Usefulness of echo-guided cardiac resynchronization pacing in patients undergoing "Ablate and Pace" therapy for permanent atrial fibrillation and effects of heart rate regularization and left ventricular resynchronization

被引:6
作者
Brignole, Michele
Menozzi, Carlo [1 ]
Botto, Gian Luca [2 ]
Mont, Lluis [3 ]
Osca Asensi, Joaquin [4 ]
Garcia Medina, Dolores [5 ]
Oddone, Daniele
Navazio, Alessandro [1 ]
Luzi, Mario [2 ]
Iacopino, Saverio [6 ]
De Fabrizio, Giuseppe [7 ]
Proclemer, Alessandro [8 ]
Vardas, Panos [9 ]
机构
[1] Osped S Maria Nuova, Dept Cardiol, Reggio Emilia, Italy
[2] Osped S Anna Como, Dept Cardiol, Como, Italy
[3] Hosp Clin Barcelona, Dept Cardiol, Barcelona, Spain
[4] Hosp Univ La Fe, Dept Cardiol, Valencia, Spain
[5] Hosp Univ Virgen de Valme, Dept Cardiol, Seville, Spain
[6] Clin S Anna, Dept Cardiol, Catanzaro, Italy
[7] Osped Moscati, Dept Cardiol, Avellino, Italy
[8] Osped Misericordia, Dept Cardiol, Udine, Italy
[9] Herakl Univ Hosp, Dept Cardiol, Iraklion, Greece
关键词
D O I
10.1016/j.amjcard.2008.05.024
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
An acute comparative study of right ventricular (RV) pacing and echocardiographically guided cardiac resynchronization pacing (CRP) was performed in patients who underwent "ablate and pace" therapy for permanent atrial fibrillation. It was hypothesized that optimized CRP guided by tissue Doppler echocardiography would exert an additive beneficial hemodynamic effect to that of rate regularization achieved through atrioventricular junction ablation. An acute intrapatient comparison of echocardiographic parameters was performed between baseline preablation values and RV pacing and CRP (performed <24 hours after ablation) in 50 patients. Optimized CRP configuration was defined as the modality of pacing corresponding to that of the shortest intra-left ventricular (LV) delay among simultaneous biventricular pacing, sequential biventricular pacing, and single-chamber pacing. The intra-LV delay was defined as the difference between the longest and the shortest activation time in the six basal segments of the left ventricle. Compared with preablation measures, the ejection fraction increased by 10.8% during RV pacing (19% in patients with intra-LV delays <47.5 ms and 3% in those with intra-LV delays >47.5 ms). Compared with RV pacing, CRP caused a 9.2% increase in the ejection fraction, a 6.8% decrease in LV systolic diameter, and a 17.3% decrease in mitral regurgitation area; LV dyssynchrony was reduced from 52 +/- 27 to 21 +/- 12 ms. Similar results were observed in patients with and without depressed systolic function and in patients with and without left bundle branch block. In conclusion, rate regularization achieved through atrioventricular junction ablation and RV pacing provides a favorable hemodynamic effect that is inversely related to the level of LV dyssynchrony. Minimizing LV dyssynchrony by means of optimized CRP yields an additional important benefit. (C) 2008 Elsevier Inc. All rights reserved.
引用
收藏
页码:854 / 860
页数:7
相关论文
共 29 条
[1]  
[Anonymous], 1994, NOMENCLATURE CRITERI, V9th, P253
[2]   Doppler myocardial Imaging to evaluate the effectiveness of pacing sites in patients receiving biventricular pacing [J].
Ansalone, G ;
Giannantoni, P ;
Ricci, R ;
Trambaiolo, P ;
Fedele, F ;
Santini, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (03) :489-499
[3]   Left ventricular electromechanical delay in patients with heart failure and normal QRS duration and in patients with right and left bundle branch block [J].
Badano, Luigi P. ;
Gaddi, Oscar ;
Peraldo, Carlo ;
Lupi, Gabriele ;
Sitges, Marta ;
Parthenakis, Frangisko ;
Molteni, Santo ;
Pagliuca, Maria Rosaria ;
Sassone, Biagio ;
Di Stefano, Paola ;
De Santo, Tiziana ;
Menozzi, Carlo ;
Brignole, Michele .
EUROPACE, 2007, 9 (01) :41-47
[4]   Cardiac-resynchronization therapy in heart failure with narrow QRS complexes [J].
Beshai, John F. ;
Grimm, Richard A. ;
Nagueh, Sherif F. ;
Baker, James H., II ;
Beau, Scott L. ;
Greenberg, Steven M. ;
Pires, Luis A. ;
Tchou, Patrick J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (24) :2461-2471
[5]   Left ventricular resynchronization is mandatory for response to cardiac resynchronization therapy - Analysis in patients with echocardiographic evidence of left ventricular dyssynchrony at baseline [J].
Bleeker, Gabe B. ;
Mollema, Sjoerd A. ;
Holman, Eduard R. ;
Van De Veire, Nico ;
Ypenburg, Claudia ;
Boersma, Eric ;
van der Wall, Ernst E. ;
Schalij, Martin J. ;
Bax, Jeroen J. .
CIRCULATION, 2007, 116 (13) :1440-1448
[6]   Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacing [J].
Bordachar, P ;
Lafitte, S ;
Reuter, S ;
Sanders, P ;
Jaïs, P ;
Haïssaguerre, M ;
Roudaut, R ;
Garrigue, S ;
Clementy, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (11) :2157-2165
[7]   Cardiac resynchronization therapy can reverse abnormal myocardial strain distribution in patients with heart failure and left bundle branch block [J].
Breithardt, OA ;
Stellbrink, C ;
Herbots, L ;
Claus, P ;
Sinha, AM ;
Bijnens, B ;
Hanrath, P ;
Sutherland, GR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (03) :486-494
[8]   Comparative assessment of right, left, and biventricular pacing in patients with permanent atrial fibrillation [J].
Brignole, M ;
Gammage, M ;
Puggioni, E ;
Alboni, P ;
Raviele, A ;
Sutton, R ;
Vardas, P ;
Bongiorni, MG ;
Bergfeldt, L ;
Menozzi, C ;
Musso, G .
EUROPEAN HEART JOURNAL, 2005, 26 (07) :712-722
[9]   Assessment of atrioventricular junction ablation and VVIR pacemaker versus pharmacological treatment in patients with heart failure and chronic atrial fibrillation - A randomized, controlled study [J].
Brignole, M ;
Menozzi, C ;
Gianfranchi, L ;
Musso, G ;
Mureddu, R ;
Bottoni, N ;
Lolli, G .
CIRCULATION, 1998, 98 (10) :953-960
[10]   Assessment of atrioventricular junction ablation and DDDR mode-switching pacemaker versus pharmacological treatment in patients with severely symptomatic paroxysmal atrial fibrillation - A randomized controlled study [J].
Brignole, M ;
Gianfranchi, L ;
Menozzi, C ;
Alboni, P ;
Musso, G ;
Bongiorni, MG ;
Gasparini, M ;
Raviele, A ;
Lolli, G ;
Paparella, N ;
Acquarone, S .
CIRCULATION, 1997, 96 (08) :2617-2624