Measuring corneal hysteresis: threshold estimation of the waveform score from the Ocular Response Analyzer

被引:15
作者
Ayala, Marcelo [1 ]
Chen, Enping [1 ]
机构
[1] Karolinska Inst, St Erik Eye Hosp, Glaucoma Dept, SE-11282 Stockholm, Sweden
关键词
Waveformscore; Ocular response analyzer; Threshold; INTRAOCULAR-PRESSURE; GLAUCOMA;
D O I
10.1007/s00417-012-2053-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The aim of this study was to determine a threshold waveform score (WS) for the best score value (BSV) in the Ocular Response Analyzer (ORA). Retrospective study. One hundred and thirty-three healthy adults were recruited. Measurements were done with the ORA 2.04. Two hundred and sixty-six eyes were analyzed. Mean age was 56.49 +/- 15.97 years. The mean waveform score of the BSV was 7.39 +/- 1.32. The waveform scores ranged from 2.8 to 9.7. Kolmogorov-Smirnov test for normality was significant (p a parts per thousand currency sign 0.0001). Linear regression showed a significant positive correlation between IOPg (measured with the ORA) and IOP measured with Goldmann applanation tonometry (p a parts per thousand currency sign 0.0001), as well as significant negative correlation between the difference IOPg-IOP Goldmann and waveform score of the BSV values. Threshold estimation considering 95 % confidence interval was 7.23. Meanwhile, threshold estimation considering the difference IOPg-IOP Goldmann, for 3 mmHg, was 6.7. When using the ORA device, we recommend that clinicians try to obtain several waveform score measurements of 7 or above. Waveform scores lower than 7 may render less reliable results.
引用
收藏
页码:1803 / 1806
页数:4
相关论文
共 9 条
[1]   Corneal hysteresis measured with the Ocular Response Analyzer® in normal and glaucomatous eyes [J].
Abitbol, Olivia ;
Bouden, Jihene ;
Doan, Serge ;
Hoang-Xuan, Thanh ;
Gatinel, Damien .
ACTA OPHTHALMOLOGICA, 2010, 88 (01) :116-119
[2]   Corneal Hysteresis in Normal Subjects and in Patients with Primary Open-Angle Glaucoma and Pseudoexfoliation Glaucoma [J].
Ayala, Marcelo .
OPHTHALMIC RESEARCH, 2011, 46 (04) :187-191
[3]   Lower corneal hysteresis in glaucoma patients with acquired pit of the optic nerve (APON) [J].
Bochmann, Frank ;
Ang, Ghee S. ;
Azuara-Blanco, Augusto .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (05) :735-738
[4]   Evaluation of agreement between intraocular pressure measurements using Goldmann applanation tonometry and Goldmann correlated intraocular pressure by Reichert's ocular response analyser [J].
Ehrlich, J. R. ;
Haseltine, S. ;
Shimmyo, M. ;
Radcliffe, N. M. .
EYE, 2010, 24 (10) :1555-1560
[5]   The Usefulness of Waveform Score from the Ocular Response Analyzer [J].
Lam, Andrew K. C. ;
Chen, Davie ;
Tse, Jimmy .
OPTOMETRY AND VISION SCIENCE, 2010, 87 (03) :195-199
[6]   Determining in vivo biomechanical properties of the cornea with an ocular response analyzer [J].
Luce, DA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :156-162
[7]  
Mandalos A, 2011, J GLAUCOMA
[8]   Ocular Response Analyzer in subjects with and without glaucoma [J].
Sullivan-Mee, Michael ;
Billingsley, Shavon C. ;
Patel, Amita D. ;
Halverson, Kathy D. ;
Alldredge, Brooks R. ;
Qualls, Clifford .
OPTOMETRY AND VISION SCIENCE, 2008, 85 (06) :463-470
[9]   Correlation between corneal hysteresis intraocular pressure, and corneal central pachymetry [J].
Touboul, David ;
Roberts, Cynthia ;
Kerautret, Julien ;
Garra, Caroline ;
Maurice-Tison, Sylvie ;
Saubusse, Elodie ;
Colin, Joseph .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (04) :616-622