Derived 12-lead electrocardiogram in the assessment of ST-segment deviation and cardiac rhythm

被引:22
作者
Chantad, D
Krittayaphong, R [1 ]
Komoltri, C
机构
[1] Mahidol Univ, Siriaj Hosp, Dept Med, Div Cardiol, Bangkok 10700, Thailand
[2] Siriaj Hosp, Dept Res Dev, Div Clin Epidemiol, Bangkok, Thailand
关键词
EASI lead system; ST-segment deviation; cardiac rhythms; standard ECG;
D O I
10.1016/j.jelectrocard.2005.01.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: There are little data on the validation of 12-lead electrocardiogram (ECG) derived by the EASI lead system used for continuous monitoring in critical care settings. Objective: The objectives of this study were to determine the accuracy of 12-lead ECG derived by the EAS1 lead system in the detection of ST-segment deviation and cardiac rhythm compared with the standard 12-lead ECG. Methods: All patients admitted to the coronary care unit were studied. Kappa statistics was used to calculate the agreement between both ECG systems in the determination of cardiac rhythm and premature ventricular complex morphology, ST-segment analysis was performed in patients with acute coronary syndromes. Pearson correlation was used to correlate the ST-segment deviation between both techniques. The sensitivity and specificity of the determination of significant ST-segment deviation by the EASI lead system were calculated. Results: There were a total of 282 patients enrolled in this study. There was a complete agreement in the interpretation of cardiac rhythm between the 2 methods (kappa = 1). Analysis of ST-segment deviation of 12-lead ECG also showed a significant Correlation (correlation coefficient varied from 0.62 in lead I to 0.823 in lead aVF with a P value of < .001 in all leads) between the 2 methods with very high sensitivity and specificity in the detection of significant ST-segment elevation and depression. Conclusion: The 12-lead ECG derived by the EASI lead system is an accurate and reliable information for the assessment of ST-segment deviation and cardiac rhythm in critically ill patients. (c) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 12 条
[1]   MORPHOLOGIC CHARACTERISTICS OF NONSUSTAINED VENTRICULAR-TACHYCARDIA DETECTED DURING HOLTER MONITORING ASSOCIATED WITH ATHEROSCLEROTIC CORONARY-ARTERY DISEASE [J].
DENES, P .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (01) :57-62
[2]   THE IMPORTANCE OF DERIVED 12-LEAD ELECTROCARDIOGRAPHY IN THE INTERPRETATION OF ARRHYTHMIAS DETECTED BY HOLTER RECORDING [J].
DENES, P .
AMERICAN HEART JOURNAL, 1992, 124 (04) :905-911
[3]   DERIVING THE 12-LEAD ELECTROCARDIOGRAM FROM 4 (EASI) ELECTRODES [J].
DOWER, GE ;
YAKUSH, A ;
NAZZAL, SB ;
JUTZY, RV ;
RUIZ, CE .
JOURNAL OF ELECTROCARDIOLOGY, 1988, 21 :S182-S187
[4]  
Drew B J, 1996, Am J Crit Care, V5, P198
[5]   Comparison of standard and derived 12-lead electrocardiograms for diagnosis of coronary angioplasty-induced myocardial ischemia [J].
Drew, BJ ;
Adams, MG ;
Pelter, MM ;
Wung, SF ;
Caldwell, MA .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (05) :639-644
[6]   COMPARISON OF A VECTORCARDIOGRAPHICALLY DERIVED 12-LEAD ELECTROCARDIOGRAM WITH THE CONVENTIONAL ELECTROCARDIOGRAM DURING WIDE QRS COMPLEX TACHYCARDIA, AND ITS POTENTIAL APPLICATION FOR CONTINUOUS BEDSIDE MONITORING [J].
DREW, BJ ;
SCHEINMAN, MM ;
EVANS, GT .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (06) :612-618
[7]   Accuracy of the EASI 12-lead electrocardiogram compared to the standard 12-lead electrocardiogram for diagnosing multiple cardiac abnormalities [J].
Drew, BJ ;
Pelter, MM ;
Wung, SF ;
Adams, MG ;
Taylor, C ;
Evans, GT ;
Foster, E .
JOURNAL OF ELECTROCARDIOLOGY, 1999, 32 :38-47
[8]   Improved EASI coefficients:: Their derivation, values, and performance [J].
Feild, DQ ;
Feldman, CL ;
Horácek, BM .
JOURNAL OF ELECTROCARDIOLOGY, 2002, 35 :23-33
[9]   Comparison of electrocardiograms recorded with standard leads and derived from the vectorcardiographic Frank leads in high-risk patients [J].
Gannedahl, PE ;
Ljungqvist, OH ;
Lundin, P ;
Edner, M .
INTENSIVE CARE MEDICINE, 1997, 23 (10) :1049-1055
[10]   Optimal electrocardiographic leads for detecting acute myocardial ischemia [J].
Horácek, BM ;
Warren, JW ;
Penney, CJ ;
MacLeod, RS ;
Title, LM ;
Gardner, MJ ;
Feldman, CL .
JOURNAL OF ELECTROCARDIOLOGY, 2001, 34 :97-111