Evaluation of a Ventricular Assist Device System: Stability in a Proton Beam Therapy

被引:10
作者
Gossman, Michael S. [1 ,2 ]
Graham, Joel D. [3 ]
Das, Indra J. [4 ,5 ]
Cheng, Chee-wai [4 ]
Coutinho, Len [4 ]
Tamez, Dan [3 ]
Larose, Jeffrey A. [3 ]
机构
[1] Tristate Reg Canc Ctr, Dept Radiat Oncol, Med Phys Sect, Ashland, KY 41101 USA
[2] Regulat Direct Med Phys LLC, Flatwoods, KY USA
[3] HeartWare Inc, Miami Lakes, FL USA
[4] Indiana Univ, Hlth Proton Therapy Ctr, Bloomington, IN 47405 USA
[5] Indiana Univ Sch Med, Dept Radiat Oncol, Indianapolis, IN USA
关键词
left ventricular assist device; HeartWare left ventricular assist device; radiation; ventricular assist device; proton beam treatment; NEUTRON DOSE-EQUIVALENT; CELL LUNG-CANCER; CARDIAC-PACEMAKERS; LATEST GENERATION; RADIOTHERAPY;
D O I
10.1097/MAT.0b013e31826feeca
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Inadequate research exists regarding testing of a ventricular assist device (VAD) for susceptibility to radiation damage. Specifically, minimal data are available to radiation oncologists prescribing treatment plans for patients with an implanted VAD. As the number of implanted devices increases, patients requiring radiation at tissue sites near or at the device will increase. The purpose of this study is to provide the first analysis of radiation effects of proton beams on VADs. Five left VAD (LVAD) pumps (HeartWare Inc., Miami Lakes, FL) were exposed to proton beam radiation at a calibrated dose rate of 5 Gy/min up to a cumulative dose of 70 Gy. The HeartWare LVAD pump recorded parameters including power (W), speed (revolutions/min), and estimated flow (L/min). Analysis of collected data after each irradiation found no deviation in pump parameters from baseline values. The HeartWare LVAD pump exhibited no change in device function when directly irradiated by a high energy proton beam. Secondary neutron fluence created in the proton beam during irradiation had no effect on external components including the system controller and batteries powering the HeartWare LVAD. ASAIO Journal 2012;58:597-600.
引用
收藏
页码:597 / 600
页数:4
相关论文
共 22 条
  • [1] Andreo P, 2000, IAEA TECHNICAL REPOR
  • [2] Proton-beam radiotherapy for early-stage lung cancer
    Bush, DA
    Slater, JD
    Bonnet, R
    Cheek, GA
    Dunbar, RD
    Moyers, M
    Slater, JM
    [J]. CHEST, 1999, 116 (05) : 1313 - 1319
  • [3] Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking
    Farr, J. B.
    Mascia, A. E.
    Hsi, W. -C.
    Allgower, C. E.
    Jesseph, F.
    Schreuder, A. N.
    Wolanski, M.
    Nichiporov, D. F.
    Anferov, V.
    [J]. MEDICAL PHYSICS, 2008, 35 (11) : 4945 - 4954
  • [4] IMPLANTED CARDIAC DEFIBRILLATOR CARE IN RADIATION ONCOLOGY PATIENT POPULATION
    Gelblum, Daphna Y.
    Amols, Howard
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (05): : 1525 - 1531
  • [5] Vagus nerve stimulator stability and interference on radiation oncology x-ray beams
    Gossman, Michael S.
    Ketkar, Amruta
    Liu, Arthur K.
    Olin, Bryan
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20) : N365 - N376
  • [6] Evaluation of a Ventricular Assist Device: Stability Under X-Rays and Therapeutic Beam Attenuation
    Gossman, Michael S.
    Graham, Joel D.
    Tamez, Dan
    Voskoboynikov, Neil
    Larose, Jeffrey A.
    [J]. ASAIO JOURNAL, 2012, 58 (03) : 212 - 216
  • [7] Influence of beam efficiency through the patient-specific collimator on secondary neutron dose equivalent in double scattering and uniform scanning modes of proton therapy
    Hecksel, D.
    Anferov, V.
    Fitzek, M.
    Shahnazi, K.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2910 - 2917
  • [8] Hecksel D, 2007, AUSTRALAS PHYS ENG S, V30, P326
  • [9] Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators
    Hurkmans, CW
    Scheepers, E
    Springorum, BGF
    Uiterwaal, H
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (01): : 282 - 289
  • [10] Influence of radiotherapy on the latest generation of pacemakers
    Hurkmans, CW
    Scheepers, E
    Springorum, BGF
    Uiterwaal, H
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) : 93 - 98