Coupling of an active middle-ear implant to the long process of the incus using an elastic clip attachment

被引:14
作者
Schraven, Sebastian P. [1 ]
Mlynski, Robert [2 ]
Dalhoff, Ernst [3 ]
Heyd, Andreas [3 ]
Wildenstein, Daniela [3 ]
Rak, Kristen [1 ]
Radeloff, Andreas [1 ]
Hagen, Rudolf [1 ]
Gummer, Anthony W. [3 ]
机构
[1] Univ Wurzburg, Dept Otorhinolaryngol Plast Aesthet & Reconstruct, Comprehens Hearing Ctr, Josef Schneider Str 11, D-97080 Wurzburg, Germany
[2] Univ Med Ctr Rostock, Dept Otorhinolaryngol Head & Neck Surg Otto Korne, Doberaner Str 137-139, D-18057 Rostock, Germany
[3] Univ Tubingen, Dept Otolaryngol, Sect Physiol Acoust & Commun, Elfriede Aulhorn Str 5, D-72076 Tubingen, Germany
关键词
Middle-ear surgery; Active middle-ear implant; Incus vibroplasty; FMT lncus Laser Doppler vibrometer; LP-coupler; SENSORINEURAL HEARING-LOSS; VIBRANT SOUNDBRIDGE IMPLANT; ROUND-WINDOW STIMULATION; HUMAN TEMPORAL BONES; STAPES SURGERY; DEVICE; FIXATION; TRANSDUCER; EXPERIENCE;
D O I
10.1016/j.heares.2016.03.012
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
The active middle-ear implant Vibrant Soundbridge (VSB) is used to treat mild-to-severe sensorineural hearing losses. The standard surgical approach for incus vibroplasty is a mastoidectomy and a posterior tympanotomy, crimping the Floating Mass Transducer (FMT) to the long process of the incus (LPI) (standard crimped application). However, tight crimping increases the risk of necrosis of the LPI, resulting in reduction of energy transfer and loss of amplification. The aim of this study was to develop a new coupling device for the LPI, that does not require crimping, and to test its vibrational transfer properties in temporal-bone preparations. An extended antrotomy and a posterior tympanotomy were performed in ten fresh human temporal bones. As a control for normal middle-ear function, the tympanic membrane was stimulated acoustically and the vibration of the stapes footplate was measured by laser Doppler vibrometry (LDV). FMT-induced vibration responses of the stapes were then measured for the standard crimped application at the LPI and for the newly designed elastic long process coupler (LP coupler). For the LP coupler, velocity amplitude responses in temporal-bone preparations showed increased mean amplitudes at around 1 kHz (similar to 10 dB) and a reduction between 1.8 and 6 kHz (13 dB on average for 2 <= f <= 5 kHz). In conclusion, attachment of the FMT to the LPI with the LP coupler leads to generally good mechanical and functional coupling in temporal-bone preparations with a notable disadvantage between 1.8 and 6 kHz. Due to its elastic clip attachment it is expected that the LP coupler will reduce the risk of necrosis of the incus long process, which has to been shown in further studies. Clinical results of the LP coupler are pending. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:179 / 184
页数:6
相关论文
共 24 条
[11]   Long-term assessment after implantation of the Vibrant Soundbridge device [J].
Schmuziger, N ;
Schimmann, F ;
àWengen, D ;
Patscheke, J ;
Probst, R .
OTOLOGY & NEUROTOLOGY, 2006, 27 (02) :183-188
[12]   Long-term Stability of the Active Middle-ear Implant with Floating-mass Transducer Technology: A Single-center Study [J].
Schraven, Sebastian P. ;
Grossmann, Wilma ;
Rak, Kristen ;
Shehata-Dieler, Wafaa ;
Hagen, Rudolf ;
Mlynski, Robert .
OTOLOGY & NEUROTOLOGY, 2016, 37 (03) :252-266
[13]  
Schraven SP, 2015, OTOL NEUROTOL, V36, P22, DOI 10.1097/MAO.0000000000000593
[14]   Alternative Fixation of an Active Middle Ear Implant at the Short Incus Process [J].
Schraven, Sebastian P. ;
Dalhoff, Ernst ;
Wildenstein, Daniela ;
Hagen, Rudolf ;
Gummer, Anthony W. ;
Mlynski, Robert .
AUDIOLOGY AND NEURO-OTOLOGY, 2014, 19 (01) :1-11
[15]   Conditions for Highly Efficient and Reproducible Round-Window Stimulation in Humans [J].
Schraven, Sebastian P. ;
Hirt, Bernhard ;
Goll, Erich ;
Heyd, Andreas ;
Gummer, Anthony W. ;
Zenner, Hans-Peter ;
Dalhoff, Ernst .
AUDIOLOGY AND NEURO-OTOLOGY, 2012, 17 (02) :133-138
[16]   Controlled round-window stimulation in human temporal bones yielding reproducible and functionally relevant stapedial responses [J].
Schraven, Sebastian P. ;
Hirt, Bernhard ;
Gummer, Anthony W. ;
Zenner, Hans-Peter ;
Dalhoff, Ernst .
HEARING RESEARCH, 2011, 282 (1-2) :272-282
[17]   Audiometric evaluation of an attempt to optimize the fixation of the transducer of a middle-ear implant to the ossicular chain with bone cement [J].
Snik, A ;
Cremers, C .
CLINICAL OTOLARYNGOLOGY, 2004, 29 (01) :5-9
[18]  
Snik AFM, 2001, OTOLARYNG CLIN N AM, V34, P373
[19]   Human temporal bones versus mechanical model to evaluate three middle ear transducers [J].
Stieger, Christof ;
Bernhard, Hans ;
Waeckerlin, Daniela ;
Kornpis, Martin ;
Burger, Juergen ;
Haeusler, Rudolf .
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2007, 44 (03) :407-415
[20]   Comparison of different vibrant Soundbridge audioprocessors with conventional hearing aids [J].
Todt, I ;
Seidl, RO ;
Gross, M ;
Ernst, A .
OTOLOGY & NEUROTOLOGY, 2002, 23 (05) :669-673