Canadian Registry of Electronic Device Outcomes (CREDO) : The Abbott ICD Premature Battery Depletion Advisory, a Multicentre Cohort Study

被引:1
作者
Davis, Jason [1 ]
Thibault, Bernard [2 ]
Mangat, Iqwal [3 ]
Coutu, Benoit [4 ]
Bennett, Matthew [5 ]
Philippon, Francois [6 ]
Sandhu, Roopinder [7 ]
Sterns, Laurence [8 ]
Essebag, Vidal [9 ]
Nery, Pablo [10 ]
Wells, George [11 ]
Yee, Raymond [12 ]
Exner, Derek [13 ]
Krahn, Andrew [5 ]
Parkash, Ratika [1 ,14 ]
机构
[1] QEII Hlth Sci Ctr, Div Med Oncol, Halifax, NS, Canada
[2] Montreal Heart Inst, Div Cardiol, Montreal, PQ, Canada
[3] St Michaels Hosp, Div Cardiol, Toronto, ON, Canada
[4] Montreal Univ Hosp Ctr, Div Cardiol, Montreal, PQ, Canada
[5] Univ British Columbia, Div Cardiol, Vancouver, BC, Canada
[6] Hamilton Hlth Sci Ctr, Div Cardiol, Hamilton, ON, Canada
[7] Southlake Reg Hlth Ctr, Div Cardiol, Newmarket, ON, Canada
[8] Laval Univ, Div Cardiol, Quebec City, PQ, Canada
[9] Mazankowski Alberta Heart Inst, Div Cardiol, Edmonton, AB, Canada
[10] Univ Hlth Network, Div Cardiol, Toronto, ON, Canada
[11] Sherbrooke Univ Hosp Ctr, Div Cardiol, Sherbrooke, PQ, Canada
[12] Royal Jubilee Hosp, Div Cardiol, Victoria, BC, Canada
[13] McGill Univ, Div Cardiol, Montreal, PQ, Canada
[14] QEII Hlth Sci Ctr, Halifax Infirm Site,Room 2501B,1796 Summer St, Halifax, NS B3H 3A7, Canada
关键词
CARDIOVERTER-DEFIBRILLATOR REPLACEMENT; RECALLS; SOCIETY;
D O I
10.1016/j.cjco.2020.09.008
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Premature or rapid battery depletion may compromise the performance and reliability of an implantable cardioverter defi- brillator (ICD), potentially resulting in harm or death to patients. We sought to describe the outcomes and clinical management of devices included in the Abbott ICD Premature Battery Depletion Advisory, using data from a Canadian registry. Methods: This prospective observational study includes patients with an Abbott device subject to the advisory, from 9 centres in Canada. The incidence and outcomes related to device revision owing to premature battery depletion were identified and adjudicated by a committee.Results: There were 2678 patients enrolled with a device subject to the advisory. Devices were implanted between 2010 and 2017; follow-up time was 5.7 & PLUSMN; 0.7 years. Device revision occurred in 222 patients (8.3%). Revision for premature battery depletion occurred in 43 pa-tients (1.6%). Devices were revised at physician discretion on notice of the advisory in 16 patients (0.6%), and at patient request in 5 patients (0.2%). A total of 63 (2.4%) devices reached routine end of battery life. A further 95 (3.5%) patients underwent revision for other reasons. There were no reported major complications or adverse events with device revision owing to the advisory. There were no deaths attributed to premature battery depletion. Conclusions: The rate of premature battery depletion associated with the Abbott ICD Premature Battery Depletion Advisory is low. There were no clinically adverse events identified that were associated with the battery performance of devices under advisory.
引用
收藏
页码:48 / 53
页数:6
相关论文
共 17 条
[1]  
Abbott, BATT PERF AL WHIT PA
[2]  
Abbott, 2018, PROD PERF REP
[3]  
Abbott, 2018, ICD PREM BATT DEPL A
[4]  
Abbott, NEW BATT PERF MAN TO
[5]  
Al-Khatib SM, 2018, CIRCULATION, V138, pE272, DOI [10.1161/CIR.0000000000000549, 10.1161/CIR.0000000000000548]
[6]   Recommendations from the Heart Rhythm Society Task Force on device performance policies and guidelines [J].
Carlson, Mark D. ;
Wilkoff, Bruce L. ;
Maisel, William H. ;
Ellenbogen, Kenneth A. ;
Saxon, Leslie A. ;
Prystowsky, Eric N. ;
Alpert, Joseph S. ;
Cain, Michael E. ;
Ching, Elizabeth A. ;
Curtis, Anne B. ;
Davies, D. Wyn ;
Hammill, Stephen C. ;
Hauser, Robert G. ;
Lampert, Rachel ;
Zipes, Douglas P. .
HEART RHYTHM, 2006, 3 (10) :1250-1273
[7]   Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories [J].
Gould, PA ;
Krahn, AD .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (16) :1907-1911
[8]   Outcome of advisory implantable cardioverter-defibrillator replacement: One-year follow-up [J].
Gould, Paul A. ;
Gula, Lorne J. ;
Champagne, Jean ;
Healyey, Jeffrey S. ;
Cameron, Doug ;
Simpson, Christophers ;
Thibault, Bernard ;
Pinter, Arnold ;
Tung, Stanley ;
Sterns, Laurence ;
Birnie, David ;
Exner, Derek ;
Parkash, Ratika ;
Skanes, Allan C. ;
Yee, Raymond ;
Klein, George J. ;
Krahn, Andrew D. .
HEART RHYTHM, 2008, 5 (12) :1675-1681
[9]   Survival rates as a guide to implanted cardioverter-defibrillator replacement strategies for device recalls - Adding statistical insight to clinical intuition [J].
Gula, Lorne J. ;
Massel, David ;
Krahn, Andrew D. ;
Skanes, Allan C. ;
Yee, Raymond ;
Klein, George J. .
AMERICAN HEART JOURNAL, 2007, 153 (02) :253-259
[10]   Pacemaker and ICD generator reliability - Meta-analysis of device registries [J].
Maisel, WH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (16) :1929-1934