Neonatal chest radiography: Influence of standard clinical protocols and radiographic equipment on pathology visibility and radiation dose using a neonatal chest phantom

被引:5
作者
Al-Murshedi, S. [1 ]
Hogg, Peter [2 ]
England, A. [2 ]
机构
[1] Univ ALZahraa Women, Coll Hlth & Med Technol, Karbala, Iraq
[2] Univ Salford, Sch Hlth & Soc, Salford M6 6PU, Lancs, England
关键词
Pathology visibility; Image quality; Radiation dose; Dose optimisation; Neonate chest radiography; CONTRAST-DETAIL PHANTOM; IMAGE QUALITY; FLAT-PANEL; COMPUTED RADIOGRAPHY; AMORPHOUS-SILICON; DETECTOR; OPTIMIZATION; EXPOSURE; REDUCTION; SYSTEM;
D O I
10.1016/j.radi.2020.02.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Little is known about the variations in pathology visibility (PV) and their corresponding radiation dose values for neonatal chest radiography, between and within hospitals. Large variations in PV could influence the diagnostic outcome and the variations in radiation dose could affect the risk to patients. The aim of this study is to compare the PV and radiation dose for standard neonatal chest radiography protocols among a series of public hospitals. Methods: A Gammex 610 neonatal chest phantom was used to simulate the chest region of neonates. Radiographic acquisitions were conducted on 17 X-ray machines located in eight hospitals, utilising their current neonatal chest radiography protocols. Six qualified radiographers assessed PV visually using a relative visual grading analysis (VGA). Signal to noise ratios (SNR) and contrast to noise ratios (CNR) were measured as a measure of image quality (IQ). Incident air kerma (IAK) was measured using a solid-state dosimeter. Results: PV and radiation dose varied substantially between and within hospitals. For PV, the mean (range) VGA scores, between and within the hospitals, were 2.69 (2.00-3.50) and 2.73 (2.33-3.33), respectively. For IAK, the mean (range), between and within the hospitals, were 24.45 (8.11-49.94) mGy and 34.86 (22.26-49.94) mGy, respectively. Conclusion: Between and within participating hospitals there was wide variation in the visibility of simulated pathology and radiation dose (IAK). Implications for practice: X-ray units with lower PV and higher doses require optimisation of their standard clinical protocols. Institutions which can offer acceptable levels of PV but with lower radiation doses should help facilitate national optimisation processes. (C) 2020 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:282 / 287
页数:6
相关论文
共 44 条
[1]   Comparative analysis of radiation dose and low contrast detail detectability using routine paediatric chest radiography protocols [J].
Al-Murshedi, Sadeq ;
Hogg, Peter ;
Meijer, Annemieke ;
Erenstein, Hendrik ;
England, Andrew .
EUROPEAN JOURNAL OF RADIOLOGY, 2019, 113 :198-203
[2]   Relationship between body habitus and image quality and radiation dose in chest X-ray examinations: A phantom study [J].
Al-Murshedi, Sadeq ;
Hogg, Peter ;
England, Andrew .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 :65-71
[3]   A novel method for comparing radiation dose and image quality, between and within different x-ray units in a series of hospitals [J].
Al-Murshedi, Sadeq ;
Hogg, Peter ;
Lanca, Luis ;
England, Andrew .
JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (04) :1344-1358
[4]   Fact or fiction: An analysis of the 10 kVp 'rule' in computed radiography [J].
Allen, Elizabeth ;
Hogg, Peter ;
Ma, Wang Kei ;
Szczepura, Katy .
RADIOGRAPHY, 2013, 19 (03) :223-227
[5]  
[Anonymous], 1982, PROC SPIE, DOI DOI 10.1117/12.967674
[6]  
[Anonymous], 2014, ACR SPR PRACT PAR PE
[7]   Radiation dose quantities and risk in neonates in a special care baby unit [J].
Armpilia, CI ;
Fife, IAJ ;
Croasdale, PL .
BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (895) :590-595
[8]   Dose reduction in patients undergoing chest imaging: Digital amorphous silicon flat-panel detector radiography versus conventional film-screen radiography and phosphor-based computed radiography [J].
Bacher, K ;
Smeets, P ;
De Hauwere, A ;
Verstraete, K ;
Thierens, H .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (04) :923-929
[9]   Risk of cancer from diagnostic X-rays:: estimates for the UK and 14 other countries [J].
Berrington de González, A ;
Darby, S .
LANCET, 2004, 363 (9406) :345-351
[10]   Copper filtration in pediatric digital X-ray imaging: Its impact on image quality and dose [J].
Brosi P. ;
Stuessi A. ;
Verdun F.R. ;
Vock P. ;
Wolf R. .
Radiological Physics and Technology, 2011, 4 (2) :148-155