Pacing in the middle cardiac vein in a patient with tricuspid prosthesis

被引:10
作者
Nguyen, LS
Swaroop, S
Prejean, CA
机构
[1] Fountain Valley Reg Hosp, Dept Cardiol, Fountain Valley, CA USA
[2] Fountain Valley Reg Hosp, Dept Cardiovasc Surg, Fountain Valley, CA USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 02期
关键词
middle cardiac vein; tricuspid prosthesis; epimyocardial pacing;
D O I
10.1046/j.1460-9592.2002.00243.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A man with acute endocarditis developed complete heart block several days after the tricuspid and aortic valve replacement. Several weeks after implantation, his epimyocardial pacing leads developed a high threshold and failed to capture the ventricle at the maximal pulse width and output of the pacemaker. An angled-tip lead was placed in the middle cardiac vein for ventricular pacing. The pacing and sensing thresholds of this lead were within the expected range during follow-up. Therefore, in patients with prosthetic tricuspid valve, pacing in the middle cardiac vein should be considered before open-chest placement of the epimyocardial lead.