laboratory) measurement of SvO(2) showed clinical applicability and good agreement, an effect not reproduced by the CO measurement. New studies are needed to further investigate the agreement between the continuous and intermittent methods for the measurement of CO in adverse hemodynamic conditions such as ANH.
引用
收藏
页码:458 / 466
页数:9
相关论文
共 41 条
[1]  
Am Soc Anesthesiologists Task Forc, 2003, ANESTHESIOLOGY, V99, P988
[2]   EFFECTS OF THE RESPIRATORY CYCLE ON CARDIAC-OUTPUT MEASUREMENTS - REPRODUCIBILITY OF DATA ENHANCED BY TIMING THE THERMODILUTION INJECTIONS IN DOGS [J].
ARMENGOL, J ;
MAN, GCW ;
BALSYS, AJ ;
WELLS, AL .
CRITICAL CARE MEDICINE, 1981, 9 (12) :852-854
[3]  
Auler J O, 2001, Minerva Anestesiol, V67, P355
[4]  
BAELE PL, 1982, ANESTH ANALG, V61, P513
[5]   Transesophageal echocardiographic hemodynamic monitoring during preoperative acute normovolemic hemodilution [J].
Bak, Z ;
Abildgård, L ;
Lisander, B ;
Janerot-Sjöberg, B .
ANESTHESIOLOGY, 2000, 92 (05) :1250-1256
[6]  
Bakker Ron, 2004, Eur J Cardiovasc Nurs, V3, P165, DOI 10.1016/j.ejcnurse.2004.02.003
[7]   Haemodynamic effects of moderate normovolaemic haemodilution in conscious and anaesthetized patients [J].
Biboulet, P ;
Capdevila, X ;
Benetreau, D ;
Aubas, P ;
dAthis, F ;
duCailar, J .
BRITISH JOURNAL OF ANAESTHESIA, 1996, 76 (01) :81-84
[8]  
Boldt J, 2002, CRIT CARE, V6, P52
[9]  
BOLDT J, 1994, CRIT CARE MED, V22, P1913
[10]  
Bottiger BW, 1996, INTENS CARE MED, V22, P312