Usefulness of Renal Volume Measurements Obtained by a 3-Dimensional Sonographic Transducer With Matrix Electronic Arrays

被引:23
作者
Kim, Hyun Cheol [1 ]
Yang, Dal Mo [1 ]
Lee, Sang Ho [2 ]
Cho, Yong Duck [2 ]
机构
[1] Kyung Hee Univ, Coll Med, E W Neo Med Ctr, Dept Radiol, Seoul 134727, South Korea
[2] Kyung Hee Univ, Coll Med, E W Neo Med Ctr, Dept Internal Med, Seoul 134727, South Korea
关键词
computed tomography; kidney; sonography; 3-dimensional sonography; volume measurement;
D O I
10.7863/jum.2008.27.12.1673
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective. The purpose of this study was to examine the feasibility of 3-dimensional (3D) sonography using a matrix array transducer to measure renal volume. Methods. One hundred consecutive patients with a normal serum creatinine level and kidney appearance on computed tomography (CT) performed within 2 months before sonography were enrolled in this study. Two hundred individual renal volumes were blindly obtained by the ellipsoid formula, the stacked ellipse method, the voxel count method using routine 2-dimensional (2D) sonographic data, 3D sonographic data using a matrix array transducer, and CT data, respectively. The voxel count method was validated as the reference standard by the water displacement method in 10 cadaveric pig kidneys (r = 0.99; P < .001). Renal volumes determined by 2D and 3D sonography were compared with volumes determined by CT Results. Volumes determined by 2D sonography were significantly lower than those determined by CT (P < .001) but similar to those determined by 3D sonography (P = .78). The percent volume error of 3D sonography (mean +/- SD, -2.2% +/- 3.7%) was significantly lower than that of 2D sonography (-15.7% +/- 11.8%) with CT as the standard (P < .001). The correlation coefficient between 3D sonography and CT (r = 0.98; P < .0001) was better than that between 2D sonography and CT (r = 0.831; P < .0001). In addition, Bland-Altman analysis revealed that the limits of agreement between 3D sonography and CT (-9.7% to 5.1 %) were narrower than those between 2D sonography and CT (-45.6% to 9.8%). Conclusions. Three-dimensional sonography with a matrix array transducer can significantly reduce renal volume measurement errors and offers a reliable means of determining renal volumes.
引用
收藏
页码:1673 / 1681
页数:9
相关论文
共 23 条
[1]   In vitro measurement of kidney size: Comparison of ultrasonography and MRI [J].
Bakker, J ;
Olree, M ;
Kaatee, R ;
de Lange, EE ;
Beek, FJA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (05) :683-688
[2]   Renal volume measurements: Accuracy and repeatability of US compared with that of MR imaging [J].
Bakker, J ;
Olree, M ;
Kaatee, R ;
de Lange, EE ;
Moons, KGM ;
Beutler, JJ ;
Beek, FJA .
RADIOLOGY, 1999, 211 (03) :623-628
[3]   Normal values for renal length and volume as measured by magnetic resonance imaging [J].
Cheong, Benjamin ;
Muthupillai, Raja ;
Rubin, Mario F. ;
Flamm, Scott D. .
CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2007, 2 (01) :38-45
[4]   Clinical utility of three-dimensional US [J].
Downey, DB ;
Fenster, A ;
Williams, JC .
RADIOGRAPHICS, 2000, 20 (02) :559-571
[5]   Volume ultrasound: the next big thing? [J].
Elliott, S. T. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (961) :8-9
[6]   Three-dimensional ultrasound imaging and its use in quantifying organ and pathology volumes [J].
Fenster, A ;
Downey, DB .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 377 (06) :982-989
[7]   A clinical MRI investigation of the relationship between kidney volume measurements and renal function in patients with renovascular disease [J].
Gandy, S. J. ;
Armoogum, K. ;
Nicholas, R. S. ;
McLeay, T. B. ;
Houston, J. G. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (949) :12-20
[8]   IN-VIVO COMPARISON OF 3D ULTRASONOGRAPHY AND MAGNETIC-RESONANCE-IMAGING IN VOLUME ESTIMATION OF HUMAN KIDNEYS [J].
GILJA, OH ;
SMIEVOLL, AI ;
THUNE, N ;
MATRE, K ;
HAUSKEN, T ;
ODEGAARD, S ;
BERSTAD, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (01) :25-32
[9]   Applications of 2-dimensional matrix array for 3-and 4-dimensional examination of the fetus -: A pictorial essay [J].
Goncalves, Luis F. ;
Espinoza, Jimmy ;
Kusanovic, Juan Pedro ;
Lee, Wesley ;
Nien, Jyh Kae ;
Santolaya-Forgas, Joaquin ;
Mari, Giancarlo ;
Treadwell, Marjorie C. ;
Romero, Roberto .
JOURNAL OF ULTRASOUND IN MEDICINE, 2006, 25 (06) :745-755
[10]   Impact of 3D sonography on workroom time efficiency [J].
Hagel, Jeffrey ;
Bicknell, Simon G. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (04) :966-969