Our aim was to investigate the changes in bone mineral density (BMD) of acetabulum and proximal femur after total hip resurfacing arthroplasty. A comparative study was carried out on 51 hips in 48 patients. Group A consisted of 25 patients (26 hips) who had undergone total hip resurfacing and group B consisted of 23 patients (25 hips) who had had large-diameter metal-on-metal total hip arthroplasty (THA). BMDs around the acetabulum and proximal femur were measured using dual-energy x-ray absorptiometry (DEXA) at 2 weeks, 6 months, 1 year and annually thereafter during the 3 years after surgery. At final follow-up, the acetabular net mean BMD decreased by 11% in group A and 10% in group B with no differences between two groups (P = .35). For the femoral side, in Gruen zone 1, the mean BMD increased by 4% in group A, whereas it decreased by 11% in group B (P = .029). In Gruen zone 7, the mean BMD increased by 8% at the final follow-up in group A, whereas it decreased by 13% in group B (P = .02). In both groups the mean BMD increased by 3% in Gruen zones 3, 4, 5, and 6. Stress-related bone loss of the acetabulum was comparable for MOM THA and resurfacing devices, but proximal femoral bone density increased in the resurfacing group and decreased in the THA group. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:1811 / 1815
页数:5
相关论文
共 20 条
[1]  
Amstutz Harlan C, 2006, J Bone Joint Surg Am, V88 Suppl 1 Pt 2, P234, DOI 10.2106/JBJS.F.00273
[2]   Metal-on-metal hybrid surface arthroplasty:: Two to six-year follow-up study [J].
Amstutz, HC ;
Beaulé, PE ;
Dorey, FJ ;
Le Duff, MJ ;
Campbell, PA ;
Gruen, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :28-39
[3]   Serum cobalt levels after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
Gottsauner-Wolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (11) :2168-2173
[4]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[5]  
GRUEN TA, 1979, CLIN ORTHOP RELAT R, P17
[6]   Dual energy X-ray absorptiometry analysis of peri-prosthetic stress shielding in the Birmingham resurfacing hip replacement [J].
Harty, JA ;
Devitt, B ;
Harty, LC ;
Molloy, M ;
McGuinness, A .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (10) :693-695
[7]  
HUISKES R, 1987, ACTA ORTHOP SCAND, V58, P620
[8]   Changes in the bone mineral density in the acetabulum and proximal femur after cementless total hip replacement -: Alumina-on-alumina versus alumina-on-polyethylene articulation [J].
Kim, Y. -H. ;
Yoon, S. -H. ;
Kim, J. -S. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (02) :174-179
[9]   Preservation of the bone mineral density of the femur after surface replacement of the hip [J].
Kishida, Y ;
Sugano, N ;
Nishii, T ;
Miki, H ;
Yamaguchi, K ;
Yoshikawa, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (02) :185-189
[10]  
Levenston M E, 1993, J Arthroplasty, V8, P595, DOI 10.1016/0883-5403(93)90006-P