Microcatheter entrapment retrieval from Onyx embolization in brain arteriovenous malformations: A technical note

被引:9
作者
Vu, Phat D. [1 ]
Grigorian, Arthur A. [2 ]
机构
[1] Mercer Univ, Sch Med, Macon, GA 31207 USA
[2] Med Ctr Georgia, Decatur, GA USA
关键词
Onyx; embolization; microcatheter; complication; entrapment; EXPERIENCE;
D O I
10.1177/1591019915583226
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: Many techniques have been use for retrieval of an entrapped microcatheter during Onyx (eV3 Neurovascular) embolization of brain arteriovenous malformations (BAVMs). We report our technique that we term "pull-push-pull" that can be utilized as first management in retrieving the microcatheter. Method: We analyzed a total of 37 patients that underwent BAVM embolization with either Onyx 18 or 34 at our institution. Standard embolization techniques were utilized with the use of Marathon (eV3 Neurovascular) microcatheter. When difficulty in retrieving the microcatheter arose, we used the "pull-push-pull" technique. The technique comprises the eV3 protocol of retraction. In addition, the microcatheter is stretched causing the Onyx cast to stretch in its inner core, creating a more thorough cohesive property amongst the Onyx mixture. Then the microcatheter is pushed back and to its point of embolization origin. Afterwards, retraction of the microcatheter is enabled as it can be easily dislodged from the cast. Multiple attempts can be repeated as needed. Result and discussion: We had three patients that had difficulty with removal of microcatheter (8.1%). Utilization of the "pull-push-pull" technique was used on two of those patients. No neurological complication was observed with our technique. We believe the cohesive property of Onyx solution helps in the retrieval of the catheter by our method and technique. Conclusion: We believe the "pull-push-pull" can be utilized and be an additional technique before attempting other catheter retrieval techniques in Onyx BAVM embolization.
引用
收藏
页码:620 / 623
页数:4
相关论文
共 13 条
[1]   Surgical handling characteristics of an ethylene vinyl alcohol copolymer compared with N-butyl cyanoacrylate used for embolization of vessels in an arteriovenous malformation resection model in swine [J].
Akin, ED ;
Perkins, E ;
Ross, IB .
JOURNAL OF NEUROSURGERY, 2003, 98 (02) :366-370
[2]   Retrieving microcatheters from Onyx casts in a series of brain arteriovenous malformations: a technical report [J].
Alamri, A. ;
Hyodo, A. ;
Suzuki, K. ;
Tanaka, Y. ;
Uchida, T. ;
Takano, I. ;
Kowata, K. ;
Iwatate, K. ;
Suzuki, R. .
NEURORADIOLOGY, 2012, 54 (11) :1237-1240
[3]   Onyx®:: a unique neuroembolic agent [J].
Ayad, Michael ;
Eskioglu, Eric ;
Mericle, Robert A. .
EXPERT REVIEW OF MEDICAL DEVICES, 2006, 3 (06) :705-715
[4]   Embolization of arteriovenous malformations with onyx:: Clinicopathological experience in 23 patients [J].
Jahan, R ;
Murayama, Y ;
Gobin, YP ;
Duckwiler, GR ;
Vinters, HV ;
Viñuela, F .
NEUROSURGERY, 2001, 48 (05) :984-995
[5]   Curative embolization of cerebral arteriovenous malformations (AVMs) with Onyx in 101 patients [J].
Katsaridis, Vasilios ;
Papagiannaki, Chrysanthi ;
Aimar, Enrico .
NEURORADIOLOGY, 2008, 50 (07) :589-597
[6]  
Kelly ME, 2008, Neurosurgery, V63
[7]   Periprocedural Bleeding Complications of Brain AVM Embolization with Onyx [J].
Liu, L. ;
Jiang, C. ;
He, H. ;
Li, Y. ;
Wu, Z. .
INTERVENTIONAL NEURORADIOLOGY, 2010, 16 (01) :47-57
[8]   Multimodality treatment of brain arteriovenous malformations with microsurgery after embolization with Onyx: Single-center experience and technical nuances [J].
Natarajan, Sabareesh K. ;
Ghodke, Basavaraj ;
Britz, Gavin W. ;
Born, Donald E. ;
Sekhar, Laligam N. .
NEUROSURGERY, 2008, 62 (06) :1213-1225
[9]  
Newman C, 2012, J NEUROINTERV SURG, V4, P1
[10]   Embolization of Intracranial Arteriovenous Malformations with Ethylene-Vinyl Alcohol Copolymer (Onyx) [J].
Panagiotopoulos, V. ;
Gizewski, E. ;
Asgari, S. ;
Regel, J. ;
Forsting, M. ;
Wanke, I. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (01) :99-106